Instruction manual

NGA 2000 MLT Hardware
TEST PROCEDURE / TEST POINTS
11 - 5
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
1 ( )
not used
not used
2 (signal)
3 (- 6,3 V dc)
4
5 (+ 6,3 V dc)
11.1.6 Physical Zero Alignment
For physical zero adjustment the raw signal of respective measuring channel is set to ± 100.000
counts while purging instrument with zero gas.
The raw measuring values can be shown by pressing "Status" (F2) "RawMeas." (F2).
11.1.6.1 IR Measurement
For set of physical zero adjustment see Item 17.5 please.
11.1.6.2 Paramagnetic Oxygen Measurement
Set physical zero adjustment by alteration the position of photodiode of the sensor.
A further test is to measure the output voltage of the sensor while purging with test gas.
Test point: “Pin 2” of P23 of PCB “PIC” (Fig. 1-32, too)
Signal: + 4 V dc
(purging with test gas 25 Vol.-% / 100 Vol.-% oxygen (O
2
), depending
on installed sensor)
11.1.6.3 Electrochemical Oxygen Measurement (EO
2
)
For set / check of sensor see Item 18.2 and 18.3.4 please.
PHYSICAL ZERO ALIGNMENT