Instruction manual
NGA 2000 MLT Hardware
TROUBLESHOOTING
10 - 2
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
10.2 No / Incorrect Measurement Screen
Possible Causes Check / Correct
a) Defective network connection: Check network termination (see section 1.11).
Check network connection between platform,
analyzer, external modules.
Replace connection cable or network module
PCB “LEM” if necessary.
b) Analyzer module has not been integrated Integrate the analyzer module in the
in the system software (network): system software (see software manual).
c) Failure in signal processing: See Section 11.1.
d) EMC radiation via PCB “SIO/DIO” Incorrect wiring of external actors (section 5.5)
e) PCB “AFP” board or LC display defective Exchange front panel (see 12.2)
f) PCB “ACU” defective: Replace ACU 02 (see 12.3)
NO MEASUREMENT SCREEN