NGA 2000 MLT Hardware
INSTALLATION
5 - 29
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
This analyzer model is supplied by an internatl wide range power supply, automatically
adapting to the voltage at site.
All cables (mains power cord and signal cables) are connected via screw terminals located
inside an junction box right above the analyzer dome.
See the label on the junction box cover for detailed terminal assignments.
Cables entering the junction box should be kept as short as possible when inside the box.
Mains terminals are capable to connect cables up to 4 mm
2
cross section.
ANALYZER SPECIFIC INSTRUCTIONS: CAT 200
EXPLOSION HAZARD!
Instruments covered by an approval for hazardous areas (identifiable by a
term „ATEX" on the nameplate label) MUST be equipped with cable glands,
marked EEx e and ATEX!
This applies, too, when the instrument currently is NOT installed in a
hazardous area!