Instruction manual
NGA 2000 MLT Hardware
INSTALLATION
5 - 29
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
❏ This analyzer model is supplied by an internatl wide range power supply, automatically
adapting to the voltage at site.
❏ All cables (mains power cord and signal cables) are connected via screw terminals located
inside an junction box right above the analyzer dome.
❏ See the label on the junction box cover for detailed terminal assignments.
❏ Cables entering the junction box should be kept as short as possible when inside the box.
❏ Mains terminals are capable to connect cables up to 4 mm
2
cross section.
ANALYZER SPECIFIC INSTRUCTIONS: CAT 200
EXPLOSION HAZARD!
Instruments covered by an approval for hazardous areas (identifiable by a
term „ATEX" on the nameplate label) MUST be equipped with cable glands,
marked EEx e and ATEX!
This applies, too, when the instrument currently is NOT installed in a
hazardous area!