Instruction manual
NGA 2000 MLT Hardware
INSTALLATION
5 - 17
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
Cables for external data processing must be double insulated for mains
voltage when used inside the instrument!
If double insulation not possible signal cables inside the analyzer must be
installed in a way that a distance of at least 5 mm is ensured permanently
(e.g. by utilizing cable ties).
The number of installed adaptor elements variies depending on analyzer configuration.
For detailed pin designations see section 21.
❍ Opening of housing (front panel) (see section 15.2).
❍ Take lines via cable glands (Fig. 5-7) inside the housing.
Connection is to be done to the respective adaptor elements (Fig. 5-8 and 5-9).
All unused cable glands need to be sealed using a sealing plug (part no.
ETC00791 or similiar, see Fig. 5-11).
Unused cable gland openings in the enclosure need to be covered using a
special screw (part no. ETC 000790 or similiar, see Fig. 5-11).
ANALYZER SPECIFIC INSTRUCTIONS: MLT 2
Fig. 5-8: MLT 2, Data line connections
Inside view from front (detail, without front door)
Serial out
4321
X10
2
1
X11
X9
1
1
F100.2
F100.1
Adaptor element option SIO:
Option serial interface /
Relay contacts (NAMUR)
Adaptor element option SIO:
analog outputs
Adaptor element Option DIO:
digital inputs/outputs
Power supply(s)