Instruction Manual 90002929 01/2007 Instruction Manual NGA 2000 Hardware Manual for MLT or CAT 200 Analyzer and MLT or CAT 200 Analyzer Module (combined with NGA 2000 Platform, MLT, CAT 200 or TFID Analyzer) 10th Edition 01/2007 www.EmersonProcess.
NGA 2000 MLT Hardware Instruction Manual 90002929 07/2006 ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications.
Instruction Manual NGA 2000 MLT Hardware 90002929 01/2007 CONTENTS Table of Contents PREAMBEL .................................................................................... S - 1 DEFINITIONS.................................................................................. S - 1 SAFETY INSTRUCTIONS ............................................................. S - 2 Operating and maintaining this apparatus................................... S - 3 Intended use statement ..............................
NGA 2000 MLT Hardware Instruction Manual CONTENTS 90002929 01/2007 1.4 1.4.1 1.4.2 1.4.3 1.5 1.6 1.6.1 1.6.2 1.7 1.8 1.9 1.9.1 1.9.2 1.9.3 1.10 1.10.1 1.10.2 1.10.3 1.10.4 1.10.5 1.10.6 1.10.7 1.10.8 1.11 1.12 1.13 MLT 1 ............................................................................................... 1 - 6 MLT 1 1/2 19" housing ................................................................... 1 - 6 MLT 1 ULCO .............................................................................
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 CONTENTS OPERATION (START-UP) 5. 5.1 5.1.1 5.2 5.2.1 5.2.2 5.2.3 5.2.4 5.3 5.3.1 5.3.2 5.4 5.4.1 5.4.1.1 5.4.1.2 5.4.2 5.4.2.1 5.4.2.2 5.4.3 5.4.3.1 5.4.4 5.4.5 5.4.5.1 5.4.5.2 5.4.5.3 5.5 5.5.1 5.5.2 5.5.3 5.5.4 Installation .................................................................................... 5 - 1 General ...........................................................................................
NGA 2000 MLT Hardware Instruction Manual CONTENTS 90002929 07/2006 6. Switching On .................................................................................. 6 - 1 7. 7.1 7.2 7.2.1 7.2.1.1 7.2.1.2 7.3 Measurement / Calibration / Switching Off .................................. 7 - 1 Measurement ................................................................................... 7 - 1 Calibration ........................................................................................
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 CONTENTS 11. 11.1 11.1.1 11.1.2 11.1.3 11.1.4 11.1.5 11.1.6 11.1.6.1 11.1.6.2 11.1.6.3 11.2 11.3 11.3.1 11.3.2 11.3.3 11.3.4 11.3.4.1 11.3.4.2 Test Procedure / Test Points ....................................................... 11 - 1 Signal processing ........................................................................... 11 - 1 Internal Voltage Supply ................................................................ 11 - 2 IR Source .......
NGA 2000 MLT Hardware Instruction Manual CONTENTS 90002929 07/2006 MAINTENANCE ............................................................................. 13 - 1 14. Leak Testing ................................................................................. 14 - 1 15. 15.1 15.1.1 15.1.1.1 15.1.1.2 15.1.2 15.1.2.1 15.1.2.2 15.2 15.3 15.3.1 15.3.2 15.3.3 15.3.3.1 15.3.3.2 15.4 15.4.1 15.4.2 15.4.3 15.4.3.1 15.4.3.2 15.4.3.3 Opening the Housing ......................................................
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 CONTENTS 18. 18.1 18.2 18.3 18.3.1 18.3.2 18.3.3 18.3.4 18.4 Check / Replacement of electrochemical Oxygen Sensor ........ 18 - 1 EO2-Sensor .................................................................................... 18 - 2 Check of the EO2-Sensor ............................................................... 18 - 3 Replacement of the EO2-Sensor .................................................... 18 - 4 Removal of the EO2-Sensor .....
NGA 2000 MLT Hardware Instruction Manual CONTENTS 90002929 07/2006 SUPPLEMENT 21. 21.1 21.2 21.3 21.3.1 21.3.2 21.4 21.5 21.6 Pin Assignments ........................................................................ 21 24 V dc Input (MLT 1/4) ................................................................ 21 230/120 V ac Input (MLT 3) .......................................................... 21 Option SIO (Standard I/O) ............................................................
Instruction Manual 90002929 07/2006 NGA 2000 MLT Hardware CONTENTS LIST OF FIGURES LIST OF FIGURES Fig. Title Fig. P-1: Fig. P-2: Fig. P-3: Fig. P-4: Analyzer Nameplate Label (example) ............................................................................. P - 3 From separate analyzers to analyzer system ................................................................. P - 4 Example of NGA-cabling .................................................................................................
NGA 2000 MLT Hardware CONTENTS LIST OF FIGURES Instruction Manual 90002929 07/2006 Fig. Title Fig. 1-32: Fig. 1-33: Fig. 1-34: Fig. 1-35: Fig. 1-36: Fig. 1-37: Fig. 1-38: Plug pin assignment PCB PIC ...................................................................................... 1 - 38 Plug pin assignment PCB DSP ..................................................................................... 1 - 39 Function blocks of SIO-PCB ..................................................................
Instruction Manual 90002929 07/2006 Fig. NGA 2000 MLT Hardware CONTENTS LIST OF FIGURES Title Page Fig. 11-1: Fig. 11-2: Fig. 11-3: Fig. 11-4: Fig. 11-5: Fig. 11-6: Fig. 11-6a: Fig. 11-6b: Fig. 11-6c: Fig. 11-7: Fig. 11-7a: Fig. 11-7b: Fig. 11-8: Fig. 11-9: Principle signal processing with PCB "PSV" ................................................................. 11 - 1 PCB "VVS" ....................................................................................................................
NGA 2000 MLT Hardware CONTENTS LIST OF FIGURES/LIST OF TABLES Fig. Fig. 20Fig. 20Fig. 20Fig. 20Fig. 20Fig. 20- Title Instruction Manual 90002929 07/2006 Page 1: 2: 3: 4: 5: 6: Dimensional sketch MLT 1 [all dimensions in mm] ........................................................ 20 - 6 Dimensional sketch / Drill drawing MLT 2 Standard version ........................................... 20 - 7 Dimensional sketch / Drill drawing MLT 2 Dual housing version .....................................
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 SAFETY INSTRUCTIONS PREAMBEL / DEFINITIONS Preambel This instruction manual provides information about MLT and CAT200 series gas analyzers /analyzer modules concerning subassemblies, functions, procedures, installation, operation and maintenance. This instruction manual covers several MLT and CAT200 series gas analyzers /analyzer modules variations and therefore may describe configurations and/or options not part of your specific analyzer.
NGA 2000 MLT Hardware Instruction Manual SAFETY INSTRUCTIONS GENERAL 90002929 07/2006 IMPORTANT SAFETY INSTRUCTIONS WIRING AND INSTALLATION OF THIS APPARATUS The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non-EU states should also comply with the following unless superseded by local or National Standards. 1.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 SAFETY INSTRUCTIONS OPERATING AND MAINTAINING THIS APPARATUS Operating and maintaining this apparatus This instrument has left the factory in compliance with all applicable safety regulations. To maintain this operating condition, the user must strictly follow the instructions and consider the warnings in this manual or provided on the instrument.
NGA 2000 MLT Hardware Instruction Manual SAFETY INSTRUCTIONS INTENDED USE STATEMENT / SAFETY SUMMARY / AUTHORIZED PERSONNEL 90002929 07/2006 Intended use statement MLT and CAT200 series gas analyzers/analyzer modules are intended to be used as analyzers for industrial purposes.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 SAFETY INSTRUCTIONS EXPLOSION HAZARD ! Do not operate nor install these instruments in hazardous areas without additional measures! MLT 2 and CAT 200 -- HEAVY INSTRUMENTS ! The analyzer variations MLT 2 and CAT 200 intended to be wall mounted and/or outdoor installed weigh up to 35 kg resp.
NGA 2000 MLT Hardware Instruction Manual SAFETY INSTRUCTIONS GASES AND GAS CONDITIONING (SAMPLE HANDLING) 90002929 07/2006 Gases and Gas Conditioning (Sample Handling) Take care of the safety instructions applicable for the gases (sample gases and test gases) and for the gas bottles containing these gases! Before opening gas paths they must be purged with ambient air or neutral gas (N2) to avoid hazards caused by toxic, flammable, explosive or harmful to health sample gas components! Flammable gases
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 SAFETY INSTRUCTIONS POWER SUPPLY Power Supply Verify the mains voltage at site of installation corresponds to the analyzer´s rated voltage as given on the nameplate label! Verify the safety instruction given by power supply unit manufacturer ! Instruments with External Power Supply Unit The mains socket has to be nearby the power supply unit and easily accessible! Disconnecting from mains requires unplugging the power plug! To comply with the
NGA 2000 MLT Hardware Instruction Manual SAFETY INSTRUCTIONS ADDITIONAL HINTS FOR MLT 2 AND CAT 200 90002929 07/2006 ADDITIONAL HINTS FOR MODELS MLT 2 AND CAT 200 Installation and connecting mains and signal cables are subject to qualified personnel only taking into account all applicable standards and legislative requirements! Failure to follow may cause warranty invalidation, property damage and/or personal injury or death! Mains and signal cables need to be connected to internal srew terminals requir
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 SAFETY INSTRUCTIONS GENERAL OPERATING INSTRUCTIONS General operating instructions DANGER TO LIFE ! EXPLOSION HAZARD ! Verify all gas lines are connected as described within this manual and tight! Improper gas connections may cause explosion, serious injury or death! Exhaust may contain hydrocarbons and other toxic gases, e.g. carbon monoxide.
NGA 2000 MLT Hardware Instruction Manual SAFETY INSTRUCTIONS ADDITIONAL HINTS FOR UV MEASUREMENT 90002929 07/2006 Additional hints for UV measurement HIGH VOLTAGE ! The optional UV lamp operates with high voltage (Power Supply UVS) ! UV SOURCE ! Ultraviolet light from UV lamp can cause permanent eye damage ! Do not look directly at the ultraviolet source ! TOXIC SUBSTANCE ! The optional UV lamp contains mercury.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 SAFETY INSTRUCTIONS MAGNETICALLY OPERATED FRONT PANEL Magnetically Operated Front Panel DANGER TO LIFE ! Persons with cardiac pacemakers should absolutely avoid magnetic fields ! Negative effects on persons beyond those described above caused by magnetic fields are not known. It is presumed that persons showing allergic reaction on contact with ceramic or metallic material show the same behavior on contact with magnetic material.
NGA 2000 MLT Hardware Instruction Manual SAFETY INSTRUCTIONS ELECTROSTATIC DISCHARGE 90002929 07/2006 Electrostatic Discharge The electronic parts of the analyzer can be irreparably damaged if exposed to electrostatic discharge (ESD). The instrument is ESD protected when the covers have been secured and safety precautions observed. When the housing is open, the internal components are not ESD protected anymore.
Instruction Manual NGA 2000 MLT Hardware 90002929 12/2006 PREFACE Preface a) Analyzer versions - Standard General Purpose Applications The MLT series of NGA 2000 analyzers offers multi-component, multi-method analysis. Different measurement methods can be combined in one analyzer. MLT 1, MLT 2 and MLT 4 (MLT 5) are designed to measure up to 5 gas components while MLT 1 ULCO, MLT 3 and CAT 200 allow up to 4 gas components (including photometer and nonphotometer channels).
NGA 2000 MLT Hardware PREFACE b) Instruction Manual 90002929 12/2006 Analyzer versions - Installation in Hazardous Areas Note! This manual does not deal with special conditions for analyzers in hazardous areas, related to installation, operation, maintenance etc. For such applications refer to the separate instruction manuals, delivered together with the analyzers.For installation in hazardous areas the MLT 2 can be equipped with the appropriate pressurization system acc.
Instruction Manual NGA 2000 MLT Hardware 90002929 12/2006 PREFACE All analyzer versions are marked as follows: MLTx y-CH1 CH2 CH3 CH4 CH5 with x = analyzer type 1, 2, 3, 4, 5, CAT 200 with 1 = 1/2 19", not thermostatted, external power supply 2 = Field housing, thermostatted, internal power supply 3 = 1/1 19"-housing, thermostatted, internal power supply 4 = 1/1 19"-housing, thermostatted, external power supply 5 = 19", 18 to 21 HU housing, thermostatted, internal power supply CAT 200 = flameproof encl
NGA 2000 MLT Hardware Instruction Manual PREFACE c) 90002929 12/2006 Analyzer System Architecture The MLT´s flexibility facilitates the most cost-effective system architecture - elegantly accommodating either “stand-alone” or integrated multi-channel analyzer requirements. The MLT is available both as an “Analyzer Module” or as an “Analyzer”. The “Analyzer Module” (AM) is a “blind” analysis unit but retains all the advanced MLT design features.
Instruction Manual NGA 2000 MLT Hardware 90002929 12/2006 PREFACE The modular erection with a bi-direction network makes possible ❏ Single devices (stand-alone analyzers) - analyzer modules in a platform (separate manual) including optional inputs and outputs (e.g. SIO/DIO). - MLT analyzer including optional inputs and outputs (e.g. SIO/DIO). ❏ The interconnection on simple way of analyzer modules including optional local inputs and outputs (e.g.
NGA 2000 MLT Hardware Instruction Manual PREFACE 90002929 12/2006 Local I/O’s via Internal System Bus DIO Analyzer Modules (AM’s) O2 PMD HC HFID NOx WCLD NO/NOx CLD System I/O’s via Internal System Bus ppm O2 TO2 HC FID HC TFID DIO SIO DIO CO/NO/SO2 / EO2 - MLT PO 2 MLT NGA 200 0 Alternatively to Platform or MLT/TFID Analyzer Alternatively SIO 19" Platform/ TFID Analyzer Field PC Workstation ROS EM OUN T 8 Digital Inputs 24 Digital Outputs 3 Relay Outputs RS 232 / RS 485 Printer
Instruction Manual NGA 2000 MLT Hardware 90002929 12/2006 PREFACE System Unit SIO/DIO Configuration MLT analyzer module (AM): • without front panel, i.e.
NGA 2000 MLT Hardware PREFACE P-8 Instruction Manual 90002929 12/2006 Emerson Process Management GmbH & Co.
Instruction Manual 90002929 01/2007 NGA 2000 MLT Hardware TECHNICAL DESCRIPTION STANDARD GENERAL PURPOSE APPLICATIONS/INSTALLATION IN HAZARDOUS AREAS 1. Technical Description 1.1 Standard General Purpose Applications The different analyzers and analyzer modules are assembled in principle identically. All components of analyzers or analyzer modules are incorporated into a housing for platform mounting (MLT 1), a 1/2 19" housing (MLT 1) or a 1/1 19" housing (MLT 3/4).
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 OPERATING FRONT PANEL: STANDARD VERSION 1.3 Operating Front Panel 1.3.1 Standard Version The front panel of the analyzer is the operating front panel (see Fig. 1-1). Measured values and the entire operating procedure are displayed on a LC display.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 1.3.2 TECHNICAL DESCRIPTION OPERATING FRONT PANEL: MAGNETICALLY OPERATED TOUCH PANEL Magnetically Operated Touch Panel Analyzers intended either to be used in hazardous area or for outdoor usage (e.g. MLT 2 or CAT 200) are provided with a magnetically operated touch panel. This special version ensures ingress protection (IP) and resistance against mechanical shocks.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 OPERATING FRONT PANEL: MAGNETICALLY OPERATED TOUCH PANEL 1.3.2.2 Accessories To operate the magnetically operated touch panel a special tool (actuator) is required.The following picture shows this tool as it should be provided together with your analyzer. If it is missing please contact your local sales office. Fig. 1-3a: Magnetic actuator 1.3.2.3 Fig.
Instruction Manual 90002929 07/2006 1.3.2.4 NGA 2000 MLT Hardware TECHNICAL DESCRIPTION OPERATING FRONT PANEL: MAGNETICALLY OPERATED TOUCH PANEL Storing the actuator tool The tool is provided with a chain and a special fixture to be installed at one of the analyzer’s mounting screws (see Fig. 1-5). It is recommended to place the actuator tool on the fixture as shown to avoid unintended interference while not in use. Fig. 1-5: Actuator stored when not in use Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 MLT 1 (1/2 19" HOUSING) 1.4 MLT 1 The different analyzers and analyzer modules are assembled in principle identically. All components of analyzers or analyzer modules are incorporated into a housing for platform mounting (MLT 1) or a 1/2 19" housing (MLT 1). The equipment is specified for an operating voltage of 24 V DC (± 5 %). There are mounted optional different components at the rear side of the front panel (Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION MLT 1 (1/2 19" HOUSING) UVS (Voltage Supply for UV Source) [Option] Fan Pressure Sensor [Option] OUT IN O2 Sensor, electrochemical [Option] Fig. 1-7: MLT 1, Front panel, Rear view Network connection (RJ 45 socket) [not for platform mounting module] optional PCBs [SIO / DIO e. g.] Gas connections Input 24 V dc [not for platform mounting module] Fig.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL CONSTRUCTION MLT1 (1/2 19" HOUSING) Gas Connections Solenoid Valves (Option in preparation) Gas Connections Photometer (MLT specific assembly) Cardcage (see section 1.10) Flow Sensor (Option) O2 Sensor, electrochemical (Option) Voltage Supply for UV Source (Option) Fan Pressure Sensor (Option) Operating Front Panel (analyzer) or Blind Plate (analyzer module) Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION INTERNAL CONSTRUCTION MLT1 (1/2 19" HOUSING) Gas Connections Flow Sensor (Option) O2 Sensor, paramagnetic (Option) Cardcage (see section 1.10) Photometer (MLT specific assembly) Voltage Supply for UV Source (Option) Fan Pressure Sensor (Option) Operating Front Panel (analyzer) or Blind Plate (analyzer module) Fig.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL CONSTRUCTION MLT1 (1/2 19" HOUSING) Gas Connections Gas Connections Solenoid Valves (Option in preparation) Photometer (MLT specific assembly) Cartridge (see Item 1.10) Flow Sensor (Option) O2 Sensor, paramagnetic (Option) Pressure Sensor (Option) Voltage Supply for UV Source (Option) Operating Front Panel (analyzer) or Blind Plate (analyzer module) Fan Fig.
Instruction Manual 90002929 01/2007 1.4.2 NGA 2000 MLT Hardware TECHNICAL DESCRIPTION MLT1 ULCO MLT 1 ULCO NGA 2000 MLT 1 ULCO gas analyzer is specially designed to measure ultra low carbon monoxide in a matrix with high water vapor and/or carbon dioxide contents. The analyzer is equipped with a 2nd optical bench including a multi detector assembly (MDA block) for cross interference compensation, especially being designed for automotive and flue gas applications.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL CONSTRUCTION MLT1-ULCO Gas Connections Detector COlow Flow Sensor (Option) Cardcage (see Item 1.10) Detector COhigh (Option, ICEE/CEMS only) Multi Detector Assembly MDA (CO2/H2O detector for cross compensation and CO2 measurement, ICEE/CEMS only) Fan Pressure Sensor (Option) Operating Front Panel (analyzer) or Blind Plate (analyzer module) Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 1.4.3 TECHNICAL DESCRIPTION MLT1 (HOUSING FOR PLATFORM MOUNTING) MLT 1 Housing for platform mounting MLT 1 analyzer module housings for platform mounting are available to built-in into a NGA platform only (M) or for external installation and platform mounting (A) connected via NGA network. Analyzer modules for platform mounting have a blind plate too (Fig. 1-x).
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL CONSTRUCTION MLT1 (HOUSING FOR PLATFORM MOUNTING) Gas Connections Gas Connections Solenoid Valves (Option in preparation) Photometer (MLT specific assembly) Cardcage (see Item 1.10) Flow Sensor (Option) O2 Sensor, electrochemical (Option) Voltage Supply for UV Source (Option) Fan Pressure Sensor (Option) 24 V dc Input / Network Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION INTERNAL CONSTRUCTION MLT1 (HOUSING FOR PLATFORM MOUNTING) Gas Connections Flow Sensor (Option) O2 Sensor, paramagnetic (Option) Cardcage (see Item 1.10) Photometer (MLT specific assembly) Voltage Supply for UV Source (Option) Fan Pressure Sensor (Option) 24 V dc Input / Network Fig. 1-15: MLT 1, Platform mounting, Top view (with paramagnetical O2 Sensor) Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL CONSTRUCTION MLT1 (HOUSING FOR PLATFORM MOUNTING) Gas Connections Solenoid Valves (Option in preparation) Gas Connections Photometer (MLT specific assembly) Cardcage (see Item 1.10) Flow Sensor (Option) O2 Sensor, paramagnetic (Option) Fan Pressure Sensor (Option) Voltage Supply for UV Source (Option) 24 V dc Input / Network Fig.
Instruction Manual 90002929 07/2006 1.5 NGA 2000 MLT Hardware TECHNICAL DESCRIPTION MLT2 (FIELD HOUSING) MLT 2 (Field Housing) With MLT 2 Emerson Process Management offers a gas analyzer designed for wall mounting and outdoor installation due to it‘s stainless steel enclosure. All MLT 2 components are incorporated into a wall-mountable housing with ingress protection code IP 65 (approx. NEMA 4/4X) according to EN 60529.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 MLT2 (FIELD HOUSING) Connection of mains and interface signals are to be established via internal screw terminal blocks. This requires working inside an opened analyzer, near by potentially live components! Installation of this analyzer is allowed by qualified personnel only, familiar with the potential risks of working near live components! The model MLT 2 (field housing) does not provide a mains switch.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION MLT2 (FIELD HOUSING) Wall mounting holder Open Open Open Fastener front panel Operation front panel Open Open Open Front panel Fig. 1-17a: Front view MLT 2 (Standard housing/Standard operation front panel) Fig. 1-17b: Front view MLT 2 (Dual compartment housing/Standard operation front panel) Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL CONSTRUCTION MLT2 (FIELD HOUSING) Terminal Strips Option SIO (RS 232/485 and Relay Contacts) Fuses (T 6.3A/250V) X11 X9 Circuit Boards (see Item 1.10) X10 Serial out 12 1234 F100.1 F100.2 11 Terminal Strips Terminal Strips Option SIO Photometer sliding carriage Option DIO (analog outputs) (Fig. 1-18b) Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 1.6 TECHNICAL DESCRIPTION MLT3 MLT 3 All components of analyzers or analyzer modules are incorporated into a 1/1 19" housing. The housings are available as rack mounting (R) or as table-top (T) versions. For analyzer modules there is mounted a blind plate instead of an operation front panel. The equipment has an internal power supply with “autoranging” for operating voltages of 230 V AC or 120 V AC resp., 47-63 Hz.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL CONSTRUCTION MLT3 Gas Connections Photometer (MLT specific assembly) Power Supply (UPS 01 T) O2 Sensor, paramagnetic (Option) Cardcage (see Item 1.
Instruction Manual NGA 2000 MLT Hardware 90002929 01/2007 1.6.1 TECHNICAL DESCRIPTION MLT3 (GAS PURITY MEASUREMENT) MLT 3 (Gas purity measurement) Special versions of MLT 3 are available with suppressed ranges for gas purity measurements and measurements requiring physics thermostated up to 120 °C - see separate manual for gas purity with suppressed ranges.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 01/2007 MLT3 (GAS PURITY MEASUREMENT) ! Span gas DIGITAL I/O 230/120 V 50/60 Hz max. 240 VA NETWORK Sub.-min. D, 9 pole (connection to PCB “SIO”) Exhaust Sample gas ANALOG I/O SERIAL I/O Sub.-min. D, 9 pole (to Solenoid Valve Block Option) Zero gas PCB “SIO” Input 230/120 V ac (PS UPS 01 T) Solenoid Valve Block (Option) Gas connections (analyzer specific) Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION INTERNAL CONSTRUCTION MLT3 (GAS PURITY MEASUREMENT) Gas connections Photometer (MLT specific assembly) Solenoid valve block (Option) Power Supply (UPS 01 T) O2 Sensor, paramagnetic (Option) Cardcage (see Item 1.10) Heating Unit Pressure regulator Fan Flow meter Manual 4/2-way-valve (Option) Pressure sensor Operating Front Panel (analyzer) or Blind Plate (analyzer module) Fig.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 MLT4 1.7 MLT 4 All components of analyzers or analyzer modules are incorporated into a 1/1 19" housing. The housings are available as rack mounting (R) or as table-top (T) versions. For analyzer modules there is mounted a blind plate instead of an operation front panel. The equipment is specified for an operating voltage of 24 V DC (± 5 %). F1 F2 F3 F4 F5 N GA 2000 Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION INTERNAL CONSTRUCTION MLT4 Gas Connections Gas Connections O2 Sensor, paramagnetic (Option) Photometer (MLT specific assembly) Photometer (MLT specific assembly) Cardcage (see Item 1.10) Heating Unit Fan Voltage Supply for UV Source (Option) O2 Sensor (Option), electrochemical Voltage Supply for UV Source (Option) Operating Front Panel (analyzer) or Blind Plate (analyzer module) Fig.
NGA 2000 MLT Hardware TECHNICAL DESCRIPTION CAT 200 1.8 Instruction Manual 90002929 07/2006 CAT 200 The CAT 200 is an analyzer with flameproof enclosure, designed especially for operation in hazardous areas. It is approved to meet the requirements of the European Directive for equipment intended to be used in hazardous areas (commonly know as "ATEX 95“) and of the North American Standards (certified by the Canadian Standards Association, CSA International).
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION CAT 200 Connection of mains and interface signals are to be established via internal screw terminal blocks. This requires working inside an opened analyzer, near by potentially live components! Installation of this analyzer is allowed by qualified personnel only, familiar with the potential risks of working near live components! The model CAT 200 does not provide a mains switch.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 CAT 200 1 11 2 3 10 9 4 8 7 6 5 Fig. 1-28: CAT 200, Exterior view 1 2 3 4 5 6 1 - 30 Eye bolts (2 pcs) Cable glands for signal cables (3 pcs max) Eye bolts (2 pcs) Mounting holes (4 pcs.) Base Extender housing 7 8 9 10 11 Dome housing Front panel behind safety glass Interlock screw Interlock screw Junction box Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION CAT 200 6 5 4 3 1 2 Fig. 1-29: CAT 200 (Dome and extender housing removed) 1 2 3 4 Analyzer module Gas fittings with internal flame arrestors Bracket for analyzer module Nameplate label Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL GAS PATHS 1.9 Internal Gas Paths The materials used for the gas paths may be selcted to suit the intendend application. In marking such selection the diffusion rates of the individual gas components, their corrosivity, and the temperature and pressure of the sampled gas must be taken into account. 1.9.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 1.9.2 TECHNICAL DESCRIPTION INTERNAL GAS PATHS Gas Path Layout (internal tubing) The principle various possible layouts of the internal gas lines are summarized in the table 1-1.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 INTERNAL GAS PATHS MLT 3 (GAS PURITY MEASUREMENT) 1.9.3 MLT 3 (gas purity measurement) The necessary gas connections are marked analyzer specific (see section 5.3.and 5.4.3.1). Different possibilities of internal layout and external tubing are shown in Fig. 1-30, depending on instrument specific features. Up to the internal pressure regulator the gas paths are designed as stainless steel tubings.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION INTERNAL GAS PATHS MLT 3 (GAS PURITY MEASUREMENT) 6 mm 6 mm sample gas SS tubing 1/8" tube piece gas outlet quick shutoff connector pressure sensor 6 mm Viton tubing Viton tubing 6 mm Viton tubing SS tubing PO2 front side rear side Viton tubing analysis cell Viton tubing analysis cell span gas Viton tubing zero gas SS tubing saftey dust filter threated plug 1/8" 1/16" 6 mm 1/8" 1/8" 1/8" 6 mm 1/16"
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 PRINTED CIRCUIT BOARDS 1.10 Printed Circuit Boards All necessary PCBs are pushed in into a cardcage, which is identically for all MLT versions (see Fig. 1-31). MLT Rear Panel Cardcage Network Inlet Module (LEM) optional PCB (e.g. SIO / DIO) Interconnection Board (ICB) Physical Interface (PIC) or Digital Signal Processing (DSP) Signal Processing (PSV) (free if using DSP) Controller Board (ACU) MLT Front Side Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION PRINTED CIRCUIT BOARDS optional PCB [e.g. SIO / DIO] Signal Processing (PSV) (free if using DSP) Inter Connection Board (ICB) out Network Inlet Module Controller Board (ACU) (LEM, Option) Physical Interface (PIC) or Digital Signal Processing (DSP) Fig. 1-31b: PCB arrangement MLT 2 [Inside view, detail (without front panel)] 1.10.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 PHYSICS INTERFACE CARD (PIC) 1.10.4 PIC (Physics Interface Card) The PIC card (Physics Interface Card) supplies the photometer components and the individual sensors with the individual required operating voltages and transmits all measuring signals to the signal processing unit PSV. 1 2 3 Fig. 1-32: Plug pin assignment PCB PIC The plugs shown in Fig. 1-20 are used as follows: Plug No.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 1.10.5 TECHNICAL DESCRIPTION DIGITAL SIGNAL PROCESSING CARD (DSP) Digital Signal Processing Card (DSP) Instead of using the 2 PCB's PIC and PSV alternatively those can be replaced by ONE board containing both functions in the Digital Signal Processing Board DSP. NGA Bus connector Input DC sensors (e.g. O2/H2 sensors) Input AC sensors (e.g. IR/VIS/UV sensors) Temp. sensor 4/IR(UV) source 4 Temp. sensor 3/IR(UV) source 3 Temp.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 PCB ACU 1.10.6 ACU The ACU 02 board is the main part of the MLT both Analyzer Module or Analyzer! It is a one-board CPU card. All user data are safe loaded via a battery buffer, even when the voltage supply is disconnected. This card contains the complete software for controlling and operating the analyzer module.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 1.10.7 TECHNICAL DESCRIPTION SIO (STANDARD INPUTS-/OUTPUTS) SIO (Standard Inputs-/Outputs) Every platform, every MLT analyzer or every MLT analysis module can be equipped with max. 1 SIO (see table P-1, too)! At mounting into a platform or an analyzer the SIO serves system functionallity of I/O´s and supports all analysis modules of the NGA system. Only the analysis module itself is supported at mounting into an analysis module.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 Position of SIF module RS232 or RS485 Position of SIA module Analog output 7 + 8 Position of SIA module Analog output 3 + 4 Position of SIA module Analog output 5 + 6 DIO (DIGITAL IN-/OUTPUTS) Fig. 1-35: SIO-PCB with extension cards 1.10.8 DIO (Digital In-/Outputs) Every platform can be equipped with max. 1 to 5 DIOs. Every MLT analyzer or every MLT analysis module can be equipped with max.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 1.11 TECHNICAL DESCRIPTION NETWORK TERMINATION Network Termination The network module LEM (see Fig. 1-19) connects the analyzer with external modules via network. RJ 45 sockets are serving for network interconnection between analyzer or platform and analyzer modules (see Fig. 1-8, 1-13, 1-20, 1-23 and 1-31b). It is necessary to terminate the ends of a twisted pair bus to minimize reflections (see Fig. 1-37).
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 90002929 07/2006 SPECIFICATIONS AT THE NAMEPLATE LABEL 1.12 Specifications at the Nameplate Label The analyzer nameplate label gives important information about the instrument‘s configuration, installed measuring principle(s), sample gas(es) and measuring range(s). In case an instrument is built according a suitability approval the related certificate number is given, too.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DESCRIPTION SPECIFICATIONS AT THE NAMEPLATE LABEL All analyzer versions are marked as follows: MLTx y-CH1 CH2 CH3 CH4 CH5 with x = analyzer type 1, 2, 3, 4, 5, CAT 200 with 1 = 1/2 19", not thermostatted, external power supply 2 = Field housing, thermostatted, internal power supply 3 = 1/1 19"-housing, thermostatted, internal power supply 4 = 1/1 19"-housing, thermostatted, external power supply 5 = 19", 18 to 21 HU housing, thermostat
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DESCRIPTION 1.13 90002929 01/2007 MLT 2HT (high temperature measurement) As a special version MLT 2HT is available for measurements requiring physics thermostated up to 120 °C. The physical compartment is parted into 2 sections: - A high temperature (HT) part with special heating and isolation for all components being in contact with sample gas - SS tubing/fittings, fine dust/safety filter, needle valve, calibration valve (depending on options).
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 2. MEASURING PRINCIPLE IR MEASUREMENT Measuring Principle Depending on analyzer model different measuring methods will be used. 2.1 IR Measurement The measuring effect devided from absorption of infrared radiation is due to the gas being measured. The gas - specific wavelengths of the absorption bands characterize the type of gas while the strength of the absorption gives a measure of the concentration of the component measured.
NGA 2000 MLT Hardware Instruction Manual MEASURING PRINCIPLE IR MEASUREMENT 90002929 07/2006 to electronic 14 14 12 13 11 11 10 10 9 9 8 8 7 7 5 6 3 4 1 2 Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 2.1.1 MEASURING PRINCIPLE IR MEASUREMENT Opto - Pneumatic Measuring Principle For measurement a heating coil in the light source generates the necessary infrared radiation (1). This radiation passes through the chopper wheel [light chopper wheel (3)].
NGA 2000 MLT Hardware MEASURING PRINCIPLE IR MEASUREMENT Instruction Manual 90002929 07/2006 When the IR radiation passes through the measurement side of the analysis cell into the detector, a part of it is absorbed depending on gas concentration. Absorption chamber gas become colder, gas pressure in the absorption chamber is reduced and some gas of compensation chamber passes through the flow channel into the absorption chamber.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 MEASURING PRINCIPLE IR MEASUREMENT CO2 CO 75 60 45 54 Interference Filter 36 CO CO2 0 HC 18 Reference Transmittance [%] 72 90 0 15 30 Transmittance [%] 90 Absorption band 3000 3200 3400 3600 3800 4000 4200 4400 4600 4800 5000 5200 5400 5600 5800 6000 Wave Length [nm] Fig.
NGA 2000 MLT Hardware MEASURING PRINCIPLE UV MEASUREMENT 2.2 Instruction Manual 90002929 07/2006 UV Measurement The absorption measurement in the UV spectral range is based on the same principle as the IR measurement (Fig. 2-1). A glow-discharge lamp [2] is used as radiation source. The UV radiation passes through the chopper [3] and a filter cell [6] into the dual-section analysis cell [7]. A second filter cell [6] is installed after the analysis cell.
Instruction Manual 90002929 07/2006 2.3 NGA 2000 MLT Hardware MEASURING PRINCIPLE OXYGEN MEASUREMENT Oxygen Measurement Depending on analyzer model different two measuring methods will be used. 2.3.1 Paramagnetic Measurement (PO2) Oxygen measurement is based on the paramagnetical characteristics of Oxygen molecules: Two nitrogen filled quartz spheres (N2 is not paramagnetic) are arranged in a dumbbell configuration and, hinged to a platinum wire, placed inside a cell.
NGA 2000 MLT Hardware Instruction Manual MEASURING PRINCIPLE OXYGEN MEASUREMENT 90002929 07/2006 10 4 11 1 5 3 2 6 7 9 8 Fig. 2-4: Principle Construction of paramagnetic Analysis Cell 1 2 3 4 5 6 7 8 9 10 11 2-8 Permanentmagnet Platinum wire Mirror Glass ball Loop Photodetector Light source Preamplifier Display Gas inlet Gas outlet Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 MEASURING PRINCIPLE OXYGEN MEASUREMENT Solvent-resistant Paramagnetic Oxygen Cell Solvents being tested: up to 20% Name Name Name Name Name Acetone Acrolein Argon Aromatics Butadiene Butadiene-1 Butadiene-2 Methane Ethane Propane n-Butane Pentane Hexane Heptane Methyl bromide Iso propanol Acetic acid Ethanol Methanol Ethylene Ethylene oxide Acetylene Butene i Butyr acid i-Butyr aldehyde i-Propyl formiat Formaldehyde Carbon dioxide Dimethy
NGA 2000 MLT Hardware Instruction Manual MEASURING PRINCIPLE OXYGEN MEASUREMENT 90002929 07/2006 Special Hints on Paramagnetical Oxygen Sensors The table below shows how accompanying gases interfere the paramagnetic oxygen measurement. If the concentration of such gases is already given at time of enquiry this interference may be taken into account during factory startup and thus minimized (option).
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 2.3.2 MEASURING PRINCIPLE OXYGEN MEASUREMENT Electrochemical Measurement (EO2) This sensor uses the principle of galvanic cells, fig. 2-5 shows the design. 10 11 6 9 11 3 5 1 7 13 2 4 8 12 12 Fig. 2-5: Structure of electrochemical Oxygen Sensor The electrochemical oxygen sensor‘s key components are a lead anode (1) and a gold cathode (2) surrounded by a special acid electrolyte (3).
NGA 2000 MLT Hardware Instruction Manual MEASURING PRINCIPLE OXYGEN MEASUREMENT 90002929 07/2006 (Red) (Black) (11) Thermistor (5) Resistor (6) (-) (+) GoldCathode (2) LeadAnode (1) O2 + 4 H+ + 4 e- → 2 H2O 2 Pb + 2 H2O → 2 PbO + 4 H+ + 4 e- Electrolyte (3) (ph 6) Summary reaktion O2 + 2 Pb → 2 PbO Fig. 2-6: Reaction of galvanic cell In consequence of it‘s design the sensor‘s lifetime is limited and depends on theoretical designed life and oxygen concentration.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 2.3.3 MEASURING PRINCIPLE OXYGEN MEASUREMENT Trace Electrochemical Measurement (TEO2) The MLT uses an electrochemical sensor technology to achieve the trace measurement of oxygen. The principle structure of the oxygen sensor is shown in Fig. 2-7. The sensor is a self contained disposable unit which requires no maintenance.
NGA 2000 MLT Hardware MEASURING PRINCIPLE OXYGEN MEASUREMENT Instruction Manual 90002929 07/2006 Notes for analyzers with electrochemical TEO2 cell! For TEO2 sensor please note that the gas inlet and outlet connections of the analyzer are sealed to prevent exposure of the sensor to air. Prolonged exposure of the sensor to air can cause extended start up time, reduction of performance or damage to the sensor.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 2.4 MEASURING PRINCIPLE THERMAL CONDUCTIVITY Thermal Conductivity To measure gases like Hydrogen (H2), Argon (Ar) or Helium (He), the measurement method of thermal conductivity (TC) will be used. 2.4.1 Sensor Design A complete in glass encapulated temperature resistor is the basis of the measuring. Four of this resisitors are arranged in a Wheatstone Bridge which is mounted into a block.
NGA 2000 MLT Hardware Instruction Manual MEASURING PRINCIPLE THERMAL CONDUCTIVITY 2.4.3 90002929 07/2006 Measurement Method The entire measurement cell is thermostated to a temperature of up to 75 °C. The four integral temperature sensors are electrically heated to a higher temperature. Their temperatures, and thus their electrical resistance, are dependent upon heat losses, which, in turn, result from heat transport in the surrounding gas to colder chamber walls.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 5. INSTALLATION GENERAL Installation This chapter describes in detail how to install the different analyzer variations. 5.1 General Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing material until the instrument is operational.
NGA 2000 MLT Hardware INSTALLATION GENERAL Instruction Manual 90002929 07/2006 To stay in compliance with regulations regarding electromagnetic compatibility it is recommended to use only shielded cables, as optionally available from Emerson Process Management or equivalent. Customer has to take care that the shield is connected in proper way. Shield and signal connector enclosure need to be conductively connected, submin-d-plugs and sockets must be screwed on the analyzer.
Instruction Manual 90002929 07/2006 5.1.1 NGA 2000 MLT Hardware INSTALLATION TRANSFER SAFETY LOCK OF MLT 1/ULCO OR MLT 2 Transfer Safety Lock of MLT 1/ULCO or MLT 2 TRANSFER SAFETY LOCK ! First of all operations unscrew transfer safety lock of MLT 1/ULCO or MLT 2 ! MLT 1/ULCO: Unscrew both knurled-head screws on bottom side of the housing (Fig. 5-1a) ! For protection against loss screw the knurled-head screws into the respective holders at housing rear side (Fig.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 GAS CONDITIONING (SAMPLE HANDLING) 5.
Instruction Manual 90002929 07/2006 5.2.1 NGA 2000 MLT Hardware INSTALLATION GAS CONDITIONING (SAMPLE HANDLING) Fine Dust Filter (Option MLT 3) MLT 3 can have two built-in fine dust filters (filter material PTFE, pore-size approx. 2 µm) optional. 5.2.2 Gas Sampling Pump (Option MLT 3) Optional MLT 3 can be equipped with a gas sampling pump (pumping rate max. 2.5 l/min.). Note! For mobile application of MLT only ! Lifetime max. 5,000 running hours ! 5.2.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 GAS CONNECTIONS 5.3 Gas Connections TOXIC GAS HAZARDS! Take care that all external gas lines are connected as described and are tight to avoid leaks! Improperly connected gas lines may cause explosion or death! Exhaust may contain hydrocarbon and other toxic components (e.g. carbon monoxide)! Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 INSTALLATION GAS CONNECTIONS The number of gas fittings as well as their assignment vary depending on analyzer model and selected options. All gas fittings are labeled and are normally located at the analyzers rear panel, in case of MLT 2 and CAT 100 they are located at the bottom side. When it is necessary to open gas paths seal the analyzer‘s gas fittings by using PVC caps to avoid pollution of the internal gas path by moisture, dust, etc.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 GAS CONNECTIONS 5.3.1 Internal Solenoid Valve Block (Option for MLT 1/CAT 200) For operation with optional solenoid valves, the following indications have to be considered: Operation with solenoid valves is not possible for 2 - channel analyzer with parallel gas paths. Supply overpressure for all gases is limited to 50 to 500 hPa when an analyzer is equipped with an internal solenoid valve block. . 5.3.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 1 5.4 Analyzer Specific Instructions 5.4.1 MLT 1 5.4.1.1 MLT 1, platform mounting only MLT 1 analyzer module housing is available for platform mounting (built-in into a NGA platform) only (type M). These analyzers need DC 24 V power supply and netwotk connection utilizing platform. Built-in the modul into platform housing correctly (see platform manual).
NGA 2000 MLT Hardware INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 1 5.4.1.2 Instruction Manual 90002929 07/2006 MLT 1, external installation and 1/2-19" Housing MLT 1 analyzer module housing is available for platform mounting (built-in into a NGA platform) or for external installation (type A). The analyzer variation MLT 1 configured as tabletop analyzer (type T) provide a handle attached to a front panel frame.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 1 Network connection (RJ 45 socket) [not for platform mounting module] optional PCBs [SIO / DIO e. g.] Gas connections Input 24 V dc [not for platform mounting module] Fig.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 ANALYZER SPECIFIC INSTRUCTIONS: MLT 1 Solenoid valve block Fig. 5-5: MLT 1, gas connections with solenoid valve option V 1 = Span gas 1 V 2 = Span gas 2 V 3 = Sample gas V 4 = Zero gas out = Gas outlet (to IN CH1)* IN CH1 = Gas input (from OUT of soelnoid valve block)* OUT CH1 = Gas exhaust * Standard.
Instruction Manual 90002929 07/2006 5.4.2 NGA 2000 MLT Hardware INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 2 MLT 2 (Field Housing) The MLT 2 variation is designed for wall mounting and is available in two different versions: One version, equipped with standard front panel is intended to be installed at protected locations whereas the version with impact tested magnetically operated front panel may be outdoor installed.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 ANALYZER SPECIFIC INSTRUCTIONS: MLT 2 550 492 300 332 approx. 355 10 18 Glands (connection cables) 18 Fig. 5-6a: Dimensional sketch / Drill drawing MLT 2 Standard version [all dimensions in mm] 550 492 approx. 355 603 635 300 38 38 38 10 glands (connection cables) 27 18 43 18 280 16 Gas connections Fig.
Instruction Manual 90002929 07/2006 5.4.2.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 ANALYZER SPECIFIC INSTRUCTIONS: MLT 2 a) Mains Supply The analyzer provides an integrated multi range power supply unit (either power supply of type SL5 or of type SL10) with voltage selector switch, to be set to the appropriate range to meet the nominal voltage (115 V / 230 V) at installation site. The switch is accessible through a safety barrier when the analyzer‘s front door is open.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 2 Cables for external data processing must be double insulated for mains voltage when used inside the instrument! If double insulation not possible signal cables inside the analyzer must be installed in a way that a distance of at least 5 mm is ensured permanently (e.g. by utilizing cable ties). The number of installed adaptor elements variies depending on analyzer configuration.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 GAS CONNECTIONS ANALYZER SPECIFIC INSTRUCTIONS: MLT 2 Alternatively terminal strips (connection of mains line) Powerline filter (connection of L and N of mains line) Analog out Adaptor element option SIO: analog outputs PE N Digital out L N L PE Adaptor element Option DIO: digital inputs/outputs Ground conductor pin (PE mains line) Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 c) INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 2 Cable Gland Assembly Instruction for Shielded Cables 1. Strip the cable insulation. 2. Uncover the shielding 3. Feed cable through gland nut and into fixing element. 4. Put the shielding net over the element the way that it covers the o-ring 2 mm. 5. Stick the fixing element into the neck and fix the gland. Fig.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 ANALYZER SPECIFIC INSTRUCTIONS: MLT 2 Gas Connections Wall mounting holder d) Gas connections Fig. 5-12a: MLT 2, gas connections 1 2 3 4 Fig. 5-12b: MLT 2, Gas fittings at analyzer’s bottom side (other configurations possible, see labels at instrument) 1 2 3 4 5 - 20 Gas inlet channel 1 Gas inlet channel 2 Gas outlet channel 1 Gas outlet channel 2 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 5.4.3 INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 3 MLT 3 All components of analyzers or analyzer modules are incorporated into a 1/1 19" housing. The housings are available as rack mounting (R) or as table-top (T) versions. For analyzer modules there is mounted a blind plate instead of an operation front panel. The equipment has an internal power supply with “autoranging” for operating voltages of 230 V AC or 120 V AC resp., 47-63 Hz.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 ANALYZER SPECIFIC INSTRUCTIONS: MLT 3 5.4.3.1 MLT 3 for gas purity measurement The necessary gas connections are marked analyzer specific. Different possibilities of internal layout and external tubing are shown in Fig. 1-24, depending on instrument specific features.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 b) INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 3 Instrument with solenoid valve block The connection of sample gas, zero gas and span gas will be done to the solenoid valve block at the rear side of the instrument (Fig. 5-14 and 5-15). The common exhaust of solenoid valve block is be lead to the real sample gas inlet of the instrument via stainless steel tube (see Fig. 1-30a). span gas sample gas exhaust (common) zero gas Fig.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 ANALYZER SPECIFIC INSTRUCTIONS: MLT 4 5.4.4 MLT 4 All components of analyzers or analyzer modules are incorporated into a 1/1 19" housing. The housings are available as rack mounting (R) or as table-top (T) versions. For analyzer modules there is mounted a blind plate instead of an operation front panel. These analyzers need DC 24 V power supply utilizing an external power supply unit.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: MLT 4 1 IN OUT 3/Purge 3/Purge IN OUT 2 IN OUT Fig. 5-18: MLT 4, gas connections in = Gas inlet out = Gas outlet 1 = measuring channel 1 2 = measuring channel 2 3 = measuring channel 3 Purge = purge gas (housing) Note! Purge option combination is only available with one or two channel analyzer. Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: CAT 200 5.4.5 Instruction Manual 90002929 07/2006 CAT 200 CAT 100 analyzers are designed for wall mounting and intended to be outdoor installed taking into account the ambient temperature limits as given in the technical data section of this manual. .
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 5.4.5.1 INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: CAT 200 Wall Mounting Mount CAT 200 analyzers to a wall or stand by utilizing 4 through holes of ø 15.9 mm (0.63 in) each.
NGA 2000 MLT Hardware INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: CAT 200 5.4.5.
Instruction Manual 90002929 07/2006 NGA 2000 MLT Hardware INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: CAT 200 ❏ This analyzer model is supplied by an internatl wide range power supply, automatically adapting to the voltage at site. ❏ All cables (mains power cord and signal cables) are connected via screw terminals located inside an junction box right above the analyzer dome. ❏ See the label on the junction box cover for detailed terminal assignments.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 ANALYZER SPECIFIC INSTRUCTIONS: CAT 200 1 1 24 5 1 1 2 3 4 24 5 6 7 8 Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 5.4.5.3 INSTALLATION ANALYZER SPECIFIC INSTRUCTIONS: CAT 200 Gas Connections 1 2 3 5 8 6 4 7 9 10 11 Fig.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION WIRING SIGNAL TERMINALS: ELECTRICAL CONNECTIONS IN GENERAL 5.5 90002929 07/2006 Wiring Signal Terminals Emerson Process Managament has taken any effort during the MLT series development process to ensure electromagnetically compatibility (EMC; concerning emission and immunity), stated by EMC measurements according EN 61326. Nevertheless EMC is not only influenced by the instrument‘s design, but widely by the installation procedure at site, too.
Instruction Manual 90002929 07/2006 ❏ NGA 2000 MLT Hardware INSTALLATION WIRING SIGNAL TERMINALS: ELECTRICAL CONNECTIONS IN GENERAL Using double-shielded cables: In this case one shield is connected to the analyzer‘s housing while the other shield is connected to the external equipment. This gives an advantage when both instruments are supplied by different supply networks (e.g. when installed in different buildings).
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 WIRING SIGNAL TERMINALS: WIRING INDUCTIVE LOADS 5.5.2 Wiring Inductive Loads Switching inductive loads is a standard application generating electromagnetic disturbances: The moment an inductive load (e.g. relay, valve, etc.) is switched off, it‘s magnetic field defies the change of current flow, generating high voltages (up to hundrets of volts) at the coil‘s contacts.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 5.5.3 INSTALLATION WIRING SIGNAL TERMINALS: DRIVING MULTIPLE LOADS Driving Multiple Loads Another popular application is driving multiple loads within one system by multiple outputs, whereat the supply voltage for the loads is taken from one common source.
NGA 2000 MLT Hardware Instruction Manual INSTALLATION 90002929 07/2006 WIRING SIGNAL TERMINALS: DRIVING HIGH CURRENT LOADS 5.5.4 Driving High Current Loads Loads with currents exceeding the rated currents specified for MLT analyzers outputs must not be driven directly by digital or relay outputs. Driving such loads requires external relays acting as decoupling devices: The MLT output drives the external relay, which itself drives the load.
Instruction Manual 90002929 07/2006 6. NGA 2000 MLT Hardware SWITCHING ON Switching On Take care of the safety instructions as given at the beginning of this manual while working at and inside the instruments! Once the instrument has been correctly assembled and installed in accordance with the general instructions given in section 5., the equipment is ready for operation. The equipment is switched on by providing the required voltage.
NGA 2000 MLT Hardware SWITCHING ON 6-2 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual 90002929 07/2005 NGA 2000 MLT Hardware MEASUREMENT/CALIBRATION//SWITCHING OFF MEASUREMENT 7. Measurement / Calibration / Switching Off 7.1 Measurement The primary step in the measurement of the concentration of a gas component is the admission of sample gas to the analyzer. Note! Take care for the start-up procedure before measurement (chapter 6.)! ❍ Admit sample gas at the respective gas inlet fitting (see Item 5.).
NGA 2000 MLT Hardware MEASUREMENT/CALIBRATION//SWITCHING OFF CALIBRATION 7.2 Instruction Manual 90002929 07/2005 Calibration To insure correct measurement results, zeroing and spanning should be carried out once a week. Other intervals are possible. The zero-level must always first be set before any other calibrations are attempted.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2005 MEASUREMENT/CALIBRATION//SWITCHING OFF CALIBRATION 7.2.1 Test Gases 7.2.1.1 Zero Gas For zeroing, the analyzer has to be purged with nitrogen (N2) or adequate zero gas [e. g. synth. air or conditionned air (but not for O2 - measurement)] . 7.2.1.2 Span Gas The calibration of all another analyzers should be done with pure span gases in order to prevent interferences between the gases (e. g.
NGA 2000 MLT Hardware MEASUREMENT/CALIBRATION//SWITCHING OFF SWITCHING OFF 7.3 Instruction Manual 90002929 07/2005 Switching Off Before switching off the analyzer, we recommend first purging all the gas lines for about 5 minutes with zeroing gas (N2) or adequate conditioned air.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 10. TROUBLESHOOTING NO FUNCTION (LCD DISPLAY) Troubleshooting Take care of the safety instructions as given at the beginning of this manual while working at and inside the instruments! Working inside the instruments is subject to briefed personnel familiar with potential risks! 10.
NGA 2000 MLT Hardware Instruction Manual TROUBLESHOOTING NO MEASUREMENT SCREEN 90002929 07/2006 10.2 No / Incorrect Measurement Screen Possible Causes a) Defective network connection: Check / Correct Check network termination (see section 1.11). Check network connection between platform, analyzer, external modules. Replace connection cable or network module PCB “LEM” if necessary.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 10.3 TROUBLESHOOTING DISPLAY MESSAGES Display Messages Several operating conditions of the analyzer / module are controlled and signalized via corresponding messages of the display. Failures can be shown at the menu “Status” → “Status details” → “Failures” (Item “4.1.1 Status details” of software manual). 10.3.1 Chopper Fail Possible Causes a) Chopper fail: Check / Correct Check connection to PCB “PIC” Fig. 1-32) or PCB “DSP” Fig.
NGA 2000 MLT Hardware Instruction Manual TROUBLESHOOTING DISPLAY MESSAGES 10.3.3 Detector signal communication failed Possible Causes a) 90002929 07/2006 Check / Correct Failure in signal processing: Check connection to PCB “PIC” Fig. 1-32) or Detector failure PCB “DSP” Fig. 1-33) resp.; See Section 11.1.4 (IR/UV measurement), Section 11.1.6 (O2 measurement). b) Contamination of the gas paths Check if photometric components are dirty, clean or replace the components if necessary (cf.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 10.3.6 TROUBLESHOOTING DISPLAY MESSAGES Heater Unit Regulation Possible Causes Check / Correct a) PCB “BHZ” failure Check PCB “BHZ” (Section 11.2) b) Voltage supply absent Check voltage supply (Section 10.1 / 11.1.1) 10.3.7 Temperature Measurement Possible Causes a) b) Check / Correct Chopper not or not correct connected Check connection of Chopper → PCB “PIC” to PCB “PIC”/“DSP” (Fig. 1-20) or Chopper → PCB “DSP” (Fig.
NGA 2000 MLT Hardware Instruction Manual TROUBLESHOOTING DISPLAY MESSAGES - NO OR INCORRECT ANALOG OUTPUTS / DIGITAL I/O'S 10.3.9 90002929 07/2006 External Input Possible Causes Check / Correct a) Configuration of PCB “DIO” incorrect Check correct software configuration of “DIO” b) EMC radiation via PCB “DIO” Incorrect wiring of external actors (section 5.5) c) PCB “DIO” not or not correct connected / Check correct connection of PCB “DIO”. defective Exchange PCB “DIO”. d) 10.
Instruction Manual 90002929 07/2006 10.5 TROUBLESHOOTING CALIBRATION NOT POSSIBLE Calibration not possible Possible Causes a) NGA 2000 MLT Hardware Incorrect nominal value of span gas Check / Correct Enter the correct nominal value (certification of span gas bottle) (see software manual) b) Incorrect zero gas / span gas in use. Check zero gas / test gas in use c) Tolerance error.
NGA 2000 MLT Hardware Instruction Manual TROUBLESHOOTING FLUCTUATING OR ERRONEOUS DISPLAY 10.6 90002929 07/2006 Fluctuating or erroneous display Possible Causes Check / Correct a) Leakage into gas circuit. Perform a leakage check (see Section 14). b) Ambient air contains gas constituent to be measured in excessive concentration. Purge the instrument. c) Gas pressure subject to excessive fluctuations. Check the gas lines preceding and following the instrument.
Instruction Manual 90002929 07/2006 10.7 TROUBLESHOOTING RESPONSE TIME TOO LONG (T90 TIME) Response time too long (t90 time) Possible Causes a) NGA 2000 MLT Hardware Incorrect response time (t90 time). Check / Correct Check the value for t90 time (see software manual) b) Pumping rate inadequate. Check diaphragms and membranes of gas sampling pump. The feeder line between the sampling point and the instrument is too long.
NGA 2000 MLT Hardware TROUBLESHOOTING 10 - 10 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TEST PROCEDURE / TEST POINTS SIGNAL PROCESSING WITH PSV 11. Test Procedure / Test Points The housing has to be opened (cf. Section 15) for troubleshooting, for checking the connections and for replacement or cleaning of any of the components of the equipment.
NGA 2000 MLT Hardware Instruction Manual TEST PROCEDURE / TEST POINTS INTERNAL VOLTAGE SUPPLY / IR SOURCE 11.1.1 90002929 07/2006 Internal Voltage Supply Test point: “+” “-” “A” “ ” “+” or “-” of detector PCB “VVS” (Fig. 11-2) R5 (UV) Measuring instrument: Digital voltmeter Signal: ≈ + 6,2 V dc at test point “+” ≈ - 6,5 V dc at test point “-” R11 (IR) R1 (UV) Failure: No signal Possible reason: Detector not connected to PCB “PIC” (Fig. 1-20) or PCB “DSP” (Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TEST PROCEDURE / TEST POINTS CHOPPER / UNAMPLIFIED MEASURING SIGNAL AT DETECTOR LED “red” 11.1.3 Chopper Test point: PCB “MOP” Measuring instrument: optical LED “green” Fig. 11-3: PCB "MOP" (partial view) Signal: red LED must be off green LED must be blinking Failure: red LED is on / green LED is on permanent Possible reason: Chopper not (correct) connected to PCB “PIC” (Fig.
NGA 2000 MLT Hardware Instruction Manual TEST PROCEDURE / TEST POINTS SIGNAL PROCESSING ON PCB “PSV” 11.1.5 90002929 07/2006 Signal Processing on PCB “PSV” Test point: P21.2 (“+”) and P21.3 “ ” of PCB “PIC” (Fig. 1-32) Measuring instrument: Oscilloscope (signal form see 11.1.4) Signal: Unamplified Measuring Signal from Detector (Measuring Channel 1) (max. 4 VPP / min. 1 VPP) green LED of PCB “PSV” must be blinking with approx.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 11.1.6 TEST PROCEDURE / TEST POINTS PHYSICAL ZERO ALIGNMENT Physical Zero Alignment For physical zero adjustment the raw signal of respective measuring channel is set to ± 100.000 counts while purging instrument with zero gas. The raw measuring values can be shown by pressing "Status" (F2) → "RawMeas." (F2). 11.1.6.1 IR Measurement For set of physical zero adjustment see Item 17.5 please. 11.1.6.
NGA 2000 MLT Hardware Instruction Manual TEST PROCEDURE / TEST POINTS HEATING UNIT 11.2 90002929 07/2006 Heating Unit Test point: PCB “BHZ” (Fig. 11.4) of heater unit (Fig. 1-21.
Instruction Manual 90002929 07/2006 11.3 NGA 2000 MLT Hardware TEST PROCEDURE / TEST POINTS TROUBLESHOOTING INSTRUCTIONS FOR PCB DSP01 Troubleshooting Instructions for PCB DSP01 Fig. 11-5: PCB "DSP01” This chapter contains purposeful instructions for troubleshooting when operating a DSP01 pcb (substitute for PIC02-pcb and PSV02-pcb, e.g. when raw count = 0 counts). This instructions manual provides a troubleshooting procedure that enables you to check DSP01 for errors step by step.
NGA 2000 MLT Hardware Instruction Manual TEST PROCEDURE / TEST POINTS TROUBLESHOOTING INSTRUCTIONS FOR PCB DSP01 11.3.1 90002929 07/2006 Inspection of LEDs on component side of DSP01 A C B Fig. 11-6: LEDs on component side of PCB "DSP01” If you presume any malfunction of DSP01 pcb, first step is to check out correct activity / status of the LEDs on component side of the pcb. In the picture above, all elements that have to be inspected are marked in yellow. .
Instruction Manual 90002929 07/2006 NGA 2000 MLT Hardware TEST PROCEDURE / TEST POINTS TROUBLESHOOTING INSTRUCTIONS FOR PCB DSP01 B On the right you see the LEDs fo area B in enlarged depiction. During operation, the LEDs V3 (CH1) and V4 (CH2) must be flashing, if a detector is employed on the concerning detector connection. In case that the appropriate channel is not B used, status of the concerning LEDs does not matter.
NGA 2000 MLT Hardware Instruction Manual TEST PROCEDURE / TEST POINTS TROUBLESHOOTING INSTRUCTIONS FOR PCB DSP01 11.3.2 90002929 07/2006 Inspection of soldering pads of DSP01 For that part of troubleshooting, and especially for inspection of soldering pads LB1 and LB2, turn the soldering side of DSP01 up. A B Fig. 11-7: Soldering pads on soldering side of PCB "DSP01” Areas A and B where soldering pads LB1, LB2 and LB18 are located in, are marked in yellow in the depiction above.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 11.3.3 TEST PROCEDURE / TEST POINTS TROUBLESHOOTING INSTRUCTIONS FOR PCB DSP01 Evaluation When inspection of the DSP is finished you can evaluate it as follows: (Schematic diagram of inspection) Inspection of the LEDs (acc. to part 1 of this description) as well as LB3 Status of LEDs not erroneus (acc. to description) Status of LEDs all right / LB3 all right (acc. to description) Inspection of soldering pads LB1, LB2, LB18 and P40 (acc.
NGA 2000 MLT Hardware Instruction Manual TEST PROCEDURE / TEST POINTS TROUBLESHOOTING INSTRUCTIONS FOR PCB DSP01 90002929 07/2006 11.3.4 Appendix 11.3.4.1 Description of soldering pad DEF (LB3): Basically, soldering pad LB3 / DEF must be open. Configuration of that soldering pad causes the following operating modes: open: „NORMAL-MODE“ closed: „DEFAULT-MODE“. Soldering pad C Fig. 11-9: Soldering pad “DEF/LB3” (PCB "DSP01”) 11.3.4.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TEST PROCEDURE / TEST POINTS TROUBLESHOOTING INSTRUCTIONS FOR PCB DSP01 Mode LB3 LB18 P40 Normal open 2-3 (flash enable 1-2 (for programmupdate), must be 2-3) open LED OK Comment Error V16 off always on or off V13 flash T / P Sensors (Pressure- and Temp.
NGA 2000 MLT Hardware TEST PROCEDURE / TEST POINTS 11 - 14 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual 90002929 07/2006 12.
NGA 2000 MLT Hardware Instruction Manual REMOVAL / REPLACEMENT OF COMPONENTS REMOVAL / REPLACEMENT OF OPERATION FRONT PANEL 12.2 90002929 07/2006 Removal / Replacement of Operation Front Panel Front panel plate together with LC display and circuit board AFP 01 constitutes an unit. Therefore, the front panel has to be disassembled completely when one component of the unit is out of order.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 12.3 REMOVAL / REPLACEMENT OF COMPONENTS REPLACEMENT OF BUFFER BATTERY OF ACU 02 Replacement of Buffer Battery on the ACU 02 Take care of the safety instructions as given at the beginning of this manual while working at and inside the instruments! Working inside the instruments is subject to briefed personnel familiar with potential risks! 12.3.1 Removal of ACU 02 Use the following steps to remove the ACU 02 board (see Fig.
NGA 2000 MLT Hardware Instruction Manual REMOVAL / REPLACEMENT OF COMPONENTS REPLACEMENT OF BUFFER BATTERY OF ACU 02 12.3.2 90002929 07/2006 Replacement of Buffer Battery Use the following procedure to replace the battery: ❍ Remove the ACU 02 (see 12.3.1). ❍ Remove the jumper “P23” (ACU02) or “P17” (ACU 01) for battery buffer (Fig. 12-1). All data and compensation values entered by the user will be deleted (RAM-fail)! ❍ Unsolder the battery from the soldering pins (see Fig. 12-1).
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 12.4 REMOVAL / REPLACEMENT OF COMPONENTS FUSES Fuses Take care of the safety instructions as given at the beginning of this manual while working at and inside the instruments! Working inside the instruments is subject to briefed personnel familiar with potential risks! In case of exchanging fuses the customer has to be certain that fuses of specified type and rated current are used.
NGA 2000 MLT Hardware Instruction Manual REMOVAL / REPLACEMENT OF COMPONENTS FUSES 12.4.2 90002929 07/2006 MLT 1 / 4 The fuses for 24 V dc Input are mounted on the PCB “LEM”. ❍ Remove PCB “LEM” (see Section 12.1). ❍ Take out and check the fuses (Fig. 12-3). Replace the fuse(s) if necessary. Two pieces of spare fuses are on the PCB (Fig. 12-3). ❍ Insert PCB “LEM” (see Section 12.1) and reclose the instrument. Network Connectors (RJ 45) Spare Fuses Fuse Fig.
Instruction Manual 90002929 07/2006 12.5 NGA 2000 MLT Hardware REMOVAL / REPLACEMENT OF COMPONENTS CONNECT/DISCONNECT UV SOURCE AND POWER SUPPLY FOR UV SOURCE Connect / Disconnect UV Source and Power Supply for UV Source The cover of the housing and the front panel plate has to be disassembled to have access to the cable from the UV source to connect it with the power supply for UV source.
NGA 2000 MLT Hardware Instruction Manual REMOVAL / REPLACEMENT OF COMPONENTS CONNECT/DISCONNECT UV SOURCE AND POWER SUPPLY FOR UV SOURCE 90002929 07/2006 ❍ Open the front panel / remove cover from MLT housing (cf. Section 15). ❍ Take the cable from the UV source (part no. 42711213) out of the thermostat controlled compartment to the power supply for UV source located behind the removed front panel ❍ Plug in the cable from UV source on the connector on top of the power supply for UV source (part no.
Instruction Manual 90002929 07/2006 NGA 2000 MLT Hardware REMOVAL / REPLACEMENT OF COMPONENTS CONNECT/DISCONNECT UV SOURCE AND POWER SUPPLY FOR UV SOURCE Connect the two white grounding cables to the UV source. ❍ One grounding cable comes from power supply for UV source . ❍ The 2nd grounding cables comes from PCB ESP10 (part no. 43002530). ❍ Connect both cables to the UV lamp (connection point) as indicated below.
NGA 2000 MLT Hardware REMOVAL / REPLACEMENT OF COMPONENTS 12 - 10 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 MAINTENANCE Maintenance In general only the gas conditionning hardware will require maintenance; the analyzer itself requires very little maintenance. The following checks are recommended for maintenance of the proper operation of the analyzer.
NGA 2000 MLT Hardware MAINTENANCE 13 - 2 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 14. LEAK TESTING Leak Testing The gas path system should be leak tested at least on a bimonthly basis and after maintenance, replacement or repair of gas path parts. Before opening gas paths they must be purged with ambient air or neutral gas (N2) to avoid hazards caused by toxic, flammable, explosive or harmful to health sample gas components! Analyzer / Analyzer Module Overpressure Valve Water Fig.
NGA 2000 MLT Hardware LEAK TESTING 14 - 2 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 15. OPENING OF HOUSING MLT 1 Opening the Housing The housing must be opened for checking the electrical connections and for replacement or cleaning of any of the components of the analyzer. The following sections describe how to open the different analyzer housings provided the instruments are fully accessible, e.g. are taken out of the rack.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 MLT 1 (PLATFORM HOUSING) 15.1 MLT 1 15.1.1 MLT 1 (Platform housing) 15.1.1.1 Housing Cover ❍ Disconnect all voltage supplies. ❍ Remove analyzer module out of platform (see platform manual). ❍ Unscrew the respective fastening screws at both housing sides (Fig. 15-1). ❍ For rear cover unscrew the additonal fastening screw at the top of the housing (Fig. 15-1) ❍ Remove the respective housing top cover panel.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 15.1.1.2 OPENING OF HOUSING MLT 1 (PLATFORM HOUSING) Front Panel Removing of front panel can be done without removing of housing covers: ❍ Disconnect module from power supply. ❍ Remove analyzer module out of platform (Item 17.1 of platform manual). ❍ Unscrew the 6 fastening screws at both housing sides (Fig. 15-2) or unscrew the 4 fastening screws at front panel (Fig. 15-2). ❍ Remove front panel to the front carefully.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 MLT 1 (1/2 19" HOUSING) 15.1.2 MLT 1 (1/2 19" housing) 15.1.2.1 Housing Cover ❍ Disconnect all voltage supplies. ❍ Unscrew fastening screws for rack mounting / front frame if necessary (Fig. 1-1). Remove analyzer out of rack or remove the front mounting frame and carrying strap to rear. ❍ Unscrew the respective fastening screws at both housing sides (Fig. 15-3) ❍ Remove the respective housing top cover panel.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 15.1.2.2 OPENING OF HOUSING MLT 1 (1/2 19" HOUSING) Front Panel ❍ Opening housing cover (Item 15.2.1). ❍ Unscrew the 6 fastening screws at both housing sides (Fig. 15-4) ❍ Remove front panel to the front carefully. Optionally internal components may be fixed to the front panel, e.g. internal fan, UVS, oxygen sensor, pressure sensor (see fig.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 MLT 2 15.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 OPENING OF HOUSING MLT 2 Fastener for front panel Fastener for front panel Fig. 15-5: MLT 2 (Field housing) (fastener for front panel) mechanical stop knurled head screws Fig. 15-6: MLT 2, photometer carriage slide Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 MLT 3 / MLT 4 15.3 ❍ MLT 3/4 (1/1 19" housing) Disconnect all voltage supplies. 15.3.1 ❍ Housing Cover Unscrew the 8 fastening screws on top. Pull the cover straight up. Repeat above steps in reverse order to assemble the housing 15.3.2 ❍ Front Panel (MLT 4 / MLT 3 standard version) Unscrew the 6 fastening screws (Fig. 15-7) Remove front panel to the front carefully.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 OPENING OF HOUSING MLT 3 FOR GAS PURITY MEASUREMENT 15.3.3 Front Panel (MLT 3 gas purity measurement) 15.3.3.1 Operation Front Panel ❍ Unscrew the 6 fastening screws (Fig. 15-8) Remove operation front panel to the front carefully. 15.3.3.2 ❍ Left Front Panel Unscrew the 4 fastening screws (Fig. 15-8). Remove left front panel to the front carefully. Gas lines and optionally internal components may be fixed to the front panel (see fig.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 CAT 200 15.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 OPENING OF HOUSING CAT 200 The CAT 200 housing consists of two main parts: • junction box • dome with extender housing Base O-Ring Extender Housing O-Ring Retaining Ring Window Dome Housing Fig. 15-9: Cat 200 Enclosure Assembly The junction box contains an encapsulated EMI mains filter and all the terminals for connecting mains supply and signal lines. The dome covers all other components, e.g. power supply unit and analyzer module.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 CAT 200 15.4.1 Junction Box ❍ Open the junction box by loosening the 6 Allen screws (fig. 15-10) utilizing the (imperial) Allen key shipped with the instrument. ❍ Remove the cover. Note! Assemble the cover utilizing ALL 6 screws to ensure proper IP rating against dust and water! Allen screws (6 pcs) Fig. 15-10: Cat 200 Junction box 15 - 12 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 15.4.2 OPENING OF HOUSING CAT 200 Dome To get access to internal components and the analyzer module the dome‘s components have to be removed (fig. 15-11): ❍ Loosen the outer screw ❍ Turn the dome counterclockwise to remove it ❍ Loosen the inner screw ❍ Turn the extender ring counterclockwise to remove it Screws fixing dome components Fig.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 CAT 200 right bracket analyzer module plate with detents first middle knurled head screw Fig. 15-12: View at analyzer module 3 detents Fig. 15-13: Plate with detents ❍ It‘s now possible to pivot the module downwards and to unscrew three additional knurled head screws to remove the module cover.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 OPENING OF HOUSING CAT 200 Repeat the steps in reverse order to assemble the housing. When having assembled the dome perform a first test to ensure the front panel sensor keys can be activated. If all sensor fields are accessible use the screws on top of the dome and extender ring to fix both units (fig. 15-11). If the sensor fields not accessible, see at section 15.4.3 to adjust the analyzer with magnetically operated front panel. 1 2 3 Fig.
NGA 2000 MLT Hardware Instruction Manual OPENING OF HOUSING 90002929 07/2006 CAT 200 15.4.3 Magnetically Operated Front Panel To ensure safe and faultless operation of the touch panel it is required to have the control elements (reed relays) installed within a defined distance from the inner side of the security glass. If it is necessary to remove the CAT 100 dome for maintenance and/or inspection purposes and to loosen the internal analyzer, the following measures must be followed: 15.4.3.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 OPENING OF HOUSING CAT 200 Front glass Dome Analyzer Key pad bar max. 2 mm Fig. 15-15: Distance between glass and key pad bar Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware OPENING OF HOUSING 15 - 18 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 16. FINE DUST FILTER (MLT 3 OPTION) Fine Dust filter (MLT 3 Option) The filter element of the in MLT 3 optional installed fine dust filter (see Fig. 1-16a) needs to be checked for pollution on a regular basis, depending on the application and to be replaced if need be. If the filter element shows contamination, the element should be immediately replaced with a new unit.
NGA 2000 MLT Hardware FINE DUST FILTER (MLT 3 OPTION) 16 - 2 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 17. REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS GENERAL Replacement and Cleaning of Photometric Components The housing has to be opened for checking the electrical connections and for replacement or cleaning of any of the components of the equipment.
NGA 2000 MLT Hardware Instruction Manual REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS REMOVAL OF THE PHOTOMETER ASSEMBLY 90002929 07/2006 Take care of the safety instructions as given at the beginning of this manual while working at and inside the instruments! Working inside the instruments is subject to briefed personnel familiar with potential risks! 17.1 Removal of the Photometer Assembly ❍ Open the housing (cf. Section 15).
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 17.2 REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS LIGHT SOURCE REPLACEMENT Light Source Replacement (IR) ❍ Open the housing (cf. Section 15). ❍ Remove the photometer assembly out of analyzer housing (see Section 17.1). ❍ Remove the two light source hexagonal mounting screws (shown in Fig. 17-2 as Item 1). ❍ Remove the light source together with its mounting flange.
NGA 2000 MLT Hardware Instruction Manual REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS REMOVAL OF ANALYSIS CELLS 17.3 Cleaning of Analysis Cells and Windows 17.3.1 Removal of Analysis Cells 90002929 07/2006 ❍ Open the housing (cf. Section 15). ❍ Remove the photometer assembly out of analyzer housing (see Section 17.1). a) For analysis cells of lengths 1 mm to 10 mm ❍ ❍ Remove the clamp (Fig. 17-3, Item 1). Remove the clamping collars and the filter cell with signal detector assembly.
Instruction Manual 90002929 07/2006 17.3.2 a) NGA 2000 MLT Hardware REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS CLEANING Cleaning Windows The shielding windows (on the filter cells, chopper housing and the analysis cell) may be cleaned with a soft, fluff free cloth. Use a highly volatile alcohol for the cleaning procedure. To remove any lint and dust particles remaining, blow off the cleaned components with nitrogen (N2).
NGA 2000 MLT Hardware Instruction Manual REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS REINSTALLING OF ANALYSIS CELLS 17.3.3 a) b) 90002929 07/2006 Reinstalling of Analysis Cells For analysis cells of lengths 1 mm and 7 mm ❍ Place the O - rings on the filter cells. ❍ Fit the components together and fix with the clamping collars. ❍ Install the clamp (Fig. 17-3, Item 1) and tighten. Analysis cells of lengths 50 mm - 200 mm: ❍ Place the O - ring on the chopper housing side of the cell body.
Instruction Manual 90002929 07/2006 17.4 NGA 2000 MLT Hardware REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS REINSTALLING OF THE PHOTOMETER ASSEMBLY Reinstalling of the Photometer Assembly ❍ Insert the photometer assembly into the analyzer housing and fasten in position using the hexagonal screws shown in Fig. 17-1. ❍ Reconnect all gas lines to the assembly. ❍ Reconnect all electrical connections between the photometer assembly and the electronic unit PCB PIC (see Section 1.9.4).
NGA 2000 MLT Hardware Instruction Manual REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS PHYSICAL ZEROING 17.5 90002929 07/2006 Physical Zeroing Adjustment of the physical zero - level will only be required if a light source, a filter cell, or an analysis cell have been replaced or repositionned. Needed for the adjustment is a 3 mm hexagon wrench SW 3. ❍ Switch on the analyzer (cf. Section 6.). ❍ Admit zero gas to the instrument.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 18. CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR GENERAL Check / Replacement of electrochemical Oxygen Sensor Before opening gas paths they must be purged with ambient air or neutral gas (N2) to avoid hazards caused by toxic, flammable, explosive or harmful to health sample gas components! CAUSTIC ! The electrochemical O2-sensor contains electrolyte which is caustic and can cause serious burns to skin.
NGA 2000 MLT Hardware Instruction Manual CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR 90002929 07/2006 EO2-SENSOR 18.1 EO2-Sensor Through measuring principle the oxygen sensor will have only a limited life time. The life time of the oxygen sensor is dependent on the sensor itself and on the measured oxygen concentration and is calculated as follows: life time = sensor time (hours) O2 concentration (%) The so-called “sensor time” (operation without oxygen at 20 °C ) is approx. 900.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 18.2 CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR CHECK OF EO2-SENSOR Check of the EO2-Sensor Exchange the sensor, if the voltage is less than 70 % of the initially output voltage. The check requires a digital voltmeter (DVM) with a range of 2 V DC. ❍ Remove front panel (see 15.). ❍ Switch On the analyzer (see Section 6.). ❍ Admit ambient air to the analyzer (approx. 21 Vol. - O2).
NGA 2000 MLT Hardware Instruction Manual CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR 90002929 07/2006 REPLACEMENT OF EO2-SENSOR 18.3 Replacement of the EO2-Sensor 18.3.1 Removal of the EO2-Sensor ❍ Remove front panel (see Item 15.). ❍ MLT 1 only: Remove the fastening hexagon nuts of the connection block (Fig. 18-2) with hexagon spanner (SW 5.5) and remove connection block including oxygen sensor from front panel. PCB “OXS” OUT IN Fastening nut connection block O2 sensor Fig.
Instruction Manual 90002929 07/2006 18.3.2 NGA 2000 MLT Hardware CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR EXCHANJGE / REINSTALLING OF EO2-SENSOR Replacing the EO2-Sensor ❍ Disconnect the connector for the sensor from “P2” of circuit board “OXS” (see Fig. 18-3). ❍ Take the consumed sensor out of the fitting. ❍ Take off the stopper from new sensor and fit in the new sensor into the fitting, so that the name plate is at the top of the sensor.
NGA 2000 MLT Hardware Instruction Manual CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR 90002929 07/2006 BASIC CONDITIONS FOR EO2-SENSOR 18.3.4 Basic conditions for the EO2-Sensor ❍ Admit ambient air for the analyzer (approx. 21 Vol. - O2) and switch on (see Section 6.). ❍ Connect the DVM to the measuring points Tp 1 (Signal) and Tp 2 ( ⊥ ) of the PCB OXS, mounted directly at the sensor block (Fig. 18-1, see also Fig. 18-2, 1-3, 1-16a and 1-17).
Instruction Manual 90002929 07/2006 18.4 NGA 2000 MLT Hardware CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR TEO2-SENSOR TEO2-Sensor The TEO2 cell is a fuel cell which measures oxygen concentration by oxidizing a lead oxide. The electrical charge per oxidation cycle yields a current directly proportional to the oxygen concentration. This is current is measured, calibrated and displayed. The over all reduction capacity of the cell gives a limited lifetime of the cell.
NGA 2000 MLT Hardware Instruction Manual CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR ❍ 90002929 07/2006 Place the lid on the housing, fasten the screws and close the cover. Purge gas paths with inert gas (nitrogen (N2)) or sample gas as soon as possible to avoid prolonged exposure of the sensor to high concentrations of oxygen. The longer the sensor is exposed to air, the longer it will take for the sensor to recover to low ppm levels.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 19. CLEANING OF HOUSING OUTSIDE Cleaning of Housing Outside Use a liquid general purpose detergent and a lint-free cloth for cleaning the analyzer´s outside.
NGA 2000 MLT Hardware CLEANING OF HOUSING OUTSIDE 19 - 2 Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware 90002929 01/2007 20. TECHNICAL DATA HOUSING Technical Data Certifications EN 61326, EN 61010-1 (for measurment of not flammable gases or not explosive gases resp. (< 50 % LEL). Bigger concentrations requires supplementary NAMUR, CSA-C/US*), C-Tick, PAC, BRML GOSST: VNIIMS, Pattern (Belarussia) protective measures !) *) PCB EXI 01 EC TYPE-EXAMINATION CERTIFICATION TUEV Nord report No.: 98 ATEX 1341 X FDA Test: 0-10 ppm CO (MLT1/3) TUEV Nord report No.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA 90002929 01/2007 HOUSING / OPTIONS Protection class MLT 1/3/4 MLT 2 CAT 200 (according to DIN standard 40050) IP 20 IP 65 (NEMA4/4X) IP 55 (IP 66 w/o breather), optional with internal tropicalization Data line & main line glands EEx e II KEMA, cable diameter 7 to 12 mm Permissible ambient temperature (operation): - MLT 1/ MLT 2/ MLT 3/ MLT 4 - CAT 200 + 5 to + 40 °C (45 °C as option, not EO2) 0 to 50 °C; -30°C (GP & CSA-C/US only) Permissibl
Instruction Manual NGA 2000 MLT Hardware 90002929 01/2007 20.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA 90002929 01/2007 GENERAL SPECIFICATIONS Specifications of MLT: Table 20-1 NDIR / VIS / UV Oxygen Sensor (PO2 and EO2) Thermal Conductivity (TC) Detection limit Linearity Zero-point drift Span (sensitivity) drift Repeatability ≤ 1 % 1) 4) ≤ 1 % 1) 4) ≤ 2 % per week 1) 4) ≤ 0.
Instruction Manual NGA 2000 MLT Hardware 90002929 01/2007 TECHNICAL DATA SPECIFICATIONS ULCO Altered NDIR/VIS/UV-Specifications of MLT-ULCO compared to table 20-1: Table 20-2 COultra low CO2, ultra low Detection limit Linearity Zero-point drift Span (sensitivity) drift Repeatability Response time (t90) Influence of gas flow Influence of temperature (pressure constant) - on zero point - on span (sensitivity) Thermostatting ≤ 0.2 ppm 3) ≤ ± 1 % of NV 3) (NV ≥ lowest range) ≤ ± 0.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA 90002929 07/2006 DIMENSIONS MLT 1 Analyzer module for platform mounting Network NETWORK ME - + Front view (network/24 V dc for platform mounting only) 24 VDC / IN Rear view (network/24 V dc for outside mounting only) CAUTION ! Use either front OR rear supply Side view Analyzer / analyzer module for rack mounting / table-top Front view (analyzer module with blank plate) Rear view Side view Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DATA DIMENSIONS MLT 2 550 492 300 332 approx. 355 10 18 Glands (connection cables) 18 Fig. 20-2: Dimensional sketch / Drill drawing MLT 2 Standard version [all dimensions in mm] 550 492 approx. 355 603 635 300 38 38 38 10 glands (connection cables) 27 18 43 18 280 16 Gas connections Fig. 20-3: Dimensional sketch / Drill drawing MLT 2 Dual housing version [all dimensions in mm] Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA 90002929 07/2006 DIMENSIONS MLT 2 SAMPLE IN SAMPLE OUT PURGE IN 95 PURGE OUT 75 38 135 60 40 105 45 113 40 25 80 113 Fig. 20-4: Dimensional sketch MLT 2 for Ex Zone 2 in standard housing [all dimensions in mm] approx. 355 5 120 182 5 90 50 Control Unit Solenoid Valve (PURGE IN) 13 220 approx. 356 PURGE OUT 38 Fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DATA DIMENSIONS MLT 3 / 4 30 40 PURGE IN SAMPLE IN SAMPLE OUT PURGE OUT 25 75 38 105 135 140 57.1 132.5 Fig. 20-6: Dimensional sketch MLT 2 for Ex Zones with “Z Purge” (MLT 2-NF) or “Continuous purge” in standard housing [all dimensions in mm] 448.8 465.1 483 approx. 537 approx. 488 Front view Top view Fig. 20-7: Dimensional sketch MLT 3/4 [all dimensions in mm] Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA 90002929 07/2006 DIMENSIONS CAT 200 (14.3) (27) (0.85) (0 .6 3) (13) (25.4) Fig. 20-8: Dimensional sketch / Drill drawing CAT 200 [all dimensions in mm (Inch)] 20 - 10 Emerson Process Management GmbH & Co.
Instruction Manual 90002929 01/2007 NGA 2000 MLT Hardware TECHNICAL DATA VOLTAGE SUPPLY: ANALYZERS 20.4 Voltage supply 20.4.1 Analyzers 20.4.1.1 MLT 1/4 Input Voltage Supply 3-pole XLR- Flange (male), lockable 24 V dc (+/- 5 %) / 3 A (MLT 1) / 5 A (MLT 4) [For ac operation {230/120 V} dc supply by options SL10, SL5 (both for rack mounting only) UPS 01 T, or equivalent power supply] 20.4.1.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA 90002929 07/2006 POWER SUPPLIES: UPS01T 20.4.2 Power Supplies 20.4.2.1 UPS 01 T (Universal Power Supply) This power supply unit may be ordered for supplying tabletop analyzers and rack mount analyzers not requiring a rail mounted power supply. Model designation UPS 01 T Input Rated input voltage Input voltage range Rated input power Input connector 120 / 230 V 50/60 Hz 99 - 138 / 187-264 V max. 240 VA IEC plug (fig.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 TECHNICAL DATA 275.3 57.7 54.2 POWER SUPPLIES: UPS01T 105.9 Fig. 20-9: Dimensional sketch UPS 01 T (Universal Power Supply), table-top version rack module turn around 90° [all dimensions in mm, without cable and plugs] PE Fig. 20-10a: IEC mains input plug (shield) 2 Pin 1: Pin 2: Pin 3: Shield: 3 1 ME + 24 V ⊥) 0 V (⊥ Housing Fig. 20-10b: 24 V DC output socket Pin assignment Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA POWER SUPPLIES: EXTERNAL 5A POWER SUPPLY FOR DIN RAIL INSTALLATION 20.4.2.2 90002929 07/2006 External 5A Power Supply Unit for DIN Rail Installation This power supply unit may be ordered for supplying analyzers installed in a rack, where mounting the power supply on a DIN rail is required.
Instruction Manual 90002929 07/2006 NGA 2000 MLT Hardware TECHNICAL DATA POWER SUPPLIES: EXTERNAL 5A POWER SUPPLY FOR DIN RAIL INSTALLATION Right side Front side Bottom side Fig. 20-11: Power Supply SL5 Dimensions (approx. [mm]) Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA POWER SUPPLIES: EXTERNAL 10A POWER SUPPLY FOR DIN RAIL INSTALLATION 20.4.2.3 90002929 07/2006 External 10A Power Supply Unit for DIN Rail Installation This power supply unit may be ordered for supplying analyzers installed in a rack, where mounting the power supply on a DIN rail is required.
Instruction Manual 90002929 07/2006 NGA 2000 MLT Hardware TECHNICAL DATA POWER SUPPLIES: EXTERNAL 10A POWER SUPPLY FOR DIN RAIL INSTALLATION Right side Front side Bottom side Fig. 20-12:Power Supply SL 10 Dimensions (approx. [mm]) Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual TECHNICAL DATA 90002929 07/2006 POWER SUPPLIES: TABLETOP POWER SUPPLY UNITS 20.4.2.4 Tabletop Power Supply Units This power supply unit may be ordered for supplying two tabletop analyzers by one shared power supply unit. A variation for installation in a module rack is available, too, with output currents of 10 A or 5 A.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 21. PIN ASSIGNMENTS VOLTAGE INPUT Pin Assignments To stay in compliance with regulations regarding electromagnetic compatibility it is recommended to use only shielded cables, as optionally available from Emerson Process Management or equivalent. Customer has to take care that the shield is connected in proper way. Shield and signal connector enclosure need to be conductively connected, submin-d plugs and sockets must be screwed to the analyzer.
NGA 2000 MLT Hardware Instruction Manual PIN ASSIGNMENTS 90002929 07/2006 OPTION SIO 21.3 Option SIO (Standard I/O) 21.3.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 PIN ASSIGNMENTS OPTION DIO 21.4 Option DIO (Digitale I/O) 1 19 20 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 (5 - max. 30 VDC / max. 500 mA) Output 1 (5 - max. 30 VDC / max. 500 mA) Output 2 (5 - max. 30 VDC / max. 500 mA) Output 3 (5 - max. 30 VDC / max. 500 mA) Output 4 (5 - max. 30 VDC / max. 500 mA) Output 5 (5 - max. 30 VDC / max. 500 mA) Output 6 (5 - max. 30 VDC / max.
NGA 2000 MLT Hardware Instruction Manual PIN ASSIGNMENTS 90002929 07/2006 CAT 200 21.5 Terminal Assignment of CAT 200 1 1 24 5 1 1 24 2 4 3 5 Fig. 21-6: CAT 200, Junction box, interior view 1 2 3 Cable gland mains Mains terminals Upper signal terminals (24) 4 5 Lower signal terminals (24). Signal cable glands (max. 3) Note! Assignment of signals to signal terminals inside junction box varies depending on selected analyzer options.
Instruction Manual NGA 2000 MLT Hardware 90002929 07/2006 PIN ASSIGNMENTS CAT 200 Mains Terminal Description 1 Hot (line In) 2 Neutral 3 Ground 4 Ground 5 Ground Table 21-1: CAT 200 Power Connections Terminal Assignments Terminal Description Upper 1 ± 22 mA, (burden 500 Ω), Output 1 Upper 2 ± 22 mA, (burden 500 Ω), Output 2 Upper 3 ⊥ (mA), Outputs 1 + 2 Upper 4 ± 22 mA, (burden 500 Ω), Output 3 (option) Upper 5 ± 22 mA, (burden 500 Ω), Output 4 (option) Upper 6 ⊥ (mA), Outputs
NGA 2000 MLT Hardware Instruction Manual PIN ASSIGNMENTS 90002929 07/2006 CAT 200 Terminal RS 232 RS 485 Lower 1 Ground Ground Lower 2 RxD RxD - Lower 3 TxD RxD + Lower 4 Not used TxD + Lower 5 Ground TxD - Table 21-5: CAT 200 RS 232 / RS 485 Serial Interface Terminal Assignments (Option SIO) Terminal Description Upper 13 (5 - max. 30 VDC / max. 500 mA) Output 1 Upper 14 (5 - max. 30 VDC / max. 500 mA) Output 2 Upper 15 (5 - max. 30 VDC / max.
Instruction Manual 90002929 07/2006 21.6 NGA 2000 MLT Hardware PIN ASSIGNMENTS CAT 200 Increased Safety Box - Label Schematic Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual PIN ASSIGNMENTS 90002929 07/2006 Table 21-7: CAT 200 Terminal Assignments - Bottom (Lower) Contacts Table 21-8: CAT 200 Terminal Assignments - Top (Upper) Contacts 21 - 8 Emerson Process Management GmbH & Co.
Instruction Manual 90002929 07/2006 22. NGA 2000 MLT Hardware WATER CONTENT CALCULATION FROM DEW-POINT TO VOL.-% OR g/Nm³ Calculation of Water Content from Dew-point to Vol. -% or g/Nm³ Table 22-1 Dew-point °C + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 Emerson Process Management GmbH & Co.
NGA 2000 MLT Hardware Instruction Manual WATER CONTENT CALCULATION FROM DEW-POINT TO VOL.-% OR g/Nm³ Dew-point °C + + + + + + + + + + + + + + + + + + + + + + + + + + + Remark: 22 - 2 42 44 45 46 48 50 52 54 55 56 58 60 62 64 66 68 70 72 74 76 78 80 82 84 86 88 90 °F + + + + + + + + + + + + + + + + + + + + + + + + + + + 107,6 111,2 113,0 114,8 118,4 122,0 125,6 129,2 131,0 132,8 136,4 140,0 143,6 147,2 150,8 154,4 158,0 161,6 165,2 168,8 172,4 176,0 179,6 183,2 186,8 190,4 194,0 Water Content Vol.
Instruction Manual 90002929 07/2006 Emerson Process Management GmbH & Co.
Instruction Manual NGA 2000 MLT Hardware WORLD HEADQUARTERS ROSEMOUNT ANALYTICAL EUROPE Emerson Process Management GmbH & Co. OHG Industriestrasse 1 63594 Hasselroth Germany T 49 6055 884 0 F 49 6055 884209 Emerson Process Management Rosemount Analytical Inc. 6565 P Davis Industrial Parkway Solon, OH 44139 USA T 440.914.1261 Toll Free in US and Canada 800.433.6076 F 440.914.1271 e-mail: gas.csc@EmersonProcess.com www.raihome.