Instruction manual
Table Of Contents
- INSTRUCTION MANUAL HFID
- ESSENTIAL INSTRUCTIONS
- TABLE OF CONTENTS
- PREFACE
- DEFINITIONS
- SAFETY INSTRUCTIONS WIRING AND INSTALLATION OF THIS APPARATUS
- OPERATING AND MAINTAINING THIS APPARATUS
- GENERAL SAFETY NOTICE / RESIDUAL RISK
- AUTHORIZED PERSONNEL
- GASES AND GAS CONDITIONING (SAMPLE HANDLING)
- POWER SUPPLY
- ELECTROSTATIC DISCHARGE
- GENERAL PRECAUTIONS FOR HANDLING AND STORINGHIGH PRESSURE GAS CYLINDERS
- DOCUMENTATION
- COMPLIANCES
- GLOSSARY OF TERMS
- ANALYZER SYSTEM ARCHITECTURE
- SECTION 1: DESCRIPTION AND SPECIFICATIONS
- SECTION 2: INSTALLATION
- SECTION 3: OPERATION
- SECTION 4: MAINTENANCE AND SERVICE
- SECTION 5: TROUBLESHOOTING
- SECTION 6: REPLACEMENT PARTS
- RETURN OF MATERIAL
- LIST OF FIGURES AND TABLES
- SUPPLEMENT

Instruction Manual
HAS64E-IM-HW
September 2008
NGA 2000 HFID
4 - 11Maintenance and ServiceEmerson Process Management GmbH & Co.OHG
c. Burner Jet Installation
1. Install new lower gasket, jet assembly and upper
gasket into burner base, finger-tight jet nut.
2. Install new sample jet (with gasket) and tighten.
CAUTION
BURNER CONTAMINATION
Do not handle internal parts of the burner with bare
hands. All tools used for maintenance must be free
of contaminates.
3. Tighten jet nut.
4. Install air baffle per Figure 4-8.
NOTE
Incorrect installation of air baffle will cause igni-
tion failure.
5. See Figure 4-7B. Insert new o-ring into burner
base.
6. Set combustion chamber into burner base being
careful not to move air baffle.
7. See Figure 4-7A. Insert new gasket on combus-
tion chamber, install burner cap and burner cap
retainer, torque screws to 6 inch lbs.