Instruction manual

Table Of Contents
NGA 2000 HFID
Instruction Manual
HAS64E-IM-HW
September 2008
4 - 10 Maintenance and Service Emerson Process Management GmbH & Co.OHG
7. Refer to Figure 4-8. Lift air baffle out of burner
base.
8. Remove the sample jet and gasket from the bot-
tom of the burner base.
9. Remove the jet nut. Grasp jet assembly and lift
out (along with upper gasket) of burner base. Re-
move bottom gasket.
Figure 4-8: Burner Jets
Air Baffle
End bent to raise it above the tab a
distance equal to the height of the tab.
Position with tab over
the air hole
(approximately 1/16" to the right).
.06
Sample Jet
657005
Gasket
102256
Sample Jet
657005
Gasket
102273
Nut, Jet
016005
Air Baffle (Ref)
102260
(see below)