Instruction manual

Table Of Contents
NGA 2000 HFID
Instruction Manual
HAS64E-IM-HW
September 2008
3 - 18 Operation Emerson Process Management GmbH & Co.OHG
c. Relay Outputs Setup
There are three relays on the SIO board. The contact
logic can be set with a jumper on the SIO board to se-
lect NO (normally open) or NC (normally closed). Full
details of the SIO board are contained in its own manual.
In the System SIO Module menu (Figure 3-16) select
"Relay outputs setup..."
Figure 3-25: Relay Outputs Setup Menu
Output number:
Corresponds to the relay number 1-3.
Invert signal:
"Disabled" signal is normal, "Enabled" signal is inverted.
Choose source module...
Figure 3-26: Choose Source Module Menu
Choose desired source module for the relay output num-
ber (1-3) being configured.
The list of modules will depend on the installed mod-
ules.
Choose signal...
Figure 3-27: Choose Signal Menu
Choose desired signal for the relay output number (1-3)
being configured.
The list of signals will depend on the chosen module. If
available, press the >>> (F5) softkey for additional sig-
nals.
The three lines displayed at the bottom of the "Relay
Outputs Setup" menu show the current status of the
selected relay output.
Signal comes from: The module chosenfrom the
"Choose Source Module" menu.
Signal name: The signal chosen from the "Choose Sig-
nal" menu.
Actual status: The current status of the signal; Off or
On.
Line#1
Measure
0.000 ppm
Back...
-- Relay Outputs Setup --
Output number:
Invert signal:
Choose source module...
Choose signal...
Signal comes from:
Signal name:
Actual status:
1
Disabled
????
????
Off
Line#1
Measure >>>
0.000 ppm
<<< Back...
-- Choose Source Module --
Control Module: 0.0
1000723 Line#1: 1.0
Line#1
Measure >>>
0.000 ppm
<<< Back...
-- Choose Signal --
1
Function control
Maintenance request
Failure
Cal. In progress
Zero in progress
Span in progress
Zero failed
Span failed
Pressure limits...