Instruction manual
Table Of Contents
- INSTRUCTION MANUAL HFID
- ESSENTIAL INSTRUCTIONS
- TABLE OF CONTENTS
- PREFACE
- DEFINITIONS
- SAFETY INSTRUCTIONS WIRING AND INSTALLATION OF THIS APPARATUS
- OPERATING AND MAINTAINING THIS APPARATUS
- GENERAL SAFETY NOTICE / RESIDUAL RISK
- AUTHORIZED PERSONNEL
- GASES AND GAS CONDITIONING (SAMPLE HANDLING)
- POWER SUPPLY
- ELECTROSTATIC DISCHARGE
- GENERAL PRECAUTIONS FOR HANDLING AND STORINGHIGH PRESSURE GAS CYLINDERS
- DOCUMENTATION
- COMPLIANCES
- GLOSSARY OF TERMS
- ANALYZER SYSTEM ARCHITECTURE
- SECTION 1: DESCRIPTION AND SPECIFICATIONS
- SECTION 2: INSTALLATION
- SECTION 3: OPERATION
- SECTION 4: MAINTENANCE AND SERVICE
- SECTION 5: TROUBLESHOOTING
- SECTION 6: REPLACEMENT PARTS
- RETURN OF MATERIAL
- LIST OF FIGURES AND TABLES
- SUPPLEMENT

Instruction Manual
HAS64E-IM-HW
September 2008
NGA 2000 HFID
3 - 15OperationEmerson Process Management GmbH & Co.OHG
Figure 3-19: Signals Menu
Signal value for 0% (100%) output:
It is possible to set the signal value for 0 % output and
for 100% output so as to output only a portion of the
entire range.
Example:
• Range from 0 to 1,000 ppm
• 0% value to be 400 ppm,
100% value to be 700 ppm.
• Analog output normally:
0 V = 0 ppm, 10 V = 1,000 ppm
• After changing the output scaling:
0 V = 400 ppm, 10 V = 700 ppm.
Move the cursor to the "Signal value for 0 % output:"
line and adjust the value to 400. Then change to the
"Signal value for 100 % output:" line and adjust the value
to 700.
NOTE
If the measurement range is changed, the settings
done in this menu will revert back to the standard
values of the range. The output values can be
changed permanently in the menu "Range Set-
tings."
NOTE
The signal range of the analog output should be
less than the smallest range of the channel. Oth-
erwise the analog output may exhibit excessive
noise.
Output current:
Select the desired output range in the "Output current
range:" line. The options are 0...20 mA or 4...20 mA.
Hold output during calibration:
Enable this option to hold the analog output to the last
value during calibration.
Pressing the F5 (More...) softkey changes the to the
submenus "Output Signal if Assigned Module Fails" and
"Fine Adjustment."
Figure 3-20: Output Signal If Assigned Module
Fails Menu
Line#1
Measure >>>
0.000 ppm
<<< Back...
-- Signals --
Case temp.:
Preamp temp.:
Flame status:
AutoIgnition:
t-90 time:
Delay time:
Linearizer:
Calibration:
>>>
Line#1
Measure >>>
0.000 ppm
<<< Back...
-- Signals --
Burner air press.:
Air sense:
Burner air:
Burner fuel flow:
Burner fuel press.:
Sample press.:
Sample flow:
Purge control status:
Line#1
Measure More...
0.000 ppm
Back...
-- Output Signal If Assigned Module Fails --
Output(s) value on analyzer failure:
-- Fine Adjustment --
Output number:
Operation mode:
Fine adjustment for 0% output:
Fine adjustment for 100% output:
BeginOfRange - 10%
1
Normal
4096
819