Instruction manual
Table Of Contents
- INSTRUCTION MANUAL HFID
- ESSENTIAL INSTRUCTIONS
- TABLE OF CONTENTS
- PREFACE
- DEFINITIONS
- SAFETY INSTRUCTIONS WIRING AND INSTALLATION OF THIS APPARATUS
- OPERATING AND MAINTAINING THIS APPARATUS
- GENERAL SAFETY NOTICE / RESIDUAL RISK
- AUTHORIZED PERSONNEL
- GASES AND GAS CONDITIONING (SAMPLE HANDLING)
- POWER SUPPLY
- ELECTROSTATIC DISCHARGE
- GENERAL PRECAUTIONS FOR HANDLING AND STORINGHIGH PRESSURE GAS CYLINDERS
- DOCUMENTATION
- COMPLIANCES
- GLOSSARY OF TERMS
- ANALYZER SYSTEM ARCHITECTURE
- SECTION 1: DESCRIPTION AND SPECIFICATIONS
- SECTION 2: INSTALLATION
- SECTION 3: OPERATION
- SECTION 4: MAINTENANCE AND SERVICE
- SECTION 5: TROUBLESHOOTING
- SECTION 6: REPLACEMENT PARTS
- RETURN OF MATERIAL
- LIST OF FIGURES AND TABLES
- SUPPLEMENT

Instruction Manual
HAS64E-IM-HW
September 2008
NGA 2000 HFID
3 - 1OperationEmerson Process Management GmbH & Co.OHG
SECTION 3
OPERATION
3-1 OVERVIEW
Once the HFID has been correctly assembled and in-
stalled, the analyzer module is ready for operation.
Before operating the system, verify that the Leak Checks
have been performed in accordance with Section 2-5.
In this section, all operations for starting up the analyzer
are explained. For more detailed information about soft-
ware screens see associated Software Manual.
For the remainder of this section, Analyzer Module in-
terconnection with a Platform or some interfacing com-
ponent is assumed. Display and Keypad information
refers to that which the user can expect to see and do
with regard to the Front Panel of the Platform.
Depending from the software version that is installed,
menu layout can change, whereas the principle of op-
eration always stays the same.
3-2 STARTUP & INITIALIZATION
a) Startup
1. Connect supply gases and outlets to/from module.
2. Turn ON the purge gas only. Wait a minimum of 6
minutes (see Section 3-2b, too)
3. Apply LON connection and power to the HFID Ana-
lyzer Module. If it is associated with a Platform, do
this by plugging in the Platform to a power source.
The Platform has no ON/OFF power button. Once
power has been supplied to the Platform, the HFID
Analyzer Module will be energized.
After switching on the HFID, the analyzer will be-
gin its booting procedure which is apparent on the
screen. The first part of the initialization procedure
is a self check of the software and analyzer com-
ponents. Various displays will show the status of
the initialization.
PRESSURIZED ENCLOSURE
This enclosure shall not be opened unless the area
is known to be free of flammable materials or un-
less all devices within have been de-energized.
Area classification for the protected enclosure:
Nonclassified.
Pressurization: Type Z
Temperature Identification Number: T4A
Power shall not be restored after enclosure has
been opened (or loss of purge) until enclosure has
been purged for a minimum of 6 (six) minutes at
the minimum pressure of 689 hPa (10 psig).
For safety, the Analyzer Module should be installed
in a non-confined, ventilated space. Do not block
any of the rear panel outlets as they are part of the
safety system.
CAUTION