Instruction manual
Table Of Contents
- INSTRUCTION MANUAL HFID
- ESSENTIAL INSTRUCTIONS
- TABLE OF CONTENTS
- PREFACE
- DEFINITIONS
- SAFETY INSTRUCTIONS WIRING AND INSTALLATION OF THIS APPARATUS
- OPERATING AND MAINTAINING THIS APPARATUS
- GENERAL SAFETY NOTICE / RESIDUAL RISK
- AUTHORIZED PERSONNEL
- GASES AND GAS CONDITIONING (SAMPLE HANDLING)
- POWER SUPPLY
- ELECTROSTATIC DISCHARGE
- GENERAL PRECAUTIONS FOR HANDLING AND STORINGHIGH PRESSURE GAS CYLINDERS
- DOCUMENTATION
- COMPLIANCES
- GLOSSARY OF TERMS
- ANALYZER SYSTEM ARCHITECTURE
- SECTION 1: DESCRIPTION AND SPECIFICATIONS
- SECTION 2: INSTALLATION
- SECTION 3: OPERATION
- SECTION 4: MAINTENANCE AND SERVICE
- SECTION 5: TROUBLESHOOTING
- SECTION 6: REPLACEMENT PARTS
- RETURN OF MATERIAL
- LIST OF FIGURES AND TABLES
- SUPPLEMENT

Instruction Manual
HAS64E-IM-HW
September 2008
NGA 2000 HFID
1 - 7Description and SpecificationsEmerson Process Management GmbH & Co.OHG
d. Gas Connections
Sample In: ........................................... 1/4" O.D. tube fitting, stainless steel
Regulated Air In: .................................. 1/4" O.D. tube fitting, brass
Burner Air In: ....................................... 1/4" O.D. tube fitting, brass
Fuel In: ................................................ 1/4" O.D. tube fitting, stainless steel
Purge Air In: ........................................ 3/8" O.D. tube fitting, brass
Purge Air Out: ...................................... 3/8" O.D. tube fitting, brass
Bypass Out:......................................... 1/4" O.D. tube fitting, stainless steel
Burner Exhaust Out: ............................ 3/8" O.D. tube connection, stainless steel
(must slope downward 6 ° min. from horizontal)
NOTE
Burner Exhaust, Bypass Out and Purge Air Out to be vented to atmospheric pressure and to non-classified
location in accordance with ANSI/NFPA-496 guidelines.
Pressure Relief Valve: ......................... See Caution below
See the Preface section of the Platform manual for specifications regarding Platform-related components (e.g., case
dimensions) and the I/O Module manual for specifications regarding I/O (e.g., relay outputs).
CAUTION
PRESSURE RELIEF VALVE
No connection shall be made to this fitting. If this caution is ignored, damage to the case seals may occur, and
the instrument will not operate properly.
HIGH TEMPERATURES
The Sample In, Byass Out and Burner Exhaust Out connections can reach temperatures of up to 250°C
(480°F). Severe burns could result from touching these connections.
WARNING