Specifications

P
REFACE
748410-A
August 1999
P3
Chemiluminescense Detector Analyzer Module
Rosemount Analytical NGA 2000
WARNING: INTERNAL ULTRAVIOLET LIGHT
Ultraviolet light from the ozone generator can cause permanent eye damage.
Do not look directly at the ultraviolet source. Use of ultraviolet filtering glasses
is recommended.
WARNING: TOXIC CHEMICAL HAZARD
The ozone generator lamp contains mercury. Lamp breakage could result in
mercury exposure. Mercury is highly toxic if absorbed through the skin or
ingested, or if vapors are inhaled.
Handle lamp assembly with extreme care. If the lamp is broken, avoid skin
contact and inhalation in the area of the lamp or the mercury spill.
Immediately clean up and dispose of the mercury spill and lamp residue as
follows:
Wearing rubber gloves and goggles, collect all droplets of mercury by means
of a suction pump and aspirator bottle with a long capillary tube. (Alternately,
a commercially available mercury spill clean-up kit, such as J.T. Baker product
No. 4439-01, is recommended.)
Carefully sweep any remaining mercury and lamp debris into a dust pan.
Carefully transfer all mercury, lamp residue and debris into a plastic bottle
which cab be tightly capped.
Label and return to hazardous material reclamation center. Do not place in
the trash, incinerate or flush down the sewer.
Cover any fine droplets of mercury in non-accessible crevices with calcium
polysulfide and sulfur dust
WARNING: HAND INJURY HAZARD
Do not place hands or fingers in the Platform front handles when front panel is
open. Dropping the front panel of the Platform while hand or fingers are inside
either handle can cause serious injury.