Specifications
P
REFACE
748410-A
August 1999
P2
Chemiluminescense Detector Analyzer Module
Rosemount Analytical NGA 2000
WARNING: TOXIC AND OXIDIZING GAS
This module generates ozone which is toxic by inhalation and is a strong irritant
to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is
detected by a characteristic pungent odor.
The module's exhaust contains both ozone and nitrogen dioxide, both toxic by
inhalation, and may contain other constituents of the sample gas which may be
toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons,
carbon monoxide and other products of combustion reactions. Carbon
monoxide is highly toxic and can cause headache, nausea, loss of
consciousness, and death.
Avoid inhalation of the ozone produced within the module, and avoid inhalation
of the sample and exhaust products transported within the module. Avoid
inhalation of the combined exhaust products at the exhaust fitting.
Keep all tube fittings tight to avoid leaks. See section 2.4.3, page 9, for leak test
information.
Connect rear exhaust outlet to outside vent with stainless steel or Teflon line.
Check vent line and connections for leakage.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair.
CAUTION: PRESSURIZED GAS
This module requires periodic use of pressurized gas. See General Precautions
for Handling and Storing High Pressure Gas Cylinders at the rear of this manual
WARNING: OVERBALANCE HAZARD
This analyzer module may tip instrument over if it is pulled out too far and the
Platform is not properly supported.