Specifications

S
TARTUP AND
O
PERATION
748410-A
August 1999
17
Chemiluminescense Detector Analyzer Module
Rosemount Analytical NGA 2000
3
F
IGURE
3-6. T
YPICAL
H
ELP
S
CREEN
3.5 C
ALIBRATION
Under the Expert Controls and Setup menu, use the
arrow key to select Analyzer Module
Setup, then select Calibration Gas List. Input appropriate data here.
Under the Expert Controls and Setup menu, select Expert Analyzer Controls and then
Range Settings. Select the range that will be used during sample analysis.
Introduce zero gas into the SAMPLE INLET, and, after a stable reading is reached, do the
following:
1. If the multi-Analyzer Module, split Run Mode display is shown, use the arrow keys to
highlight the Analyzer Module to be calibrated and press the select key. The desired
Analyzer Module's Run Mode display is then shown.
2. Press the MENUS softkey to enter the Main Menu.
3. Make the following selections from the Main Menu: Expert Controls and Setup, Expert
Analyzer Controls, and Zero/Span Calibration.
4. Press the ZERO softkey to enter the Analyzer Zero menu.
5. Press the ZERO softkey and wait.
6. Introduce a suitable span gas into the SAMPLE INLET.
7. Press the SPAN to enter the Analyzer Span menu, press SPAN again and wait.
8. Press the HOME softkey to re-enter the Main Menu.
9. Press DISPLAY softkey for the Run Mode display.
23.2 ppm NO Analyzer XXXXXXXX
Main Menu Hel
p
Home Escape Map
F1 F2 F3 F4 F5
The Main Menu for the analyzer system.
Note that this menu refers to the particular
analyzer selected from the run screen, when
used in a system. The softkey marked “HOME”
will always return you to this screen.
Help menu system...
Help on help...
Keyboard controls...
Editing controls...