Specifications

I
NSTALLATION
748410-A
August 1999
7
Chemiluminescense Detector Analyzer Module
Rosemount Analytical NGA 2000
2
MAXIMUM NOX
LEVEL IN PARTS
PER MILLION
GAS SUPPLIED TO
REAR PANEL AIR
INLET
SETTING ON OZONE
PRESSURE GAUGE
(PRESSURE VALUES:
GAUGE)
SETTING ON SAMPLE
PRESSURE GAUGE
(PRESSURE VALUES:
GAUGE)
800 Air
690 hPa (10 psig); provides
flow of approximately 500
cc/min. to ozone generator
344 hPa (5 psig); provides
flow of approximately 200
cc/min. to reaction
chamber
2,500 Air
690 hPa (10 psig); provides
flow of approximately 1000
cc/min. to ozone generator.
344 hPa (5 psig); provides
flow of approximately 70
cc/min to reaction
chamber.
10,000 Oxygen
1035 hPa (15 psig);
provides flow of
approximately 1000 cc/min
to ozone generator.
103 hPa (1.5 psig);
provides flow of
approximately 20 cc/min.
to reaction chamber.
CAUTION: EXTERNALLY LIMIT SAMPLE FLOW RATE TO LESS THAN 2200 CC/MIN.