Specifications

I
NSTALLATION
748410-A
August 1999
5
Chemiluminescense Detector Analyzer Module
Rosemount Analytical NGA 2000
2
2.1 U
NPACKING
If the Chemiluminescence (CLD) Analyzer Module is received as a separate unit, carefully
examine the shipping carton and contents for signs of damage. Immediately notify the
shipping carrier if the carton or contents is damaged. Retain the carton and packing material
until all components associated with the CLD Analyzer Module are operational.
2.2 A
SSEMBLY
If the CLD Analyzer Module requires assembly with other components (e.g., the Platform
and associated I/O Modules), do so at this time. Following the guides on the bottom left and
bottom center of the Platform, carefully slide the Analyzer Module halfway into place.
CAUTION: HAND INJURY HAZARD
Do not place hands or fingers in the Platform front handles when front panel is
open. Dropping the front panel of the Platform while hand or fingers are inside
either handle can cause serious injury.
Lift the spring-loaded pins on the front of the Analyzer Module, and carefully slide it the rest
of the distance. Secure the module in position by releasing the pins, which seat in the
available holes in the bottom of the case (see Figure 2-1). If the module and Platform are
difficult to assemble, remove the module, ensure the top cover of the module is firmly
seated on the hold-down screws, and repeat the assembly procedure.
Refer to Section 2-4 for electrical connections.
Install I/O Module(s) according to guidelines in the I/O manual. After startup and calibration
have been performed, secure the front panel with the six screws provided.
2.3 L
OCATION
Install the CLD Analyzer Module in a clean, weather-protected, vibration-free location free
from extreme temperature variations and moisture. For best results, install the instrument
near the sample stream to minimize sample transport time.
The analyzer should be installed within
±
15
°
of horizontal.