Specifications

I
NTRODUCTION
748410-A
August 1999
3
Chemiluminescense Detector Analyzer Module
Rosemount Analytical NGA 2000
1
TEE
1/8T
KYNAR
VENT CAPILLARY
70cc/min 4 PSIG
PUMP CAPILLARY
1/8 OD TEFLON
VALVE
C
NC NO
TEE 1/8T
KYNAR
TOP
CONNECTOR,
GLASS BLUE
NOx TO NO
CONVERTER
CONNECTOR,GLASS
BLUE
CONNECTOR, MALE
1/8T-3/8MPT SS
CROSS
1/8T SS
SAMPLE
PRESSURE
SENSOR
15 PSIG
TEE 1/8T
KYNAR
OZONE GENERATOR
AIR RESTRICTOR
430cc/min @ 12 PSIG
TEE ASSEMBLY
1/8T-1/4MPT
ELBOW
1/8T-1/4MPT
SS
BACK
PRESSURE
REGULATOR
FLOW
BALANCE
VALVE
ELBOW
1/8T-1/8MPT KYNAR
ELBOW
1/8T-1/8MPT
KYNAR
BYPASS
FLOW
SENSOR
1 70 cc/min. 5 psig Capillary: PN 659657
200 cc/min. 5 psig Capillary: PN 659658
PRESSURE SENSOR
30 PSIG AIR
116 ID VITON BLK
PRESSURE
MALE CONNECTOR
1/8T
-
3/8NPT BRS
CROSS
1/8T BRS
MALE CONNECTOR
1/8T
-
3/8NPT BRS
CONNECTOR
1/8T-1/8NPT BRS
BULKHEAD REDUCER
1/4T - 1/8T BRS
OZONE AIR
BULKHEAD REDUCER
1/4T - 1/8T SS
REDUCER
1/4T - 1/8T SS
CONNECTOR
1/4T SS
TEE
1/4T - 1/8FPT SS
BULKHEAD
1/4T SS
ELBOW
1/8T - 1/8MPT
SS
SAMPLE
EXHAUST
ELBOW 1/8T
SS
REDUCING UNION
1/8T
-
1/16T SS
1/8 OD SS
1/8 OD SS
ELBOW 1/8T
SS
REACTION
CHAMBER
EXHAUST
SAMPLE
OZONE
1
F
IGURE
1-1. CLD A
NALYZER
M
ODULE
F
LOW
D
IAGRAM