Specifications
P
REFACE
748410-A
August 1999
P8
Chemiluminescense Detector Analyzer Module
Rosemount Analytical NGA 2000
S
PECIFICATIONS
- S
AMPLE
T
EMPERATURE
:
0
°
C to 55
°
C (32
°
F to 131
°
F)
T
OTAL
F
LOW RATE
:
(Externally measured) 900 to 2200 cc/min. with
backpressure regulator pressure adjusted to 344 hPa
(5 psig)
P
ARTICLES
:
Filtered to <2 microns
D
EWPOINT
:
5.5
°
C below ambient temperature, no entrained liquid
M
ATERIALS IN CONTACT
WITH
S
AMPLE
:
Stainless steel, Teflon, glass, brass and neoprene
Optional: Stainless steel, Teflon, glass and Kynar
O
ZONATOR
G
AS
:
Clean, dry air or oxygen; flow rate: 1 L/min. maximum;
pressure: 689 hPa to 1034 hPa-gauge (10 to 15 psig);
maintain a constant pressure
±
34 hPa (
±
0.5 psig)
S
PECIFICATIONS
- G
AS
C
ONNECTIONS
O
ZONE
A
IR
:
1/4” O.D. tube fitting, brass
E
XHAUST
:
1/4” O.D. tube fitting, stainless steel
S
AMPLE
I
N
:
1/4” O.D. tube fitting, stainless steel
See the Preface section of the Platform Components manual for specifications regarding
Platform-related components (e.g., case dimensions) and the Preface of the I/O Module
manual for specifications regarding I/O (e.g., relay outputs).