Specifications

Trace Oxygen Analyzer Module
Rosemount Analytical NGA 2000
748420-A
August 1999
21
M
AINTENANCE AND
T
ROUBLESHOOTING
4
4.2 F
USES
Remove power to the Analyzer Module prior to fuse replacement. To replace the Power
Fuse, locate the fuse cover on the front panel of the Analyzer Module, as shown partially in
Figure 2-3. Push and turn the fuseholder cover 1/4 turn counterclockwise. Remove and
replace the fuse as required. There are no other fuses in the Analyzer Module.
4.3 E
LECTROLYTE
R
EPLACEMENT
Before replacing the electrolyte, be sure to turn off and disconnect all gas connections to the
analyzer module. Turn off or disconnect the power to the analyzer module.
To replace the Sensor electrolyte, remove the Analyzer Module from its mounting location
and place on a sturdy work surface. Be careful not to tilt the module from its horizontal
position as the Sensor contains liquid that can spill. Remove the cover of the Analyzer
Module and locate the 5 mounting screws that hold the Sensor Assembly onto the Analyzer
Module chassis (see Figure 4-1). Remove the 5 screws and retain. Do not lose the screws -
they have metric threads.
Disconnect the Sensor signal connector (J5) and the Flow Sensor connector (J6) from the
power board. Remove the complete Sensor Assembly from the Analyzer Module. Remove
the black sensor cover and invert the Sensor Assembly over a suitable receptacle. Flush
the Sensor twice with deionized water. Dispose of the discarded electrolyte and rinse water
in accordance with National, Federal, State and Local regulations.
Refill the Sensor with electrolyte as instructed in Section 2.2.1. Reinstall the Sensor
Assembly and reconnect J5 and J6 to the power board.
4.4 S
ENSOR
R
EPLACEMENT
If the Sensor cannot be regenerated by the addition of water or the replacement of
electrolyte, or if the Sensor shows signs of leakage, it may be necessary to replace the
Sensor. To replace the Sensor, remove the Sensor Assembly and remove the electrolyte
as described in section 4.3 above. Reinstall the black sensor cover to catch any residual
electrolyte. Invert the Sensor Assembly and locate the four (4) mounting screws which hold
the Sensor to the Sensor Assembly mounting plate. Remove and retain the four screws.