Specifications

748420-A
August 1999
6
Trace Oxygen Analyzer Module
Rosemount Analytical NGA 2000
2
I
NSTALLATION
Note
Do not add water. The volume and concentration of the bottled electrolyte is pre-
measured.
Reinstall the black Sensor cover and carefully reinstall the Sensor Assembly inside the
Analyzer Module. Do not the tilt the Sensor Assembly excessively as electrolyte may leak
out.
2.3 L
OCATION
(See Figure 2.2) The TO2 Analyzer Module comes standard with mounting ears for easy
installation on flat, horizontal surfaces. Install the TO2 Analyzer Module in a clean, weather-
proofed, vibration-free location free from extreme temperature variations and moisture. For
best results, install the instrument near the sample stream to minimize sample transport
time.
Operating ambient temperature is 0
°
C to 45
°
C (32
°
F to 81
°
F). Temperature change
should not exceed 10
°
C (18
°
F) per hour. The same temperature restrictions apply to the
location of the zero and span gas cylinders.
2.4 G
ASES
2.4.1 R
EQUIREMENTS
The TO2 Analyzer Module requires only a standard of accurately known composition for use
as a span gas. The span gas should be supplied from a cylinder equipped with a clean,
metallic diaphragm, two-stage regulator. A shutoff valve is recommended.
Calibration Gases
The TO2 module does not require routine zero calibration. The zero is factory set and does
not experience routine drift. Over long periods of time, the zero may experience minor drift.
For low ppm range analyzers, you may wish to check the zero at one year intervals.
Oxygen-free nitrogen is recommended for use as zero gas. This gas is certified to <0.5
ppm oxygen and can be improved by passing the zero gas through an oxygen scrubber
such as Millipore™ Waferpure or Semigas Nanochem® resin purifiers. A mixture of trace
oxygen in a background of nitrogen is recommended as span gas. For maximum accuracy,
the concentration of trace oxygen in the span gas should be as high as possible for the
range of measurement.