Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Trace Oxygen Analyzer http://www.processanalytic.
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications.
Instruction Manual Model NGA 2000 TO2 760008-A September 2001 TABLE OF CONTENTS PREFACE...........................................................................................................................................P-1 intended use statement ......................................................................................................................P-1 Definitions ..........................................................................................................................
Instruction Manual 760008-A September 2001 3-5 3-6 3-7 4-0 4-1 4-2 4-3 4-4 4-5 Model NGA 2000 TO2 e. Display Controls..........................................................................................................3-12 Expert Controls and Setup .......................................................................................................3-13 a. Calibration Procedure .................................................................................................3-14 b. Alarms .......
Instruction Manual Model NGA 2000 TO2 760008-A September 2001 LIST OF ILLUSTRATIONS Figure 1-1. Figure 1-2. Figure 2-1. Figure 2-2. Figure 2-3. Figure 2-4. Figure 3-1. Figure 3-2. Figure 3-3. Figure 3-4. Figure 3-5. Figure 3-7. Figure 3-8. Figure 3-9. Figure 3-10. Figure 3-11. Figure 3-12. Figure 3-13. Figure 3-14. Figure 3-15. Figure 3-16. Figure 3-17. Figure 3-18. Figure 3-19. Figure 3-20. Figure 3-21. Figure 3-22. Figure 3-23. Figure 3-24. Figure 3-25. Figure 3-26. Figure 4-1.
Instruction Manual 760008-A September 2001 iv Contents Model NGA 2000 TO2 Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 PREFACE INTENDED USE STATEMENT The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the Model NGA 2000 TO2 and the System Accessories of the NGA 2000 System. Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SAFETY SUMMARY If this equipment is used in a manner not specified in these instructions, protective systems may be impaired. DANGER. ELECTRICAL SHOCK HAZARD Operate this equipment only when covers are secured. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
Instruction Manual Model NGA 2000 TO2 760008-A September 2001 GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981 Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission 1. Never drop cylinders or permit them to strike each other violently. 2.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 DOCUMENTATION The following NGA 2000 TO2 instruction materials are available. the local representative to order. Contact Customer Service Center or 760008 Instruction Manual (this document) COMPLIANCES This product may carry approvals from several certifying agencies for use in non-hazardous, indoor locations. If so, the product will carry approval insignia on the product name-rating plate. ® NRT L/C Rosemount Analytical Inc.
Instruction Manual Model NGA 2000 TO2 760008-A September 2001 GLOSSARY OF TERMS Analyzer Module The module that contains all sensor/detector components for development of a Primary Variable signal; includes all signal conditioning and temperature control circuitry. Backplane The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and network cables, I/O Modules and Expansion Modules plug into. Control Module The Operator Interface plus the Controller Board.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Softkeys The five function keys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software. System Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s). P-6 Preface Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 1 DESCRIPTION AND SPECIFICATIONS 1-1 The TO2 Analyzer Module uses the coulometric principle of oxygen detection. This technology is based on the fact that oxygen in the sample is reduced by an electrochemical reaction. This reduction occurs at the cathode and results in the generation of hydroxyl ions. These hydroxyl ions migrate to the anode where they are oxidized to reform oxygen.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECONDARY ANODE (+) BI-STRATA™ DIFFUSION BARRIER e- ELECTRONICS S SAMPLE IN KOH O2 OH- O2 SAMPLE OUT S eO2 SENSING CATHODE (-) SECONDARY CATHODE (-) Figure 1-1. 1-2 Description and Specifications O2 SENSING ANODE (-) Trace Oxygen Detector Coulometric Principle Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Sensor Assembly Network/Power Module Sensor Computer Board Sample Flow Sensor Power Board Figure 1-2. Trace Oxygen Analyzer Module - Top View Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 1-5 Model NGA 2000 TO2 SPECIFICATIONS1 a. General Measurement Species................................... Trace Oxygen Ranges .......................................................... 0 to 100 ppm (output scalable down to 0-2 ppm fullscale) Accuracy........................................................ ±3% of reading or ±0.02% of range (except for ranges ≤ 100 ppm: ±3% of reading or ±0.05% of range) Sensitivity .................................................
Instruction Manual Model NGA 2000 TO2 760008-A September 2001 d. Gas Connections Sample In ...................................................... 1/4 inch O.D. tube fitting Sample Out.................................................... 1/4 inch O.D. tube fitting Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 1-6 Description and Specifications Model NGA 2000 TO2 Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 2 INSTALLATION Figure 4-1). Be careful not to lose these screws as they have metric threads. Carefully lift out the Sensor assembly and remove from the analyzer module. Place on a flat surface and remove the black Sensor cover by unscrewing counterclockwise. WARNING. Before starting to install this equipment, read the Safety Summary in the Preface section of this manual.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 The same temperature restrictions apply to the location of the zero and span gas cylinders. 2-4 Module to control flow, the inlet pressure to the needle valve should not exceed 345 hPa (5 psig). A constant sample flow rate between 1.0 to 3.0 SCFH (0.5 to 1.5 l/min) is recommended for best results. The Analyzer Module must vent to atmosphere to avoid back pressure influences on the oxygen reading. GASES a.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 2-6 ELECTRICAL CONNECTIONS a. Leak Test WARNING. The TO2 Analyzer Module is completely tested at the factory for gas leakage. The user is responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. CAUTION. SENSOR DAMAGE Do not expose the Sensor to pressure in excess of 1.0 psig as this may cause damage. Flow Indicator Method Supply air or inert gas such as nitrogen, at 1 psig (6.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 ANALYZER MODULE CONNECTIONS BACKPLANE CONNECTIONS Network 1 Backplane Network 2 Power Controller Board Connector Fuse Network Power Indicator Light Power Figure 2-1. Analyzer Module Interconnection with Instrument Platform 2-4 Installation Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 .23 [6] 7.75 [197] 7.75 [197] 1.61 [41] .23 [6] 5.78 [147] .266 DIA [6.75] 18.56 [471] 12.00 [305] 11.00 [279] 6.00 [152] 7.00 [178] 8.25 [206] 8.10 [206] 10.15 [258] 2.70 [68] 6.62 [168] 1.2 [31] .06 [1.5] 1.2 [31] Figure 2-2. Outline and Mounting Dimensions Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Sample Inlet 1/4" Tube Sample Exhaust 1/4" Tube Figure 2-3. Back Panel Connections Network Connections Power Connection Fuse Power Indicator Light Figure 2-4. Trace Oxygen Analyzer Front Panel 2-6 Installation Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 3 OPERATION 3-1 In case of power failure, all user defined specific module parameters are saved by a battery powered memory. OVERVIEW Once the TO2 has been correctly assembled and installed in accordance with the instructions in Section 2, the analyzer is ready for operation. The Function Keys, also called softkeys, are assigned values depending on the menu or screen being displayed. The legend is displayed above the keys.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Variable Line – A line ending with a colon (:) indicates that it displays a module variable parameter. Some parameters can be changed and some parameters display only a status and cannot be changed. Paramters that cannot be changed will be displayed below a line within the menu. a. Menu Lines & Softkey Functionality Menu lines can be selected with the ↑ key or the ↓ key.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 the channels of the current Analyzer or Analyzer module. F3 if available, F4 in the single component display. the sign from positive to negative if applicable. Lock – Changes to the main menu and locks all three operation levels, if a security code is enabled in the system configuration. F4 in the main menu. 4.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Main menu it is possible to change all operating values of the Analyzer to set up and control the parameters of measurement, calibration and data transfer. e. Measure Mode Display The Measure Mode is the normal mode of operation. In this mode, the Display will show the current gas measurement, the component of interest, the current operations of the softkeys, and several graphics.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 TO2 37.50 ppm -- Main Menu -- Analyzer basic controls (calibration) & setup… Analyzer and I/O, expert controls & setup… System configuration and diagnostics… Display controls… Time & Date: System tag: 10:30:05 August 10 2001 Fisher-Rosemount Status… Measure Channel Lock… TO2 MFG Data 37.
Instruction Manual 760008-A September 2001 3-3 Model NGA 2000 TO2 Pressing the F1 key during initializing will reset the LCD brightness and contrast to factory settings (See Section 3-4e). Pressing the F3 key will abort the network initializing, aborting any connection to other analyzers. In that case, only the menus of the local analyzer will be available. STARTUP & INITIALIZATION Establish sample or zero gas flow through the analyzer module at a nominal flow rate of 2 SCFH (1 l/min).
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 tration in the sample gas, the longer the analyzer will take to reach equilibrium. a. Quick Start This analyzer module is equipped with a quick start feature which allows the sensor to begin measuring low ppm oxygen faster. This feature can be used when the analyzer is first turned on to decrease the time required for the sensor to reach equilibrium. This function is most effective for gas sample measurements below 100 ppm.
Instruction Manual 760008-A September 2001 3-4 Model NGA 2000 TO2 BASIC CONTROLS, SETUP AND STATUS The following sections describe the basic control of the analyzer and the viewing of channel parameters. Examples of stepping through the menus are shown so that the user can become familiar with the operation, keeping in mind that displays and menu choices may be different depending on actual analyzer configuration and any customization of the menus. a. Analyzer Channel Status TO2 37.50 ppm O2 0.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 are connected to the Platform, it is possible to display them using the following steps to change the channel of the single component display: b. Single Component Display The Measure menu that displays after startup is the Single Component display of the analyzer. If other analyzer modules TO2 37.50 ppm O2 0.00 Range: 3 Electrolyte level: 50.00 OK Temp current: 263 uA 0.0 Range: 600 3 Sensor Current: 0.0500 uA 0.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 to five using the following steps to change from the single component display to the multi component display as follows: c. Multi Component Display If other analyzer modules are connected to the Platform, it is possible to display up TO2 37.50 ppm O2 0.00 Range: 3 Electrolyte level: OK Temp current: 263 uA 0.0 Range: 600 3 Sensor Current: Display 50.00 0.0500 uA 0.0000 Status… Main… TO2 ppm O2 > 37.50 100.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 d. Basic Controls and Setup The Basic Controls and Setup menu is used to set the range, initiate and exit a quick start and exit the sleep mode. See section 0 to change the begin and end concentration values for each of the four ranges. TO2 37.50 ppm O2 0.00 Range: 3 Electrolyte level: OK Temp current: 263 uA 0.0 Range: 600 3 Sensor Current: Display 50.00 0.0500 uA 0.0000 Status… Main… 100.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 e. Display Controls To adjust the display parameters, from the Main menu choose “Display controls…” as follows: TO2 37.50 ppm -- Main Menu -- Analyzer basic controls (calibration) & setup… From the Main menu, move the cursor to the “Display controls…” line and press the ↵ key. The menu will display as shown below.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 3-5 EXPERT CONTROLS AND SETUP The Expert Controls and Setup menus provide for the configuration of system and network I/O (SIO & DIO), and for the TO2 37.
Instruction Manual 760008-A September 2001 a. Calibration Procedure The TO2 analyzer module is fully factory calibrated prior to shipment using certified gas standards. If the analyzer is operated within its specified operating conditions, no regular calibration is required. The zero calibration is very stable and does not require checking more than once a year. Gas Scale Factor The Gas Scale Factor is used to correct for background gases other than nitrogen.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Main Menu ↓ Analyzer and I/O, expert controls and setup… ↓ Analyzer module controls… (Or Analyzer module setup…) ↓ TO2 37.50 ppm -- Analyzer Module Setup and Controls -- Calibration… Alarms… Ranges… Linearization… Measurement display configuration… Concentration measurement… Flow measurement… Temperature measurement… Measure Move the cursor to the “Calibration…” line and press the ↵ key. The “Calibration” menu will display as shown below.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 span calibration of the analyzer module every 3-4 months. The following procedure illustrates how to initiate a span calibration. Span Calibration Depending upon the nature of your application, it may be beneficial to verify the Main Menu ↓ Analyzer and I/O, expert controls and setup… ↓ Analyzer module controls… (Or Analyzer module setup…) ↓ TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 b. Alarms From the Analyzer Module Setup and Controls menu, choose Alarms as follows: Main Menu ↓ Analyzer and I/O, expert controls and setup… TO2 ↓ 37.50 ppm -- Analyzer Module Setup and Controls -- Calibration… Alarms… Ranges… Linearization… Measurement display configuration… Concentration measurement… Flow measurement… Temperature measurement… Measure Back… Status… TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 c. Ranges From the Analyzer Module Setup and Controls menu, choose Ranges as follows: Main Menu ↓ Analyzer and I/O, expert controls and setup… TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 d. Linearization e. Measurement Display Configuration Linearization is only performed when the sensor is changed. See section 4-3 below. TO2 37.
Instruction Manual 760008-A September 2001 f. Model NGA 2000 TO2 Concentration Measurement Gas measurement units: This menu establishes several features and parameters of the concentration measurement and display. The values on this menu are not usually changed. ppm Parts per million. mg/Nm3 Reading adjusted by the conversion factor below. ppb Parts per billion. % Percent should only be used for higher range sensor models. Main Menu ↓ Analyzer and I/O, expert controls and setup… TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Sensor model: Clamp low PVA to 0.02: Adjustable for both “Standard” and “Stabel” models in the following ranges: This parameter is used for very low oxygen concentration levels where the sensor current may be unstable about zero. Normally this would be set to “No.” 0-50 ppm 0-100 ppm ppm to mg/Nm3 conversion factor: 0-500 ppm The ppm to mg/Nm3 conversion factor in the range of 1 to 100000.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 g. Flow Measurement Sample flow should be maintained within a range suitable for the desired accuracy. See Effect of Flow in Section 1-5a. Main Menu ↓ Analyzer and I/O, expert controls and setup… TO2 37.50 ppm -- Analyzer Module Setup and Controls -- Calibration… Alarms… Ranges… Linearization… Measurement display configuration… Concentration measurement… Flow measurement… Temperature measurement… Measure Status… Back… TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 h. Temperature Measurement Used to set the temperature display units. Main Menu ↓ Analyzer and I/O, expert controls and setup… TO2 37.50 ppm -- Analyzer Module Setup and Controls -- Calibration… Alarms… Ranges… Linearization… Measurement display configuration… Concentration measurement… Flow measurement… Temperature measurement… Measure Back… Status… TO2 37.
Instruction Manual 760008-A September 2001 3-6 Model NGA 2000 TO2 SYSTEM & NETWORK I/O MODULE CONTROLS (SETUP) This menu provides access to several submenus for setting parameters of the SIO (Signal Input/Output) and DIO (Digital Input/Output) of the analyzer. Main Menu ↓ Analyzer and I/O, expert controls and setup… TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 eral configuration of the SIO board is contained in its own manual. a. System SIO This menu provides submenus for setting up the output configurations of the SIO signals. The SIO board can contain 2 to 8 analog outputs, a serial interface (RS232 or RS485), and three relay outputs. Gen- If the SIO board is installed in the analyzer, the line “Module installed:” must be set to “Yes.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 turn to the previous menu automatically and the selected reference channel will be displayed in the “Source module:” line. b. Analog Output Setup In the System SIO Module menu (Figure 3-25), select Analog output setup… Choose signal TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 analog output may exhibit excessive noise. Signal value for 0% (100%) output TO2 37.50 ppm Output current -- Analog Output Setup -Output number: Choose signal source module… Choose signal… Signal value for 0% output: Signal value for 100% output: Output current: Hold output during calibration: Signal name: Current signal value: Source module: Measure 1 TO2 37.50 ppm -- Analog Output Setup -- 0.00 100.00 0…20 mA No Sample flow 8.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Normal: The absolute measurement signal will be sent to the analog output. Output(s) value on analyzer failure TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 c. Serial Interface Setup Special Scaling for Concentration Signal TO2 From menu System SIO Module (Figure 3-25), select Serial interface setup… 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 open) or NC (normally closed). Full details of the SIO board are contained in its own manual. Communication protocol: AK MODBUS RTU The three lines displayed at the bottom of the “Relay Outputs Setup” menu show the current status of the selected relay output. None (not applicable to TO2) NOTE: The “special protocol definitions…” line accesses a submenu for setting the parameters of the AK and MODBUS TRU communication protocols.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Choose signal… Brings up menu shown below. TO2 37.50 ppm -- Choose Signal -Function control Maintenance request Failure Cal. In progress Zero in progress Span in progress Zero failed Span failed Back… Measure Choose desired source module for the relay output number (1-3) being configured. The list of modules will depend on the installed modules. e.
Instruction Manual 760008-A September 2001 3-7 Model NGA 2000 TO2 SYSTEM CONFIGURATION AND DIAGNOSTICS This menu and its submenus provides for setup of the system parameters for the platform. The menu is accessed from the Main Menu. TO2 37.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Load/Save Module Configuration a. Diagnostic Menus TO2 37.50 ppm -- Diagnostics Menus -TO2 Control module diagnostics… Analyzer module diagnostics… 37.50 ppm -- Load/Save Configuration (CM/MCA) -- Send configuration to seral interface! Load configuration from serial interface! - BE CAREFUL with this function <<< Measure Back… >>> This menu has two submenus for viewing and resetting any software errors.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Date and Time CAUTION. This menu is used to set the date, time and format for the analyzer. TO2 37.50 ppm If a security code is lost or forgotten, there is no possibility of entering the locked security level.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Entering a level locked by security code For example, return to the Main Menu by pressing the F4 key twice. Attempt to enter an enabled level by choosing the menu line. A new menu will appear requesting entry of the security code. Enter the correct code using the correct sequence of function keys. The asterisk (*) symbol will appear for each entry.
Instruction Manual 760008-A September 2001 3-36 Operation Model NGA 2000 TO2 Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 4 MAINTENANCE AND SERVICE 4-1 OVERVIEW The TO2 Analyzer Module requires very little maintenance during normal operation. DANGER. ELECTRICAL SHOCK HAZARD Operate this equipment only when covers are secured. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 label. Be careful not to spill water, splash electrolyte or overfill sensor. CAUTION. REFILLING SENSOR 5. Replace the sensor cover securely. When refilling the sensor, only use distilled or deionized water. Do not use electrolyte or tap water as they can cause damage to the sensor. Take care not to overfill. 6. Replace the top cover of the analyzer module.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 b. Electrolyte Replacement Before replacing the electrolyte, be sure to turn off and disconnect all gas connections to the analyzer module. Turn off or disconnect the power to the analyzer module. 7. Refill the Sensor with electrolyte as instructed in Section 4-2b. 8. Reinstall the Sensor Assembly and reconnect J5 and J6 to the power board. c. Sensor Replacement To replace the Sensor electrolyte: CAUTION.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 possible leakage caused by damage in shipment. 1. If not already removed from the analyzer, remove the top cover of the analyzer module and locate and remove the five mounting screws which hold the Sensor Assembly (Sensor, flowmeter, plumbing) to the module. Care should be taken not to lose the screws as they have metric threads. 2. Carefully lift out the Sensor Assembly.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 TO2 37.50 ppm TO2 -- Linearization Set Point Values -- -- Analyzer Module Setup and Controls -Y1: Y2: Y3: Y4: Calibration… Alarms… Ranges… Linearization… Measurement display configuration… Concentration measurement… Flow measurement… Temperature measurement… Measure Status… 37.50 ppm 0.0000 ppm 1.050 ppm 7.000 ppm 72.800 ppm Measure Back… Select Linearization… Back… Enter the data for each set point value. TO2 37.
Instruction Manual 760008-A September 2001 4-4 FLOW SENSOR REPLACEMENT See Figure 1-2 for Flow Sensor location. To replace Flow Sensor, remove all connecting hardware and undo connections to the sample line. The Flow Sensor is mounted to the Sensor Assembly mounting plate by two screws. Be sure to install the new Flow Sensor with the flow indication toward the outlet. 4-5 ELECTRONIC COMPONENTS a. Fuses Remove power to the Analyzer Module prior to fuse replacement.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 5 TROUBLESHOOTING 5-1 TROUBLESHOOTING The following provides a short list of common troubleshooting tips. Additional information is contained in the Platform Manual. a. Failure to purge down to ppm levels Prior to conducting any changes to the system, try running a quick start sequence (see section 3.4) to see if the oxygen reading goes lower.
Instruction Manual 760008-A September 2001 5-2 Maintenance and Service Model NGA 2000 TO2 Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 6 REPLACEMENT PARTS 6-1 WARNING. PARTS INTEGRITY Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair. TO2 MATRIX Each analyzer is configured per the customer sales order. Below is the TO2 sales matrix which lists the various configurations available. To identify the configuration of an analyzer, locate the analyzer name-rating plate.
Instruction Manual 760008-A September 2001 6-2 6-2 Model NGA 2000 TO2 REPLACEMENT PARTS 658350 Computer Analysis Board 657466 LON/Power Board 658300 Power Supply Board 902931 Sensor, Gas Flow 904675 Sensor, Oxygen 0-100 ppm 904676 Electrolyte Solution 903347 Fuse, Time-Delay 6A 250 VAC Replacement Parts Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 7 RETURN OF MATERIAL 7-1 RETURN OF MATERIAL If factory repair of defective equipment is required, proceed as follows: 1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
Instruction Manual 760008-A September 2001 7-2 Return of Material Model NGA 2000 TO2 Rosemount Analytical Inc.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 SECTION 8 INDEX A F Analog output setup, 3-26 Analog output number, 3-26 Choose analyzer module, 3-26 Choose signal, 3-26 Hold output value, 3-27 Operation mode, 3-28 Output current, 3-27 Output number, 3-28 Output rate, 3-29 Output value on analyzer failure, 3-28 Signal value, 3-27 Special scaling for concentration signal, 3-29 Analyzer Module Installation conditions, 2-1 Installation location, 2-1 Leak test, 2-3 Analyzer module diagnostics,
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Q Quick start Aborting, 3-7 Accessing, 3-7 R Relay outputs Choose signal, 3-31 Choose source module, 3-30 Invert signal, 3-30 Output number, 3-30 Replace current configuration with factory settings, 3-33 Return Authorization, 7-1 S Sample Installation conditions, 2-2 Security code enter an enabled level, 3-35 locked, 3-35 security codes, 3-34 Send configuration to serial interface, 3-33 Sensor Adding water, 4-2 Calibration data, 4-4 Leak chec
Rosemount Analytical 748377-B Material Safety Data Sheet PRODUCT: PART NUMBER: ELECTROLYTE1 904676 24 HOUR EMERGENCY TELEPHONE NUMBER: CHEMTREC (800) 424-9300 SECTION I - GENERAL Distributor: Rosemount Analytical Inc.
MATERIAL SAFETY DATA SHEET 748377-B SECTION VI – HEALTH HAZARD DATA Threshold limit value Routes of entry Effects of overexposure Emergency & first aid procedures NA Inhalation, eyes, skin, ingestion Corrosive to tissue. Inhalation of mist may cause respiratory tract damage, Eyes - Corrosive, immediately flush with water for at least 15 minutes. Call a physician. Skin - Corrosive, remove contaminated clothing. Wash with soap or mild detergent and large amounts of water at least 15 minutes.
ROSEMOUNT WARRANTY Rosemount warrants that the equipment manufactured and sold by it will, upon shipment, be free of defects in workmanship or material. Should any failure to conform to this warranty become apparent during a period of one year after the date of shipment, Rosemount shall, upon prompt written notice from the purchaser, correct such nonconformity by repair or replacement, F.O.B. factory of the defective part or parts.
Instruction Manual 760008-A September 2001 Model NGA 2000 TO2 Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com Fisher-Rosemount GmbH & Co. Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209 ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.