Network Router User Manual
Table Of Contents
- Contents
- Before You Begin
- Installation and Setup
- 2.1 Overview
- 2.2 Installation and setup
- 2.2.1 Ensure required privileges
- 2.2.2 Install the ProLink II software
- 2.2.3 Generate the temporary license
- 2.2.4 Determine your connection type
- 2.2.5 Install the signal converter and connect the wires
- 2.2.6 Configure ProLink II connection parameters and connect to the transmitter
- 2.2.7 Obtain and configure a site key
- 2.3 Troubleshooting the ProLink II installation
- 2.4 Troubleshooting the ProLink II connection
- Using ProLink II Software
- Initial Transmitter Startup Procedures
- Transmitter Configuration, Characterization, and Calibration
- Meter Verification
- Data Logger
- Transmitter Terminal Reference
- Configuring the Discrete Batch Application
- Index
98 ProLink
®
II Software for Micro Motion
®
Transmitters
Configuring the Discrete Batch Application
B.5 Performing Batch AOC calibration
Batch AOC (Automatic Overshoot Compensation) keeps the actual quantity delivered as close as
possible to the batch target, by minimizing the amount of overshoot.
If batch AOC is enabled (see Section B.3.2), batch AOC calibration is required to provide data for the
compensation process. Thereafter, batch AOC calibration is recommended:
• Whenever consistent overshooting or undershooting of the target occurs
• If equipment (valve or pump) is changed
Note: If batch AOC is not enabled, you cannot perform batch AOC calibration.
To perform batch AOC calibration using ProLink II:
1. In the
File menu, click on Calibration.
2. Click on
AOC calibration.
3. Press the
Start Calibration button. The AOC Calibration Active light turns red, and will
remain red while batch AOC calibration is in progress.
4. Run 2–10 batches.
5. When overshoot has been adequately minimized, press the
Save Calibration button.