Specifications

MAINTENANCE and SERVICE
5.2 Converter
Refer to Figure 5-1, Item 97, and Figure 5-3. To replace the converter or
sensor, disconnect the two pneumatic tubes and two electrical connections.
Unlace the heater blanket, and remove the converter. Reassemble in
reverse order, ensuring that the converter is oriented with the glass cloth at
the bottom and the sensor is oriented correctly inside the heater jacket.
Figure 5-3. Converter Assembly
ASSEMBLED SIDE VIEW
Converter
Tube 655227
Wrap with
aluminum foil
Glass
Cloth
Sensor
Heater
Jacket
655228
Sensor
655282
5.3 Ozonator
Refer to Figure 5-1, item 98.To replace the ozonator, remove the gas
fittings, the two large straps and all tie-wraps, and disconnect the one
electrical connection. Reassemble in reverse order.
5.4 Personality Modules
There are seven different personality modules. Depending on your unit, you
may have three, four five, six, or seven modules installed. These
personality modules are installed on a custom backplane. see figure 5-4 for
more information.
Tag each cable and its location before disconnecting any wiring. This helps
in re-assembly
To remove any on the personality modules. Remove cables from the
module to be removed, there are two screws at the bottom of each module.
You will have to loosen each screw before you can remove the personality
module.
Rosemount Analytical ยตCEM Continuous Analyzer Transmitter 5-3