Specifications
MAINTENANCE and SERVICE
5. Maintenance and Service
CAUTION: QUALIFIED PERSONNEL
This equipment should not be adjusted or repaired by
anyone except properly qualified service personnel.
WARNING: PARTS INTEGRITY
Tampering with or unauthorized substitution of components
may adversely affect safety of this product. Use only factory-approved
components for repair.
WARNING: ELECTRICAL SHOCK HAZARD
Disconnect power to the module(s) prior to replacing
components.
The uCEM Analyzer Module requires very little maintenance during normal
operation.
5.1 Overview
The uCEM Analyzer Module requires very little maintenance during normal
operation.
Occasionally, the detector's reaction chamber and sapphire window may
require cleaning, refer to Section 5.7.
White crystal deposits on the windows of the reaction chamber and plugging
of capillaries and vent are usually due to sample contaminates such as
ammonia reacting with the high ozone levels and NO components. To
eliminate the contaminates, the sampling system should be reworked or a
preventive maintenance program developed (if dropout is not excessive).
Another source of crystalline formation is contaminated air.
Several components may require replacement. These are discussed in the
following sections.
Rosemount Analytical ยตCEM Continuous Analyzer Transmitter 5-1