Specifications
STARTUP and OPERATION
zone may show than the AC/Heater unit fan may have failed or a possible defective
thermocouple.
*PDD Temp: For systems with NOx a cylindrical NOx detector assy is located in the
detector section. Internal to the detector assy a small peltier device is operating and
must operate at 0 degrees C. The temperature should never deviate +/- .05 degrees C
from the setpoint of zero or the NOx readings may drift. Integral will typically run
between 40 to 70%.
*PMT Temp: This temperature is for the detector assy heater core. Setting is set to 52
degrees C. Temperature should not drift more than 0.2 degrees C or NOx drift may
occur. Excessive temperature variation may be caused by either poor zone temperature
control or a faulty heater.
*Conv Temp: This temperature is for the NOx converter assembly. Temperature setting
is 330 degrees C. Temp should not vary more than 1 degrees C. or NOx
measurements will drift. A faulty heater will cause temp variations.
Note that when the enclosure door is opened that all of the above temperature setting
may be affected and will take a short about of time to react and control to the desired
temperatures.
Figure 4.16 - View Data Logs
Table 4.10 - Average Period Selection
Note: The page header
was scrolled out of view
to show all the selection
options, but it can be
seen in Fi
g
ure 4.17
If Most Recent is
selected, the month day
and hour do not need to
be selected.
Select the ending hour
to view (applicable only
to 1 minute averages)
Select 1 min., 15 min.,
1 hour or 24 hour
averages.
Average Period Time Range
Displayed
1 Minute 1 Hour
15 Minutes 1 Day
1 Hour 3 Days
24 Hours 3 Months
Rosemount Analytical ยตCEM Continuous Analyzer Transmitter 4-2