Table 4.4 - [Stream X] section
Stream Setting Description
DiluentCorrectionPercent Diluent Correction Percent used in calculations for the Stream
O2R1Range Range 1 Setting for O2 (Range 2 can be changed from the Pocket PC)
COR1Range Range 1 Setting for CO (Range 2 can be changed from the Pocket PC)
NOXR1Range Range 1 Setting for NOx (Range 2 can be changed from the Pocket
PC)
PostCalibrationDelay Number of seconds to keep the C(Calibration) flag after the Auto
Calibration process is over
R1O2ZeroDriftLimit O2 Allowed Zero Drift
Limit for Range 1.
R1COZeroDriftLimit CO Allowed Zero Drift
Limit for Range 1.
R1NOXZeroDriftLimit NOx Allowed Zero Drift
Limit for Range 1.
R1OSMidDriftLimit O2 Allowed Mid Drift Limit
for Range 1.
R1COMidDriftLimit CO Allowed Mid Drift
Limit for Range 1.
R1NOXMidDriftLimit NOx Allowed Mid Drift
Limit for Range 1.
R1O2SpanDriftLimit O2 Allowed Span Drift
Limit for Range 1.
R1COSpanDriftLimit CO Allowed Span Drift
Limit for Range 1.
R1NOXSpanDriftLimit NOx Allowed Span Drift
Limit for Range 1.
If the drift exceeds the allowed amount
a drift alarm will occur, and the
readings on the channel will no longer
be valid until a successful calibration is
completed.
R2O2ZeroDriftLimit O2 Allowed Zero Drift
Limit for Range 2.
R2COZeroDriftLimit CO Allowed Zero Drift
Limit for Range 2.
R2NOXZeroDriftLimit NOx Allowed Zero Drift
Limit for Range 2.
R2OSMidDriftLimit O2 Allowed Mid Drift Limit
for Range 2.
R2COMidDriftLimit CO Allowed Mid Drift
Limit for Range 2.
R2NOXMidDriftLimit NOx Allowed Mid Drift
Limit for Range 2.
If the drift exceeds the allowed amount
a drift alarm will occur, and the
readings on the channel will no longer
be valid until a successful calibration is
completed.