Rosemount Analytical µCEM Continuous Analyzer Transmitter 4-19
4.3.7 µCEM -Manual Calibration
A dry-run Calibration may be initiated from the Manual Calibration page of the µCEM
Settings by pressing the Manual Calibrate All icon. A full zero and span calibration will
be run by the MicroCEM but the end result corrections of the calibration will not be
applied to the O2/Nox/CO measurement values. If desired a partial calibration may be
invoked for one or more of the emission types. While the manual calibration is in
process, a calibration progress dialog will be displayed as shown in Figure 4.24. When
the manual calibration is completed, the results are displayed in the Manual Calibration
Results dialog as shown in Figure 4.10. If the Local Calibration checkbox is checked,
the Local Calibration valve will be used during the calibration rather than the probe
Calibration valve.
Note that “Start Auto Cal now” will invoke a calibration and will apply new correction
factor to all measurement when done.
Figure 4.9 - Manual Calibration Menu