Specifications

STARTUP and OPERATION
Rosemount Analytical µCEM Continuous Analyzer Transmitter 4-10
4.2.7 Stream Switching Control
Typically a dual stream system is in Automatic Stream Switching mode. That means that
it runs the timing schedule specified in User Settings Configuration file. If Automatic
switching is not desirable, the user may turn it off using Tools-> Stop Auto Switching
menu. In this case the system will remain on the current stream indefinitely. When
Automatic switching is needed again, user may turn it back on with Tools->Start Auto
Switching menu.
The same task can be accomplished remotely, by clicking Stop Auto Switching button on
the µCEM Real-Time Web page.
Note, that this option is sustained even if the system is rebooted.
The operator may also force a switch between the streams at any time whether the
system is in Auto-Switching mode or not. Tools menu has an option “Switch to
StreamName”, where StreamName is a user-specified name of the stream.
The same task can be accomplished remotely by clicking Switch to “StreamName”
button on the µCEM Real-Time Web page.