Specifications

STARTUP and OPERATION
Rosemount Analytical µCEM Continuous Analyzer Transmitter 4-1
Installation,” the analyzer is ready for operation.
Before operating the system, verify that the Leak Checks have been
performed and that the sample handling unit is performing correctly.
MicroCEM analysis enclosure On/Off switch is located inside the door on
the bottom right hand corner. Push switch to the “on” position to start
system.
The unit will immediately run thru a self diagnostic mode. This may take up
to 2 minutes. The user will know the system has passed all diagnostic test
and is “ready” upon the green LED (located above on/off switch) flashing. If
the green LED does not start to flash verify that proper power is connected
to the unit and restart. If AC/Heater fan is running but the green LED still
will not flash then call the factory immediately for help.
NOTE: After startup a warm-up time from 20 to 60 minutes (Depending
upon ambient temp) is required for accurate measuements.
Analyzer operation can be confirmed by the green LED light flashing. The
pocket pc can then be connected for viewing menus. Upon power up, the
analyzer will perform a self-test routine. The test will take approximately 6
minutes.
4.2 Analyzer Operation
4.2.1 User Interface
The µCEM User Interface runs on a Pocket-PC with Windows CE operating system. It
communicates with the µCEM via serial communication port. All input to the Pocket-PC
is done using a pointing device that comes with the Pocket-PC. The Pocket PC can be
plugged into two different ports. The first port is located on the front panel below the
on/off switch inside the front door. The second port is from the bottom of the uCEM via
J3 connector.
The pocket PC can be found behind the door behind the glass piece. Note that upon
shipment the pocket PC may be located in a separate box.
To connect the pocket PC to the: µCEM via the inside connection.
1. Open µCEM door.
2. Plug RS232 plug into adapter located on front panel
3. Plug power supply cable into 5V adapter
4. Turn Pocket PC on
5. In order to assure no other windows are open press the reset button. Reset
button is located on the back of the pocket PC.
6. Go to tools menu (Icon in upper left hand corner) and click on µCEMTS.
7. Unit will display data in 3 to 5 seconds. If unit does not show data in 3 to 5
seconds repeat procedure starting with number 5.
To connect the pocket PC to the: µCEM via the outside connection.