Specifications
INSTALLATION
3.3.5.3 Troubleshooting Leaks
Liberally cover all fittings, seals, and other possible sources of leakage with a suitable
leak test liquid such as SNOOP
™
(part 837801). Bubbling or foaming indicates
leakage. Checking for bubbles will locate most leaks but could miss some, as some
areas are inaccessible to the application of SNOOP. For positive assurance that
system is leak free, perform one of the preceding tests.
NOTE:
Refer to Specification in Preface for maximum pressure limitations.
For differential measurement, the leak check must be performed for the
measurement and reference side separately.
For analyzers with parallel gas paths, the leak check must be performed
for each gas path separately.
Figure illustrates MicroCEM analysis enclosure (Left) wire connections to the Sample Handling box
™
Trademark of NUPRO Company
Rosemount Analytical µCEM Continuous Analyzer Transmitter 3–3