Specifications
INSTALLATION
3.3.4.3 SHU #1 / #2 Interface Connector (J6 & J7). These wires are to be connected directly to
the MicroCEM sample handling enclosure (SHU) and will control the operation of the
sample pump, drain pump, purge valve and calibration valve respetively. All toggle
switches in sample handling enclosure should be set to “remote” mode upon hookup of wire
so the MicroCEM analysis enclosure will control the full system.
The Digital Interface connector has a shell size of 14, 15 contacts. Each pin will accept a wire
size of 20 AWG. Connector and 6’ pigtail by Rosemount.
PIN NAME DESCRIPTION COLOR Sample Handling
Enc. Termination
1 SPUMP1/2NO BLK Not Used
2 SPUMP1/2C BRN 1
3 SPUMP1/2NC
Sample Pump #1/2 Control,
Dry contact, 110V 1A
RED 8
4 DPUMP1/2NO ORG Not Used
5 DPUMP1/2C YEL 1
6 DPUMP1/2NC
Drain Pump #1/2 Control,
Dry contact, 110V 1A
GRN 3
7 PURG1/2NO BLU 4
8 PURG1/2C VIO 1
9 PURG1/2NC
Purge Valve #1/2 Control,
Dry contact, 110V 1A
GRY Not Used
10 CAL1/2NO WHT 5
11 CAL1/2C WHT/BLK 1
12 CAL1/2NC
Calibration Valve #1/2
Control, Dry contact,
110V 1A
WHT/BRN Not Used
Internal Jumper terminals 2 and 9 set by Rosemount
Table 3-2. Sample Handling Unit Terminal Assignments
Rosemount Analytical µCEM Continuous Analyzer Transmitter 3–6