Instruction manual
Instruction Manual 
748468-E 
August 2003 
Rosemount Analytical Inc. A Division of Emerson Process Management  Maintenance and Service 4-3 
Model MicroCEM SHS 
WHERE 
USED 
COMPONENT 
MAINTENANCE 
ITEM 
DAILY WEEKLY MONTHLY
EVERY 
3 MO. 
EVERY 
6 MO. 
EVERY 
9 MO. 
EVERY 
12 MO. 
Packing Adjust-
ment 
   I,T  
F
2 
Regulating/Needle 
Valves/Toggle Valves 
Seat and Seals              I,R 
 Flow Regulators  Flexible Orifice          I,R 
RV
1 
Check/Relief Valves 
O-Ring Seal, 
Spring 
      I,R 
Seals           I,R 
Manual or Remote-
Operated Ball Valves - 2, 
3, 4, or 5- Way 
Packing Adjust-
ment 
   I,A  
EOV
1 
Plug Valves - 2,3,4, or 5-
Way, Manual or Remote-
Operated 
Plug Assembly       I,R 
Packless Valve (Dia-
phragm or Bellows 
Sealed), Manual or Re-
mote- Operated 
Seat and Seals              I,R 
  Quick Connect Fittings  O-Rings              I,R 
Coils        
Valve Internals 
(check for cross 
leakage) 
         I,R 
SV
1 
Solenoid Valves 
Valve Seats and 
Springs 
         I,R 
Air Operators         I,R 
EOV
1 
Valve Operators 
Motorized Op-
erators 
   I,R  
Legend: 
A – Tighten as required 
B – Blowback/Blowdown 
C – Clean 
I – Inspect (some disassembly may be required) 
R – Replace as required. 
T – Test for proper function by stimulating operating/alarm condition 
V – Visual check.   
Table 4-2. Component Maintenance Schedule - Valves 
IB-103-102