Instruction manual

Instruction Manual
748468-E
August 2003
4-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model MicroCEM SHS
WHERE
USED
COMPONENT
MAINTENANCE
ITEM
DAILY WEEKLY MONTHLY
EVERY
3 MO.
EVERY
6 MO.
EVERY
9 MO.
EVERY
12 MO.
SP
1
Primary sintered metal
sample filters
Filter Element B I,C or R
Filter Element I,C or R
Sintered Metal Sample
Guard Filters
Seal Kit I,R
Filter Element B I,R
Y-Strainers
Seal Kit I,R
Disposable End-of-Line
Filters
Element I,R
Filter Element R F
2
Sample Filters with Non-
Metallic Elements
Seal Kit I,R
Filter Element R
Sample Coalescing Fil-
ters with Non-Metallic
Elements
Seal Kit I,R
Filter Element I,R
Instrument Air Filters with
Non-Metallic Elements
Seal Kit I,R
Legend:
A – Tighten as required
B – Blowback/Blowdown
C – Clean
I – Inspect (some disassembly may be required)
R – Replace as required.
T – Test for proper function by stimulating operating/alarm condition
V – Visual check.
Table 4-1. Component Maintenance Schedule - Filters
IB-103-102