Instruction manual
Instruction Manual
748468-E
August 2003
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model MicroCEM SHS
5. Connect analog output signal as required
per Electrical Interconnect and Outline
and Mounting Dimensions Drawings.
Connect Status ID contacts and System
Alarm contacts as required.
6. Connect 115 VAC 60 Hz single phase
power (utility). Observe polarity when in-
stalling power - hot to black and neutral to
white. Incorrect connections may cause
damage to system. Provide grounding for
power as required by local codes for fi-
berglass enclosures. Verify all wiring is
per the Electrical Interconnect diagram.
7. Connect customer supplied zero and
span calibration gas cylinders and Rose-
mount supplied dual stage regulators as
required. Connect Instrument Air (per ISA
Standard S7.3).
NOTE
Refer to all NOTES on drawings located in
the rear of this for additional installation
information.
System should now be ready for start-up by
authorized personnel. Prior to start-up, the
operating instructions should be read and
thoroughly understood.
IB-103-102