Satellite Radio User Manual
Installation Manual 53
Output Wiring – Configurable I/O SpecificationsOutput Wiring – Fieldbus/PROFIBUSOutput Wiring – Intrinsically Safe
Chapter 7
Output Wiring – Model 2700 FOUNDATION fieldbus 
and PROFIBUS-PA Transmitters
7.1 Overview
This chapter explains how to wire outputs for Model 2700 transmitters with the F
OUNDATION fieldbus 
and PROFIBUS-PA output boards (output option code E, N, or G).
Note: If you don’t know what outputs option board is in your transmitter, see Section 1.4.
It is the user’s responsibility to verify that the specific installation meets the local and national safety 
requirements and electrical codes.
7.2 F
OUNDATION fieldbus wiring
Wire the transmitter to the fieldbus segment according to the diagram in Figure 7-1. Follow all local 
safety regulations. The transmitter is either FISCO or FNICO approved (see Section A.1.1). For 
FISCO-approved transmitters, a barrier is required. Refer to the F
OUNDATION fieldbus wiring 
specification.
Figure 7-1 Connecting the fieldbus communication wires 
Fieldbus 
power 
supply
F
OUNDATION fieldbus per FOUNDATION 
fieldbus wiring specification
Spur to fieldbus per
F
OUNDATION fieldbus
wiring specification
Terminals 3–6
Terminals 1–2
{
Note: Terminals 3 through 6 are not used.
Note: The fieldbus communication terminals (1 and 2) are polarity insensitive.