Operating instructions
STS2 Touch Screen Display 85
Figure 42 Source Set Points
2. Select 1 to configure the settings for Source 1.
3. Configure the settings using the keypad or dialog box that is displayed when you touch a
button.
4. Select 2 for Source 2.
5. Configure the settings, as described in step 3.
6. Select SAVE to keep the settings for both sources.
The Set Points dialog is closed and the User Settings dialog is displayed.
7. Select MANUAL MAX XFER PHASE ANGLE.
A keypad is displayed.
This setting configures the maximum allowed phase difference between the sources and applies
to all manual transfers as well as any auto retransfer.
a. Configure this setting with a range of +/- 1-30 degrees.
The default setting is +/-15 degrees and the resolution is 1 deg.
b. Select OK to save the setting.
The setting is now displayed in the adjacent field in the User Settings dialog box.
8. Select FREQUENCY TRIP POINT.
A keypad is displayed.
This choice verifies that the source is running at the proper frequency. This choice sets the fre-
quency deviation in the frequency that triggers the frequency trip point alarm. The frequency
Table 14 Set Point Parameters
Button Range Default Comments
Fast UV Setting (%) -10 to -30% of Nominal
voltage rating
-20% Set in increments of 1%.
Slow UV Setting (%) -5 to -20% -10% Set in increments of 1%.
Slow UV detection delay 1-60 cycles 5 cycles Units: Line Cycle
OV Setting (%) +5% to +20% +10% Set in increments of 1%.
OV Detection Delay 1-255 cycles 3 cycles Units: 1/4 Line Cycle
I-PK Xfer Lockout 0 – 3.0 multiplied by
the unit current rating.
2.5 * Unit
Current Rating
I-peak xfer lockout detection point.
Units: Amps RMS. Select the HI or
LO buttons to configure your setting.