Model IFT9701 Transmitter Instruction Manual 35
Configuration with Communicator TroubleshootingConfiguration with ProLink IIFlowmeter Startup
Chapter 6
Configuration with a HART Communicator
6.1 Configuration overview
You may use a Rosemount hand-held HART Communicator to set up the basic configuration and 
zero the flowmeter. The complete HART Communicator menu for the IFT9701 transmitter is 
provided in Appendix B; menus for specific operations are provided in this chapter.
Configuration parameters include such items as flow cutoff and damping values, flow direction, and 
milliamp output scaling. If requested at time of order, the meter is configured at the factory according 
to customer specifications. For factory-calibrated sensors, field calibration is not normally needed nor 
recommended. Basic configuration is described in Section 6.2.
Calibration accounts for performance variations in individual meters and peripheral devices. 
Calibration procedures include autozeroing and flow calibration. Zeroing is required upon initial 
flowmeter startup (see Section 5.3); flow calibration may or may not be required. For information on 
calibration procedures using a HART Communicator, see Section 6.3.
Characterization is the process of using a communication device to enter calibration factors for flow, 
density, and temperature directly into flowmeter memory, instead of performing field calibration 
procedures. Calibration factors can be found on the flowmeter serial number tag and on the certificate 
that is shipped with the meter. For information on characterization using a HART Communicator, see 
Section 6.4.
Set control devices for manual operation during flowmeter configuration.
CAUTION
Failure to set control devices for manual operation 
during configuration will affect flowmeter output 
signals.
Before using the basic setup menu, set process control 
devices for manual operation.
When the display depicted below appears, isolate the 
flowmeter from devices that use flowmeter outputs for 
process control, then press F4 (OK).