a. Connect the VDC wires from the core processor (see 
Figure 12) to an independent power supply. This power 
supply must connect only to the core processor. A 
recommended power supply is the SDN series of 24-VDC 
power supplies manufactured by Sola/Hevi-Duty.
b. Do not ground either connection of the power supply. 
c. Connect the RS-485 wires from the core processor (see 
Figure 12) to the RS-485 terminals at the direct host. Refer to 
the vendor documentation to identify the terminals.
d. Ensure that there is no electromagnetic interference coupled 
onto the cables between the core processor and the power 
supply or direct host. 
STEP 6. Grounding the sensor
The sensor must be grounded. The sensor may be grounded via the piping 
if the joints in the pipeline are ground-bonded. If the sensor is not 
grounded via the piping, connect a ground wire to the sensor ground 
screw.
If using an integral booster amplifier, it is grounded with the sensor. If 
using a remote booster amplifier, it must be grounded separately, using 
the internal or external ground screw as required.
CAUTION
Improper grounding could cause measurement error.
To reduce the risk of measurement error:
• Ground the flowmeter to earth, or follow ground 
network requirements for the facility.
• For hazardous area installations in Europe, refer to 
standard EN 60079-14 if national standards do not 
apply.