Instruction manual
5
STEP 1. Choosing a location
Choose a location for the sensor based on the requirements described in 
this section. 
General guidelines
The following conditions must be met:
• Before operation, you must be able to stop flow through the sensor. 
(During the zeroing procedure, flow must be stopped completely, and 
the sensor must be full of process fluid.)
• During operation, the sensor must remain full of process fluid.
• The sensor must be installed in an area that is compatible with the 
classification specified on the sensor approvals tag (see Figure 1 and 
Figure 4).
Hazardous area installations
Make sure the hazardous areas specified on the approvals tags are suitable 
for the environment in which the sensor and booster amplifier will be 
installed. See Figures 1 and 4. For installation in an area that requires 
intrinsic safety, refer to Micro Motion UL, CSA, or ATEX 
documentation, shipped with the sensor or available from the Micro 
Motion web site.
For a complete list of hazardous area classifications for Micro Motion 
sensors, refer to the EXPERT
2
™
 system at www.expert2.com.
Environmental limits
Temperature limits for the D600 sensor and booster amplifier are listed in 
Table 2.
Table 2. D600 temperature limits
Component Temperature limits
Sensor with integral booster amplifier
(1)
(1) Limits apply to process fluid temperature.
–58 to +140  °F (–50 to +60  °C)
Sensor with remote booster amplifier
(1)
–400 to +400  °F (–240 to +204  °C)
Remote booster amplifier
(2)
(2) Limits apply to ambient temperature.
–40 to +140  °F (–40 to +60  °C)