 Model D and DT Sensors Instruction Manual 29
Wiring continued
Remote booster amplifier drive wiring
Wiring to a transmitter (D600 
sensor with junction box)
The instructions in this section explain how to connect a fully prepared 
9-wire Micro Motion flowmeter cable to the sensor and transmitter.
• The procedure for preparing Micro Motion cable and cable glands is 
described in the instructions that are shipped with the cable.
• Install cable and wiring to meet local code requirements.
Cable connections to sensor and transmitter
The wiring procedure is the same for the sensor and transmitter. Refer to 
the wiring diagrams on the following pages, and follow these steps:
Factory-supplied 9-wire
cable for intrinsically
safe wiring
(RTD and pickoffs)
Wiring to 
transmitter
Remove screw and terminal cover 
before installing wiring. Re-install 
cover before operating
Connection diagram
From remote 
booster amp 
terminal
To sensor 
explosion-proof 
J-box terminal
11
22
CAUTION
Failure to seal the sensor and transmitter housings 
could cause a short circuit, which would result in 
measurement error or flowmeter failure.
• Ensure integrity of gaskets and O-rings.
• Grease all O-rings before sealing.
• Install drip legs in cable or conduit.
• Seal all conduit openings.