Instruction manual
Micro Motion
®
 Model D and DT Sensors Instruction Manual 27
Wiring continued
Power supply wiring to the 
integral booster amplifier
Provide 85-250 VAC power to terminals L2 and L1 as shown in the 
diagram below.
This unit is provided with an external terminal for supplementary 
bonding connections. This terminal is for use where local codes or 
authorities permit or require such connections.
Integral booster amplifier power-supply wiring
85-250 VAC
50/60 Hz
N/L2 L/L1
Power supply
ground
Wiring
compartment
Supplementary bonding connection