Instruction manual
Micro Motion
®
 Model D and DT Sensors Instruction Manual 21
Wiring continued
Model DT sensor cable
Connecting and shielding 9-wire 
cable
A 9-wire connection is required between the junction box and the core 
processor or transmitter. Micro Motion offers two types of 9-wire cable:
• Shielded
•Armored
Both cable types contain shield drain wires. You may also use jacketed 
cable with conduit.
Cable connections to sensor and transmitter
The wiring procedure is the same for the sensor and transmitter. Refer to 
the wiring diagrams on the following pages, and follow these steps:
Drip leg
1/2" NPT
conduit fitting
3 ft. (1 m) factory-supplied flex conduit
• Liquid tight to meet CE requirements 
for European installations
• Permanently attached to sensor
Grounding screw
1/2" NPT conduit fitting
• Factory-supplied fitting
• Ensure 360
° contact
CAUTION
Failure to seal the sensor and transmitter housings 
could cause a short circuit, which would result in 
measurement error or flowmeter failure.
• Ensure integrity of gaskets and O-rings.
• Grease all O-rings before sealing.
• Install drip legs in cable or conduit.
• Seal all conduit openings.