Rosemount Analytical Inc. A Division of Emerson Process Management  Appendix A. Menu Displays 7-23 
Model NGA2000 PMD 
Current measurement parameters help
Shows the main measurement parameters.
These can be controlled in the various 
set up menus. 
HOME 
ESCAPE INFO
Barometric pressure parameters
Pressure transducer:
Barometric pressure compensation: 
Measured pressure: 
Transducer offset: 
Transducer slope: 
Transducer PGA gain: 
000.0
000.0
000.0
000.0
000.0
000.0
HOME 
ESCAPE INFO
Barometric pressure parameters
Shows whether the optional pressure transducer  
is installed or not. Also this screen allows you to 
set the transducer. Use the offset at atmospheric 
pressure, and pull a vacuum of about 25cm H2O  
on the transducer, and use the slope to make it read 
correctly. 
Choose whether to enable barometric compensation, 
and whether to allow the analyzer to emit the 
pressure value. Only one analyzer at a time in a 