Instruction Manual HASXEE-IM-HS 10/2010 Gas Analyzer Series Instruction Manual www.EmersonProcess.
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications.
HASXEE-IM-HS 10/2010 X-STREAM XE Table of contents Instruction manual SHORT FORM GUIDE FOR THIS MANUAL see chapter TOC To find information about Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . S The different instruments designs . . . . . . . . . . . . . . . 1 The instruments technical data . . . . . . . . . . . . . . . . . 2 Measuring principles characteristics . . . . . . . . . . . .
Instruction manual X-STREAM XE HASXEE-IM-HS 10/2010 Table of Contents Introduction S-1 Definitions S-1 Safety instructions S-5 Chapter 1 Technical description 1-1 Terms used in this instruction manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-2 Symbols used on and inside the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-3 Symbols used in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . .
Instruction manual Chapter 2 Technical data 2-1 Chapter 3 Measuring Principles 3-1 Chapter 4 Installation 4-1 Chapter 5 Startup 5-1 2.1 Common Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 2.2 Model-Specific Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 2.2.1 X-STREAM XEGC: ½ 19 Inch Tabletop Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 2.
Instruction manual X-STREAM XE HASXEE-IM-HS 10/2010 Table of Contents 5.3.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 5.3.2 Status Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 5.3.3 Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4 5.4 Software . . . .
Instruction manual 7.4.8 Restoring a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40 7.4.7 Cancelling an Ongoing Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40 7.5 Replacing the Electrochemical Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-41 7.5.1 General Hints on Handling the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Instruction manual X-STREAM XE HASXEE-IM-HS 10/2010 Table of Contents Chapter 9 Modbus Functions 9-1 9.1 Abstract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1 9.1.1 Modbus TCP/IP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1 9.2 Supported Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Instruction manual Tab. 3-1: Tab. 3-2: Tab. 3-3: Tab. 3-4: Tab. 3-5: Tab. 3-6: Tab. 3-7: Tab. 3-8: Tab. 3-9: Standard paramagnetic sensor -cross interference by accompanying gases.............. 3-6 Solvent resistant paramagnetic sensor - approved solvents.....................................3-7 Solvent resistant paramagnetic sensor - medium affected materials .......................3-7 Solvent resistant paramagnetic sensor -cross interference by gases..........................
Instruction manual X-STREAM XE HASXEE-IM-HS 10/2010 Index of Figures Fig. 1-1: Fig. 1-2: Fig. 1-3: Fig. 1-4: Fig. 1-5: Fig. 1-6: Fig. 1-7: Fig. 1-8: Fig. 1-9: Fig. 1-10: Fig. 1-11: Fig. 1-12: X-STREAM front panel (here the X-STREAM XEGP).............................................1-3 Optional heated area...............................................................................................1-8 Gas flow diagram: single channel or in series.........................................................
Instruction manual Fig. 3-8: Wheatstone Bridge................................................................................................3-11 Fig. 3-9: TC cell, exterior view ,thermal isolation removed..................................................3-12 Fig. 3-10: TC cell, sectional view...........................................................................................3-12 Fig. 4-1: Fig. 4-2: Fig. 4-3: Fig. 4-4: Fig. 4-5: Fig. 4-6: Fig. 4-7: Fig. 4-8: Fig. 4-9: Fig. 4-10: Fig. 4-11: Fig.
Instruction manual X-STREAM XE HASXEE-IM-HS 10/2010 Index of Figures Fig. 6-1: Fig. 6-2: Fig. 6-3: Fig. 6-4: Fig. 6-5: X-STREAM XE Software menu structure................................................................6-2 Measurement display elements.............................................................................6-25 Usage of labels and tags ......................................................................................6-31 Measurement display with labels and tags (example)..........
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 The manual covers several X-STREAM analyzer models and so may contain information about configurations and/or options not applicable to your analyzer. The installation and operation of units for use in explosive (hazardous) environments is not covered in this manual. Analyzers intended to be used in such environments are supplied with further instruction manuals, which should be consulted in addition to this.
X-STREAM XE Instruction Manual HASXEE-IM-HS 04/2010 TERMS USED IN THIS INSTRUCTION MANUAL Explosive Gas(es) Flammable Gases and gas mixtures in a mixture with air within the explosive limits. Flammable Gas(es) Gases and gas mixtures are assigned to be flammable if they might become ignitable when in a mixture with air.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 ... means dangerous voltages may be exposed. Covers may only be removed when power to the unit is disconnected, and only be trained personnel. hot surfaces may be exposed. Covers may only be removed by trained personnel and when power is disconnected. Certain surfaces may remain hot. further information and instructions are required: read the instruction manual. Emerson Process Management GmbH & Co. OHG S-3 S This symbol attached to the unit ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 SYMBOLS USED IN THIS MANUAL Wherever one or more of the following symbols are used in this instruction manual, read the accompanying information and instructions carefully. Follow these warnings and notes carefully to minimize risk. This symbol used in the manual ... ...
Instruction Manual INTENDED USE STATEMENT X-STREAM series gas analyzers are intended to be used as analyzers for industrial purposes. They must not be used in medical, diagnostic or life support applications nor as safety devices. Using X-STREAM XE analyzers as safety devices, requiring redundant design or SIL classification, is also not permitted.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Safety Instructions AUTHORIZED PERSONNEL In-depth specialist knowledge is an absolutely necessary condition for working with and on the analyzer. Authorized personnel for installing, operating, servicing and maintaining the analyzer are instructed and trained qualified personnel of the operating company and the manufacturer.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1. Suitable grounding connections should be made at all connectors provided for this purpose. 2. All safety covers and grounding connections must be properly reinstated after maintenance work or troubleshooting. 3. A fuse should be provided at the installation site which will completely disconnect the unit in case of failure. Installing an isolating switch may also be beneficial.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Safety Instructions EXPLOSION HAZARD The units described in this manual may not be used in explosive atmospheres without additional safety measures. ELECTRICAL SHOCK HAZARD Do not operate without covers secure. Do not open while energized. Installation requires access to live parts which can cause death or serious injury. For safety and proper performace this instrument must be connected to a properly grounded three-wire source of power.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 The models intended for outside and wall mounted use (X-STREAM XEF, XDF and XEFD) weigh between 26 kg (57 lb) and 63 kg (139 lb) depending on options installed. Two people and/or lifting equipment is required to lift and carry these units.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Safety Instructions GASES AND PREPARATION OF GASES GASES HAZARDOUS TO HEALTH Follow the safety precautions for all gases (sample and span gases) and gas cylinders. Before opening the gas lines, they must be purged with air or neutral gas (N2) to avoid danger from escaping toxic, flammable, exposive or hazardous gases.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 CONNECTING UNITS FOR PERMANENT INSTALLATION Only qualified personnel following all applicable and legal regulations may install the unit and connect it to power and signal cables. Failure to comply may invalidate the unit’s warranty and cause exposure to the risk of damage, injury or death. This unit may only be installed by qualified personnel familiar with the possible risks.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 General Operational Notes GENERAL OPERATING NOTES EXPLOSION HAZARD Exhaust gases may contain hydrocarbons and other toxic gases such as carbon monoxide. Carbon monoxide is toxic. Faulty gas connections may lead to explosion and death. Ensure that all gas connections are connected as labelled and airtight. • The unit must be installed in a clean and dry area protected from strong vibrations and frost.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Emerson Process Management GmbH & Co. OHG Modified resistant measuring cells are available for use with corrosive gases and/or gases containing solvents. Special configurations (e.g. intrinsically safe or infallible measuring cells) for the analysis of combustible gases are also available. Chapter 3 gives a detailed description of the various measuring techniques.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1 Technical Description • • Non-incendive assembly for installation in Zone 2 and Div 2 for the measurement of non-flammable gases. Simplified purge system (Z-purge) for installation in North American Div 2 environments. The cast aluminium field housing is designed to withstand an explosion and intended to be used in hazardous areas of Zone 1. More information about EXanalyzers can be obtained from your Emerson Process Management sales office.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.1 Overview All X-STREAM gas analyzers feature an easy-to-use graphical user interface, which displays measurement values, status and error messages, and menus for the input of parameters. For ease of use, the operator can select one of three languages for the display: By default any analyzer is configured with English and The Front Panel 1 1.1.1 German language sets, while a third can optionally be added.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.1 Overview Further analyzer features include (some of them optional): • Configurable measurement display: • gas measurements results and/or secondary results (e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.2 Configuration of Gas Lines 1.2.1 Materials Used The physical and chemical properties of the sample gas as well as the conditions under which measurement takes place influence the choice of materials. Among those available are Viton®, PFA and stainless steel. 1.2.2 Safety Filter The analyzers are generally fitted with an internal stainless-steel filter.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.2 Configuration of Gas Lines 1.2.6 Optional Components for Gas Lines The analyzers can, as an option, be fitted with further components. Not all components are available for all analyzer types: • internal sample gas pump • internal valve block • internal flow sensors • internal flow monitor switch • internal barometric pressure sensor • internal temperature sensors. 1.2.6.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.2 Configuration of Gas Lines Internal Barometric Pressure Sensor sate the atmospheric pressure variation ( measurement specification, page 3-14). If such a sensor is installed in the unit, the related menu shows the entry Internal ( 6-91).
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.2 Configuration of Gas Lines 1.2.6.7 Optional Heated Area The physical components can be optionally separated from the electrical components by means of a special box (not an option for ½ 19 in units). This can be done for one or both of the following purposes: Firstly, the box allows the physical components to be regulated to a temperature of approx.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.2 Configuration of Gas Lines 1.2.7 Alternative Configurations 1 Technical Description Depending on the application and the selected analyzer options, alternative gas line configurations are available, exemplified in the following diagram of a dual-channel analyzer: Fig. 1-3: Gas Flow Diagram: Single Channel Or in Series Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.3 Interfaces 1.3 Interfaces All analyzer types are fitted with one analog electrical output for each channel, four status relays, 2 Ethernet interfaces and a serial service interface as standard. As an option, further interfaces can be added. 1.3.1 Analog Outputs By default each X-STREAM analyzer is fitted with one output per channel, which can transmit data on concentration levels to an external data acquisition system.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 All supported Modbus parameters are listed in chapter 9. Information about web-browser access is provided in chapter 7 1.3.4 Fig. 1-4: Ethernet Interface Marking Note! All analyzers provide 2 Ethernet connectors: the upper marked "2" is configured to be used for network connection.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.3 Interfaces 1.3.6 1.3.6.1 Optional Interfaces Analog Inputs Two d. c. analog inputs enable connection to external devices. Their signals can be used for e.g. • cross compensation • pressure compensation, or • handled as a separate measurement channels. 1.3.6.2 Digital Outputs In addition to the 4 default digital outputs, analyzers can optionally be upgraded with 9 or 18 more digital outputs ( 1-13), to be used for various purposes, e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.4 Comparison of Analyzer Models 1.4 Comparison of the Various X-STREAM Analyzer Models X-STREAM XEGP X-STREAM XEGC 1 ⁄2 19 in housing, table-top or rackmountable, protection type: IP 20 1 24V input with external power supply unit Max. 3 channels: 2 photometer + 1 WLD/O2 max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.4 Comparison of Analyzer Models X-STREAM XEFD 1 Technical Description X-STREAM XEF / XDF Stainless steel wallmountable field housing, protection type: IP66 / NEMA 4X Single or dual compartment design Internal wide range power supply unit Max. 4 channels in any combination max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.5 X-STREAM XEGC Any free tube fittings can be used for purging the device to minimize interference from the ambient atmosphere, or when measuring corrosive and/or flammable gases. 1 This compact model for general purposes can be fitted with up to three photometer measurement channels (IR or UV) and one additional non-photometer channel (eO2, pO2 or WLD). Power is supplied by a separate external power supply unit.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.5 X-STREAM XEGC 1 1 2 3 4 5 6 Fig. 1-7: 1-16 2 3 4 5 Status line Graphic display Messages line home key enter key 4 keys for adjustment and menu selection 6 7 7 8 9 10 8 9 10 Signal connectors (some optional) Gas fittings Power input fuse Power input connector X-STREAM XEGC - Views Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.6 X-STREAM XEGP Connection to power supply Main power is supplied via the IEC chassis plug mounted on the rear panel, with integrated power switch and fuse holders. The internal wide range power supply unit enables the analyzers to be used worldwide.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Front view 1.6 X-STREAM XEGP 1 Status line Graphic display Messages line 2 3 4 5 4 5 6 6 home key enter key 4 keys for adjustment and menu selection Rear view 1 2 3 1 2 3 5 1 Gas connector fittings Space for additional fittings Optional purge gas inlet 2 4 3 Power inlet with filter, fuses & switch Signal input/output connectors (some optional) 4 5 Optional screw-type terminal adapter Strain-reliefs, top view details 1 1 Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.7 X-STREAM XE Field Housing Front panel The analyzer’s display is covered by an impact tested glass for enhanced protection against breakage in harsh environments. Electrical connections Electrical connections are provided via internal tube fittings, the cables being fed through cable glands on the underside of the unit (fig. 1-8). The front cover of the housing swings open to the left once the fasteners have been released.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.7 X-STREAM XE Field Housing Fig. 1-9: X-STREAM XEF / XDF- Front Views HEAVY INSTRUMENT X-STREAM field housings, intended for outside and wall mounted use, weigh approx. (XEF) 26 kg (57 lb) or (XDF) 45 kg (99 lb), depending on options installed. Two people and/or lifting equipment is required to lift and carry these units.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1 Technical Description 1.7 X-STREAM XE Field Housing 1 2 1 2 3 4 5 3 Cable gland for power cable Cable glands for signal cables 4 brackets for wall-mounting Gas in- & outlets (max. 8) Cutouts, to combine 2 housings (here closed) 5 4 5 Note! In case of XDF, the cable glands are located at the upper compartment, while the gas in- & outlets are at the bottom side of the lower compartment. Also only 2 brackets are at each compartment. Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.7 X-STREAM XE Field Housing 1 5 2 (shown with front panel removed) 1 2 3 4 5 Screw-type terminals for signal cables Power line filter Cable glands Power supply terminals with integrated fuses Ethernet and USB connectors 4 3 Note! In case of XDF, the terminals and connectors are located at the upper compartment, while physical components and gas fittings are in the lower compartment. Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.7 Field Housings in Hazardous Areas EXPLOSION HAZARD X-STREAM XEF and XDF models CAN NOT be used in explosive environments (hazardous areas) without additional safety features. This instruction manual does NOT describe the special conditions necessary to operate gas analyzers in hazardous areas. Please refer to the separate instruction manual supplied with units for use in hazardous areas.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.8 X-STREAM XEFD 1.8 X-STREAM XEFD: Cast Aluminum Flameproof Housing The most obvious X-STREAM XEFD analyzer feature is its flameproof housing (fig. 1-10). This enables its use in Ex-zone 1 hazardous environments. With a protection type of IP66/ NEMA Type 4X and sturdy cast aluminum housing designed for wall-mounting, it can also be used in other tough environments. IPx6: In case of occasional flooding, e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 situation special conditions apply for operation in hazardous areas, described in the separate manual addendum for hazardous areas. For further information on operation in general purpose environments, see 1-5. 1 gases. The assignments of the connectors is given on an adhesive label situated near the connectors.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1.8 X-STREAM XEFD 3 2 1 2 3 3 1 Cable inlets for power and signal cables Gas tube fittings, protected by flame arrestors 4 brackets for wall mounting Fig. 1-13: X-STREAM XEFD - Bottom View HEAVY INSTRUMENT The model X-STREAM XEFD, intended for outside and wall mounted use, weighs approx. 63 kg (139 lb), depending on options installed. Two people and/or lifting equipment is required to lift and carry these units.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1 Technical Description 1.8 X-STREAM XEFD (shown with front panel removed) 5 1 2 3 4 5 1 2 3 4 Terminals for signal cables (shown fully populated) Power line filter Cable inlets for power and signal cables Power supply terminals with integrated fuses Ethernet and USB connectors Fig. 1-14: X-STREAM XEFD - Terminals Emerson Process Management GmbH & Co.
X-STREAM XE 1-28 Instruction Manual HASXEE-IM-HS 04/2010 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Chapter 2 Technical Data page 2-2 X-STREAM XEGC page 2-6 X-STREAM XEGP page 2-12 X-STREAM XEF, XDF page 2-15 X-STREAM XEFD page 2-19 Emerson Process Management GmbH & Co. OHG 2 Common technical data Technical Data This chapter contains all the technical details of the analyzers, divided into common and model-specific data.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.1 Common Technical Data 2.1 Common Technical Data Site of installation Humidity (non-condensing) Degree of pollution Installation category Elevation Ambient atmosphere < 90 % RH at +20 °C (68 °F) < 70 % RH at +40 °C (104 °F) 2 II 0 to 2000 m (6560 ft) above sea level Units may not be operated in corrosive, flammable or explosive environments (except flameproof XEFD) without additional safety measures.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.1 Common Technical Data Interfaces, signal inputs and outputs standard: optional: subminD plugs and sockets screw-type terminal adapters X-STREAM XEF, XDF, XEFD: internal screw-type terminals All versions provide a 2 RJ45 plugs for Ethernet connections as well as 1 USB and 1 mini USB connector (field housings internally only).
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.1 Common Technical Data Optional interfaces for all models Digital I/O board electrical specification 7 or 14 digital inputs (X-STREAM XEGC: max. 7 inputs) function electrical specification 9 or 18 additional relay outputs (X-STREAM XEGC: max. 9 add. outputs) 2-4 function max. 30 V, internally limited to 2.3 mA HIGH: min. 4 V; LOW: max. 3 V common GND Each input can be configured to one of the following functions: Open valve V1 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.1 Common Technical Data Optional interfaces for all models Analog I/O board 2 analog inputs function protected against overload up to ± 15 V or ± 20 mA Input analog signals from external devices, such as e.g. pressure transmitters, flow sensors, analyzers, etc.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.1 Model-Specific Technical Data: X-STREAM XEGC 2.2 Model-Specific Technical Data 2.2.1 X-STREAM XEGC: ½ 19 Inch Tabletop Unit All dimensions in mm [in] Fig. 2-1: X-STREAM XEGC - Dimensions 1 2 3 Fig. 2-2: 2-6 Dimensional drawings show tabletop version with strain-reliefs and screw-type terminals.
Instruction Manual HASXEE-IM-HS 04/2010 X-STREAM XE 2.2.1 Model-Specific Technical Data: X-STREAM XEGC 2 Technical Data Analyzer with blindplate to the left Analyzer with blindplate to the right Two analyzers in a rack All dimensions in mm [in] Fig. 2-3: X-STREAM XEGC - Rack Mounting Options Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.1 Model-Specific Technical Data: X-STREAM XEGC Temperatures operational storage Weight, max IP or Type rating Gas connections max number max for purging (incl. / separate) material sizes Power supply unit Power supply nominal voltage voltage range nominal input current, max Power input fuses Electrical in- and outputs power signals optional special 2-8 0 ... +50 °C / 32 ... 122 °F -20 ... +70 °C / -4 ... -158 °F 8 ... 12 kg / 17.6 ... 26.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.1 Model-Specific Technical Data: X-STREAM XEGC 2.2.1.1 Data for Optional External Power Supply Units Model UPS 01 T PE 57.7 54.2 This PSU can be ordered as an option for supplying power to one tabletop unit. Technical Data All dimensions in mm [in] 105.9 IEC power input socket 275.3 (shielding) 2 3 2 Pin 1: ME Pin 2: + 24 V Pin 3: 0 V (⊥) shield: housing flange 1 Pin configuration for 24 V DC output socket Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.1 Model-Specific Technical Data: X-STREAM XEGC This PSU can optionally be ordered for rack installation. Two variations are available: • with blind front panel, connectors at the rear side • with rear panel, connectors to the front. Both variations are fixed to the rack by means of screws at the panels. Fig. 2-5: 2-10 UPS 01 Power Supply Unit for Rack Installation Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.1 Model-Specific Technical Data: X-STREAM XEGC Model 10 A tabletop PSU This PSU can optionally be ordered for powering 2 tabletop units with a single common power supply. PE Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.2 Model-Specific Technical Data: X-STREAM XEGP 2.2.2 X-STREAM XEGP: 19 Inch Tabletop and Rack-Mount Models approx. values in mm [in] Strain relief bracket, detail (model with clamping adapters) Fig. 2-7: 2-12 X-STREAM XEGP - Dimensions Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.2 Model-Specific Technical Data: X-STREAM XEGP Gas connections max number max for purging (incl. / separate) material sizes Power supply unit Power supply nominal voltage voltage range nominal input current, max standard, max w/ temperature control, max Power input fuses Electrical in- and outputs power signals optional special Emerson Process Management GmbH & Co. OHG Technical Data IP or Type rating 0 ... +50 °C / 32 ... 122 °F -20 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.2 Model-Specific Technical Data: X-STREAM XEGP 1 2 Fig. 2-8: 1 Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.3 Model-Specific Technical Data: X-STREAM XE Field Housing 2.2.3 X-STREAM XEF/XDF: Single/Dual Compartment Field Housing Connector for potential equalization 2 Cable glands Technical Data Gas fittings All dimensions in mm [inches in brackets] Fig. 2-10: X-STREAM XEF - Dimensions Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.3 Model-Specific Technical Data: X-STREAM XE Field Housing All dimensions in mm [inches in brackets] Gas fittings Connector for potential equalization Cable glands Fig. 2-11: X-STREAM XDF - Dimensions 2-16 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.3 Model-Specific Technical Data: X-STREAM XE Field Housing Gas connections max number max for purging (incl.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.3 Model-Specific Technical Data: X-STREAM XE Field Housing 1 2 1 2 3 Power supply terminals with fuse holders Grounded conductor clamp (PE) Power supply cable entry 3 Fig. 2-12: X-STREAM XE Field Housing - Power Supply Terminals / Fuse Holders 1 1 2 3 2 Ethernet & USB (CAN projected) Analog & digital I/O terminal strips Max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.4 Model-Specific Technical Data: X-STREAM XEFD 2.2.4 X-STREAM XEFD: Flameproof Housing 2 Technical Data Transport lugs to be removed after installation Flame arrestors with gas fittings (enclosure threads: M18 x 1.5) Cable inlets (enclosure threads; M20 x 1.5) Eyebolt detail All dimensions in mm [inches in brackets] Fig. 2-14: X-STREAM XEFD - Dimensions Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.2.4 Model-Specific Technical Data: X-STREAM XEFD Temperatures operational storage Weight, max IP or Type rating Gas connections max number max for purging (incl.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 1 2 3 4 Power terminals with integrated fuse holders Protective earth terminal (PE) Power cable entry EMI power supply filter Technical Data 2.2.4 Model-Specific Technical Data: X-STREAM XEFD 1 4 3 2 2 Fig. 2-15: X-STREAM XEFD - Power Supply Terminals / Fuse Holders 1 1 2 3 Ethernet & USB (CAN projected) Analog & digital I/O terminal strips Max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 2.3 Information on Name Plate 2.3 Information on Name Plate The name plate provides details on the configuration of the unit, installed measuring techniques, sample gases and measuring ranges. It also indicates the unit’s serial number. The plate is located on either the side or the rear of the unit. Note! Analyzers configured to be installed in hazardous areas have special name plates, described in the associated manuals.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.1 Infrared Measurement (IR) Ultraviolet Measurement (UV) The non-dispersive measurement methods described in this section utilize gas specific light absorption in order to discriminate between different gases. This is possible, as any gas possesses distinct absorption characteristics.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.1 Infrared (IR) and Ultraviolet (UV) Measurement NDIR Detector Absorption chamber absolute flow, detected by the micro flow sensor, in both cases is therefore a measure for the light absorbed while the chopper is open. This directly correlates to the amount of light not absorbed in the analysis cell and therefore to the concentration of the measurement gas inside the analysis cell.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.1 Infrared (IR) and Ultraviolet (UV) Measurement 3.1.3 Technical Implementation pyrodetectors, limiting the bandwidth of radiation passing the chopper. The radiation emitted by an IR or UV source passes an adaptor cell, widening the beam to completely fill out the analysis cell´s diameter. At the opposite side of the cell, another adaptor cell is installed to reduce the beam to the diameter of the opening in the chopper.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.2 Oxygen Measurement 3.2 Oxygen Measurement Two different principles are used for measuring oxygen concentrations. The principle used in your specific instrument is given by the channel code (sample gas designator) on the nameplate label ( figure on page 2-22): pO2 = paramagnetical sensor eO2 = electrochemical sensor The paramagnetic oxygen detector also contains a temperature sensor and a heating element to hold the detector at approx. 55 °C.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.2 Oxygen Measurement 3.2.1.1 Cross Interferences by Accompanying Gases The table below shows, how accompanying gases interfere the paramagnetical oxygen measurement. If the concentration of such gases is already given at time of enquiry, this interference may be taken into account during factory startup and thus minimized (option). 100 % Gas Acetylene Allene Ammonia Argon Bromine 1.2-Butadiene 1.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.2 Oxygen Measurement Applications With Corrosive Or Solvent Components Approved solvents (inclusive accompanying disturbing components) Component Acetic acid Acetone Acrolein Aromatics Butadiene Butadiene-1 Butadiene-2 C2H2 C4H8 C5 C6H12 CH3COOH Cyclohexane Cyclohexanon Dimethyl sulfide Ethanol Ethene Ethylene Ethylene oxid Concentration Component Concentration 0.1- 20 % Heptane 0.1- 20 % 0.1- 20 % Hexane 0.1- 20 % 0.1- 20 % Isopropanol 0.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.2 Oxygen Measurement 3.2.2 Electrochemical Measurement (I) This sensor utilizes the principle of galvanic cells, fig. 3-5 shows the design. 10 11 6 9 11 3 5 1 7 13 2 4 8 1 2 3 4 5 6 7 8 9 10 11 12 13 Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.2 Oxygen Measurement (black) (11) Thermistor (5) Resistance (6) (-) (+) Gold Lead cathode (2) anode (1) Principles (red) Electrolyte (3) Electrochemical reaction: Electrochemical Reaction of Oxygen Sensor 3 Fig. 3-7: In consequence of its design the sensor‘s lifetime is limited and depends on theoretical designed life and oxygen concentration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.2 Oxygen Measurement 3.2.3 Special Hints Due to the measuring principle the electrochemical oxygen cell requires a minimum internal consumption of oxygen (residual humidity avoids drying of the cell). Supplying cells continuously with dry sample gas of low grade oxygen concentration or with sample gas free of oxygen could result in a reversible detuning of O2 sensitivity. The output signal will become unstable, but response time remains constant.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.3 Thermal Conductivity Measurement 3.3.2 Technical Implementation A block made of stainless steel contains two gas paths. Both, the volume of the block and the mass of the sensors have been minimized in order to obtain short response times. To suppress influences by changing ambient temperature the block is thermostatted and isolated against ambience. The sensors are fully glass packaged to withstand aggressive gases. 1 2 3 4 1 2 3 4 5 Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.6 Measurement Specification 3.4 Measurement Specification Sample gas components and measuring ranges (standard configurations) Note! The following table gives an overview only. Not all data is applicable to all analyzer variations. The sample gas(es) and measuring ranges for your specific analyzer are given by the order acknowledgement and on the analyzer‘s name plate label.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.6 Measurement Specification General Performance Specifications NDIR/UV/VIS Oxygen Sensor (PO2 and EO2) Thermal Conductivity Detection limit <1%1 4 <1%1 4 <2%1 4 Linearity <1%1 4 <1%1 4 <1%1 4 < 2 % per week Zero-point drift < 0.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 3.6 Measurement Specification Special Performance Specifications for Gas Purity Measurements (ULCO & ULCO2) Noise <1%1 Detection limit <2% Linearity < +1 % 1 < +2 % 1 Span (sensitivity) drift < +2 % Repeatability Response time (t90) < +2 % 1 2 3 < +2 % 1 2 4 < 10 s resp. < 0.2 ppm 2 3 1 2 4 2 9 1 2 9 2 resp. < 0.2 ppm 1 resp. < 0.2 ppm < 10 s 2 3 9 1 2 4 9 < +2 % 1 2 resp. < 0.2 ppm 1 2 7 < +2% 1 2 9 7 0.2 - 1.5 l/min.
X-STREAM XE 3-16 Instruction Manual HASXEE-IM-HS 04/2010 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Chapter 4 Installation This chapter describes the correct installation procedure for the various X-STREAM analyzer versions. On receipt, check the packaging and its contents thoroughly for damage. Inform the carrier immediately of any damage to packaging or contents. Keep the packaging in a safe place until unit is installed.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.2 Installation - Introduction 4.2 Introduction ELECTRICAL SHOCK HAZARD Before connecting the analyzer to mains power, please read the chapter on safety warnings and the following instructions carefully. The place of installation must be clean, dry and protected against strong vibrations and frost. Please observe the admissable operating temperatures given in the technical data. Units must not be subjected to direct sunlight or sources of heat.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Purging option The purging medium (e.g. to minimize CO2 interference or for enhanced safety when measuring corrosive or poisonous gases) must be dry, clean and free of corrosives or components containing solvents. To minimize cross interferences the purge gas also has to be free of components to be measured. Its temperature must correspond to the ambient temperature of the analyzer, but be at least within the range 20–35 °C (68–95 °F).
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.3 Installation - Gas Preparation TOXIC GAS HAZARDS Take care that all external gas pipes are connected in the described way and that they are gastight to avoid leakages! Faulty connected gas pipes lead to explosion hazard or even to mortal danger! Don‘t take a breath of the emissions! Emissions may contain hydrocarbons or other toxic components (e.g. carbon monoxide)! Carbon monoxide may cause headache, sickness, unconsciousness and death.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.3 Installation - Gas Preparation The analyzer should be mounted close to the sample gas source to minimize transport time. A sample gas pump can be used to reduce the reaction time; this requires that the analyzer be operated in bypass mode or fitted with a pressure control valve to protect against excessive gas flow and pressure (Fig. 4-3). Exhaust Analyzer Exhaust Filter Installation Pressure control valve Flow sensor Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.4 Installation - Electrical Connections 4.4 Electrical Connections ELECTRICAL SHOCK HAZARD Only qualified personnel, observing all applicable technical and legal requirements, may install these devices and connect power and signal cables. Failure to comply may render the guarantee void and cause exposure to risk of damage, injury or death. The devices may only be installed by personnel who are aware of the possible risks.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5 Analyzer Specific Instructions for Installation 4.5 Analyzer Specific Instructions for Installation Installation instructions: X-STREAM XEGC & XEGP page 4-8 X-STREAM XE field housings page 4-19 Notes for wiring signal inputs and outputs Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP 4.5.1 X-STREAM XEGC, X-STREAM XEGP be used instead, provided it conforms to the specifications on 2-9. X-STREAM XEGC / XEGP analyzers should be operated in a horizontal position. An optional kit with brackets and blindplate enables to install XEGC models into a rack; this is accomplished by means of four screws ( 2-7 ). The external PSU is optionally available for rack mounting, too.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Purge gas inlet (optional) Fig. 4-5: Fuses Power in Power switch Serial CAN interface Serial interface ( RS 232 / Modbus 485) 4 Installation Gas outlets Gas inlets Digital in- / outputs (Board 2) Digital in- / outputs (Board 1) USB connectors RJ45 ethernet connectors Analog outputs / relay outputs Service interface 4.5.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Purge gas inlet (optional) Fig. 4-6: Fuses Power in Power switch Strain reliefs and shielding contacts for signal cables Serial CAN interface Serial interface ( RS 232 / Modbus 485) 4 screw holes for rack mounting (7.5 x 10.5 mm / 0.29" x 0.41") Gas outlets Gas inlets Digital inputs / outputs USB connectors RJ45 ethernet connectors Analog outputs / NAMUR status relays Services interface 4.4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP Signal inputs and outputs The number of signal outputs actually available varies according to the unit’s configuration. Analog signals Relay outputs Analog signals and relay outputs are located on a shared 25-pin submin socket (X1; Fig. 4-7), 4-13). or on an optional terminals adaptor XSTA ( 4 (0) - 20 mA; burden: RB ≤ 500 Ω Dry relay change-over contacts can be used as NO or NC. max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP Serial interface For specifications and notes on control, see chapter 7 Pin 1 Pin Pin 1 6 2 7 3 8 4 9 5 Pin 8 Pin no. 1 2 3 4 5 6 7 8 9 Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP 1 2 Analog Outputs 0.14 ... 1.5 mm2 Recommended (AWG 26 ... AWG 16) wire gauge: end sleeves not required Skinning length: 5 mm (0.2") Thread: M2 Min. tightening 0.25 Nm torque: (2.21 in.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP Digital inputs & outputs 7 or 14 digital inputs (X-STREAM XEGC: max. 7 inputs) 9 or 18 additional relay outputs (X-STREAM XEGC: max. 9 add. outputs) max. 30 V, internally limited to 2.3 mA HIGH: min. 4 V; LOW: max. 3 V common GND Dry relay change-over contacts can be used as NO or NC electrical specification electrical specification max. load.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP 1 0.14 ... 1.5 mm2 Recommended (AWG 26 ... AWG 16) wire gauge: end sleeves not required Skinning length: 5 mm (0.2") Thread: M2 Min. tightening 0.25 Nm torque: (2.21 in.lb) Note! The configuration illustrated here is that of the first adapter (on socket X4.1). Inputs 8-14 and outputs 14-22 are on the second adapter (on socket X4.2), if installed. 2 1 Connector for socket X4.1 / X4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP Analog inputs Analog inputs are located on a 9-pole submin-D-connector (plug X5; Fig. 4-12) or on an optional terminals adaptor XSTI (Fig. 4-13).
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP A XSTI adapter can optionally be used to connect analog IN cables to screw-type terminals instead of a submin-D plugs and sockets. The adapter is plugged onto the unit´s submin-D connector X5. 0.14 ... 1.5 mm2 Recommended (AWG 26 ... AWG 16) wire gauge: end sleeves not required Skinning length: 5 mm (0.2") Thread: M2 Min. tightening 0.25 Nm torque: (2.21 in.lb) Signal P3.1 P3.2 P3.3 P3.4 P3.5 P3.6 P3.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.1 Installation - X-STREAM XEGC, X-STREAM XEGP Power supply X-STREAM XEGC 24 VDC is supplied to the unit by means of a three-pin XLR connector on the rear panel of the instrument. Depending on the order, the following is supplied as an accessory: either • an external power supply unit which can be connected directly to the analyzer using the supplied cable or • a connector which can be used with a cable and PSU as specified by the customer.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings 4.5.2 X-STREAM XEF, X-STREAM XDF Fitted with four eyebolts and featuring IP66/ Type 4X protection, the X-STREAM XE field housing can be mounted in the open air on a wall or frame with no extra work.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings Note! Take care to reserve space at the right side of the instrument for laying the cables! Gas fittings Connector for potential equalization Cable glands All dimensions in mm [inches in brackets] Fig. 4-16: X-STREAM XDF - Dimensions for Installation 4-20 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings 8 1 2 3 4 Note on XDF! In case of the dual compartment version XDF, the electrical connections are established in the upper compartment, and the gas connections to fittings at the lower compartment. Besides this, the design and layout of terminals and fittings are the same as with the single compartment version XEF.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings Gas connections Gas connections max number max for purging (incl. / separate) material sizes 8 2 incl. PVDF; stainless steel (opt.) 6/4 mm; 1⁄4" Signal in- and outputs Preparation of signal cables All signal cables are connected to screw-type terminals located inside the housing.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings ELECTRICAL SHOCK HAZARD Verify the power supply at installation site meets the specification given on the analyzer´s nameplate label, before installing the instrument! Verify power cables are disconnected and/or instrument is de-energized prior to working at the terminals! A maximum of three modules with 36 terminals each can be fitted.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings Analog signals Relay outputs 1-4 Terminals for analog signals and relay outputs 1 - 4 are located on the outer left module (terminal block X1; Fig. 4-18). Specification of analog signal outputs: 4 (0) - 20 mA; burden: RB ≤ 500 Ω Dry relay change-over contacts can be used as NO or NC. max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings Serial interface The 9 terminals on the left (28 - 36) of the right most strip carry the serial interface signals.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings RJ45 Ethernet connection The RJ45 connectors are located on an electronics board in the card cage section of the unit (Fig. 4-17, pg. 4-21). For connecting the analyzer into a network for analysis purposes, use the leftmost Ethernet connector. To install this connection, a cable must be fed through the cable entry without a connector.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings Digital inputs and relay outputs (option) Terminals for these signals are located on the terminals board XSTD (terminal block X4; Fig. 4-21). electrical specification Notes! Depending on configuration, an analyzer can be fitted with up to two of these terminal blocks (the unit will then feature 14 digital inputs and 18 digital outputs). To aid identification, the sockets are labelled X4.1 and X4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings Analog inputs (option) Terminals for analog input signals are located on the terminals board XSTI (terminal block X5; fig. 4-22).
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings Connecting the power cord The power cord is connected to screw-type terminals located inside the housing. Skinning length: Thread: Min.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.5.2 Installation - X-STREAM XE Field Housings ELECTRICAL SHOCK HAZARD BY MISSING EARTHING CONDUCTOR Before completing the electrical connection of the instrument, verify cables are inserted and connected in correct manner! Ensure the earthing conductor (protective earth; PE) is connected! When all connections are correctly made and checked, • 4-30 close the front panel and secure with the two fasteners. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.6 Installation - Notes on Wiring 4.6 Notes On Wiring Signal Inputs and Outputs Emerson Process Managament has made every effort during the development process, to ensure that the X-STREAM analyzer series ensures electromagnetic compatibility (EMC) with respect to emission and interference resistance, as confirmed by EMC measurements. Electrical Shielding of Cables Installation 4.6.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.6 Installation - Notes on Wiring On-site conditions often differ from test environments and may require special precautions. Such a case arises when strong electromagnetic fields which could induce an interference current in the shielding. This type of current creates a potential difference between the connected housings. Two possible methods of eliminating this are described here.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.6 Installation - Notes on Wiring Installation Fig. 4-27: Shield Connector Terminal With Cable 4 X-STREAM XEGP/XEGC with screw-type terminal adapters In order to avoid measured values being influenced by external interference signals when terminal adapters are in use, the signal cable shieldings must be connected to the analyzer housing by means of shield connector terminals: • Strip the signal cable to a length of 20 cm (8").
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.6 Installation - Notes on Wiring 4.6.2 Wiring Inductive Loads Switching inductive loads creates electromagnetic interference: When an inductive load (e.g. relay, valve) is switched off, the magnetic field resists the change in current; this induces a high voltage across the coil contacts (several hundred volts). This impulse propogates through the connected cables and can influence any electrical devices nearby or destroy signal inputs and outputs.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 4.6 Installation - Notes on Wiring 4.6.4 Driving Multiple Loads Frequently, several loads in one system are controlled by several analyzer outputs, whereby the power for the loads derives from a common source. Special care is needed when wiring the loads to minimize interference from switching these loads: avoid connecting the loads to their power supply by a common line: L Installation • 4 Fig.
X-STREAM XE 4-36 Instruction Manual HASXEE-IM-HS 04/2010 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Chapter 5 Startup 5.1 Introduction Once the unit has been unpacked and installed, it is recommended to first check the settings, and if necessary adjust them to the user’s needs. e.g: • What hardware is installed? • Is the unit configured to your needs (alarms, inputs, outputs, etc.) The following pages describe how to navigate through the menus and what is to be observed when configuring the unit.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.2 Symbols Used 5.2 Symbols and Typographical Conventions In the following sections, the symbols and typographical conventions explained below are used to describe the software menus and navigation. Symbol Description Convention Within Process Descriptions Setup.. Menu title Setup.. Analog outputs.. Parent (SETUP) and current menu (ANALOG OUTPUTS) Analog outputs.. Output1...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.3 Front Panel Elements 5.3 Front Panel Elements All X-STREAM XE gas analyzers have an graphic display to show measuring and status information, and the easy-to-use menu-based user interface for entering parameters. For ease of understanding, the user can at any time select one of three languages stored in Ch1 Ch2 Ch3 Ch4 Ch1 Ch2 Ch3 Ch4 5.3.2 X-STREAM XE Front Panel Startup 5.3.1 0.000 ppm CO 0.000 ppm CO 0.000 ppm CO 0.000 ppm CO Display 10.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.3 Front Panel Elements 5.3.3 Keys Six keys enable the use of the menu system. Depending on the operational mode (measuring, browsing menus, editing) they have the following functions: enter key: Mode Function Measuring Enter main menu Open submenu (..
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.3 Front Panel Elements left key: Function Enter main menu or Measuring open 2nd measurement display page (if configured) Go up 1 level or page in menu Browsing system Move cursor 1 char to the left Leave channel selection Cancel editing of given paraEditing meter Go to previous menu page, if shows in first menu line key: Mode Function Measuring Enter main menu or open 2nd measurement display page (if configured) Browsing Open submenu (..
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.4 Software 5.4 Software The analyzer software displays measurement results and status messages, allows parameters to be set and edited, and maintenance functions (e.g. calibration) to be carried out. The software is organised hierarchically: The topmost level is called MEASUREMENT DISPLAY, followed by a MAIN MENU; all other menus and submenus are arranged below ( fig. 6-1, page 6-2).
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.4 Software Editing The editing mode enables changing parameters. It is initiated by pressing enter. If the selection is a parameter list, the current entry is highlighted and may be changed by up and down. If the selection is a value, the cursor is placed over the last character. Use up and down to change it. Use left and right to select another character.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.4 Software Access Levels & Codes Level four has the highest priority and is used for factory settings — only qualified EMERSON service personnel have access to this level. Level three gives access to system admin parameters, e.g. for configuring data acquisition systems communication. Level two covers the expert settings, e.g. basic settings for calibrations and measurements.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.4 Software 5.4.2 Special Messages 5.5 Depending on the last action performed by the user, one of the following messages may be displayed to assist or inform the user. Information on incorrect entry: The value entered by the user is outside valid limits. The display indicates what limits apply. Press to return to the previous screen to enter an acceptable value. Wrong input Range 0.1 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.5 Powering Up The setup menu enables several additional configurations, e.g. • 2 pages measurement display • different font sizes 6-29. Note! There are also functions, that do activate a relay, but are not shown on the display (e.g. concentration alarms). In such cases, check the status menu for more information. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.6 Selecting the Language 5.6 Selecting the Language If the analyzer is operational and it becomes clear that the incorrect language has been set, which is unintelligible to the operator, the following sequence of keypresses (starting at the measurement display) can be used to set the language. If the system has been set up accordingly, the code for access level 1 must be entered at this point to enable access to the following menu.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings 5.7 Checking the Settings ) Setup Display.. Alarm.. Measurements.. In-/Outputs.. Communication.. Alarms.. Installed Options.. Save-Load.. Display Contrast.. Language: English Phrase Version EN1.04B Measurements.. Measurement Display.. Menu Access.. 10 Min Auto Home: 10 Min Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings 5.7.1 Installed Options All X-STREAM gas analyzers can be fitted with a variety of optional components: follow these steps to see which options are installed on your analyzer. Press left to return to SETUP, highlight "Installed options" and press enter. Installed Options 1of2 Licenses.. Valves: Pumps: DIO#1 Installed: DIO#2 Installed: Anal.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings 5.7.2 Configuring the Display Press left to return to SETUP. Line 1: Line 2: Line 3: Line 4: Line 5: Comp1 Comp2 Comp3 Comp4 Blank Labels.. Note! X-STREAM currently supports one pressure sensor only. Values Press1 to Press5 thus refer to the same sensor. Display 1 labels Line 1: Line 2: Line 3: Line 4: Line 5: When entering LABELS..
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings 5.7.3 Calibration Setup Once the display settings have been checked, press left to return to SETUP and open CALIBRATION to check the calibration settings. Setup.. Calibration.. Multi-channel unit: Select the component to be set in the gas component selection menu. Note! For more detailed information on calibration 7-5. procedures, Ch1 Calibration Gases.. Tolerances.. Procedure.. Valve Assignment.. Program Sequence..
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings Press left several times to return to SETUP and open MEASUREMENT. Measurement 1of2 Ranges.. Damping.. Linearization.. Cross Interference.. Average.. Delay.. Ch1 Measurement 2of2 Cut-off: None Pressure Compensation .. Ch1 Pressure compensation Manual Pressure: 1013 hPa Pressure Pressure Status 1013 hPa Good Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings 5.7.4 Setting the Analog Outputs Press left to return to SETUP, and then open IN/OUTPUTS, and from there ANALOG OUTPUTS. Setup.. In/Outputs.. Analog Outputs.. Analog Outputs Output1.. Output2.. Output3.. Output4.. Output5.. Signal: OutRange: Low Scale: Max Scale: AutoScale: FailMode: 0/4 mA: 20 mA: Hold: 5-18 Comp1 0-20 mA 0.00 100.00 Yes Live 0.00 100.00 No Select the analog output you like to check.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings Description The analog output is deactivated A 0 or 4 mA signal is generated, e.g. to check the signal processing in an external system. Whether a 0 or 4 mA signal is generated, is set by the "Out 0 mA range” line ( next page). A 20 mA signal is generated, e.g. to check the signal processing in an exter20 mA nal system. Comp1 ... 5 Gas concentration Temp1 ... 5 Temperature Press1 ... 5 Pressure Flow1 ... 5 Flow Calc A ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings Signal: OutRange: Low Scale: Max Scale: AutoScale: FailMode: 0/4 mA: 20 mA: Hold: Comp1 0-20 mA 0.00 100.00 Yes Live 0.00 100.00 No Next, select the output range: • 0-20 mA (dead zero) generates a 20 mA signal, if a concentration is measured at the upper limit of the signal range. A 0 mA signal is generated if the sample gas concentration equals the value specified with "LowScale" .
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings Operation modes conforming to NAMUR 43 (NE 43) recommendations The operation modes described above do not generate a signal which enables detection of a failure in the measurement system. In such cases the behaviour of the output signal is undefined: either the last value is held, or a random NE 43 contains recommendations for setting analog outputs in order to avoid these situations.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings "0/4 mA" enables to finetune the analog output: Set "Signal" to 0 mA and, while measuring the output current, adjust it to the expected value. Accepted range: -10,000 ... +10,000 Signal: OutRange: Low Scale: Max Scale: AutoScale: FailMode: 0/4 mA: 20 mA: Hold: 5-22 Comp1 0-20 mA 0.00 100.00 Yes Live 0.00 100.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings 5.7.5 Setting Concentration Alarms Note! If concentration alarms are not being used, continue with 5-24. Press left until SETUP is displayed, then open ALARMS - CONCENTRATION Concentration Alarm Monitor: LoLo Level: Lo Level: Hi Level: HiHi Level: Hysteresis: States.. On 50.000 ppm 100.000 ppm 400.000 ppm 600.000 ppm 10.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings Ch1 Concentration Alarms LoLo Alarm: Lo Alarm Hi Alarm HiHi Alarm Off On Off Off Concentration 5.7.6 75.000 ppm You may turn the alarm function On or Off separately for each channel ("Alarm Monitor"). It´s also possible to use only some of the thresholds. In this case, set the not used to a level outside the range limits (for this, "Lo" and "LoLo" support entering negative values).
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.7 Checking the Settings Local Backup Save.. UsrBack Date Restore.. Undo Restore! Busy Progress (0..1000) 7/29/09 14:26 For now, make a local backup: Enter LOCAL BACKUP and then select "Save..". 0 0 Save new local backup and overwrite old one! Are you sure? Confirm the operation (select "Yes!"). 5 Startup No! Yes! Copying Data Busy Progress (0..1000) 0 1000 Wait until "Progress (0..
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 5.8 Perform a Calibration 5.8 Perform a Calibration We recommend to perform at least a zero calibration, after startup of the instrument, to ensure proper measuring results. Refer to chapter 7 for a ����������������� comprehensive description of calibration procedures. 5-26 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 Chapter 6 User Interface and Software Menus This chapter describes the structure and content of the X-STREAM gas analyzer software menus. While this chapter describes all software menus in hierarchical order, chapters 5 & 7 explain by practical examples, how to navigate through the menus to perform certain basic setup operations, or maintenance functions. 6.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2 Menu System 6.2 Menu System Note! This overview does only show menu branches up to the 3rd menu level, not functions nor parameter lines! E.g. the lines "Pump" or "Lock menus" of CONTROL are not shown. The analyzer´s menu system has a dynamic behavior in that it does not show entries not supported by the current analyzer configuration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2 Menu System 6.2.1 Switching On Input Ch1 Output Ch2 Flow 1 10.0000 10.0000 Ch1 Ch2 Ch3 Ch4 ppm CO ppm CO 0.00 l/min When the unit is powered up, a self-test (POST) is initiated, after which the unit shows the MEASUREMENT DISPLAY. Note! Two different measurement display layouts are available and user selectable. DISPLAY SETUP, 6-29. Depending on the setup, either a 2-channel layout or a 4-channel layout is used. 0.000 ppm CO 0.000 ppm CO 0.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2 Control Menu Control.. Perform zero calibrations 6-6 Perform span calibrations 6-7 Perform advanced calibrations Off Off 6-16 Press enter in this line to lock all menus, for which "Code" in MENU ACCESS has been set to On ( 6-32) Acknowledge messages 6-17 Available if the related internal pump is installed: Switch internal pump 1 / 2 On or Off.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.1 Zero Calibration Menu Control.. Zero Calibration.. If the system is setup accordingly, the access code for level 1 must be entered to gain access to this menu. Multi-channel unit: In SELECT COMPONENT, select the channel to be calibrated. Press enter in this line to cancel any ongoing calibration Press tion Zero Calibration Cancel! Start! Zero Gas 0.000 ppm Concentration 0.000 ppm Flow 0.00 l/min Status.. Results.. Restore..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.2 Span Calibration Menu Control.. Span Calibration.. If the system is setup accordingly, the access code for level 1 must be entered to gain access to this menu. Multi-channel unit: In SELECT COMPONENT, select the channel to be calibrated. Span Calibration Cancel! Start! Span Gas 0.000 ppm Concentration 0.000 ppm Flow 0.00 l/min Status.. Results.. Restore..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.3 Advanced Calibration Menu Control.. Advanced calibration.. Start and control valve supported calibration procedures Note! This menu is only available if “Valves” in INSTALLED OPTIONS is set to a value other than none. If the system is setup accordingly, the access code for level 1 must be entered to gain access to this menu.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu Advanced Calibration 2of2 Status.. Results.. Next Automatic Calibrations.. Open a submenu to see calibration status 6-11 summary for all channels Open a submenu to see calibration results 6-13 summary for all channels 6 Software Menus Open a submenu to view the scheduled dates for next automatically performed calibrations 6-118 . Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.3.1 Zero & Span Single Menu Control.. Advanced Calibration.. Zero&Span Single.. Start and control a valve supported calibration procedure for zero and span for a single channel only. Multi-channel unit: In SELECT COMPONENT, select the channel to be calibrated.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.3.2 Calibration Status Summary Control.. Advanced Calibration.. Status.. Press enter in this line to cancel any ongoing calibration Indicates the currently ongoing procedure (Purging, Zeroing, Spaning, Ready) Shows the current procedure, or Off Shows the remaining time for the current procedure Shows the time for the previous procedure Information about the step currently carried out Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.3.3 Calibration Status Single Control.. ... ... Calibration Status Single.. This menu gives a channel specific status. Note! This menu can be opened from several higher-level menus, dependent on which it now may be necessary to select a component: Multi-channel unit: In SELECT COMPONENT select the channel to be viewed.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.3.4 Calibration Results Control.. Advanced Calibration.. Results.. Result OK Any ZeroFail Any SpanFail No No Calibration Results Single.. Emerson Process Management GmbH & Co. OHG Open a submenu to see detailed, channel specific calibration results information 6-14 6-13 6 Calibration Results Summary Software Menus Open this menu, to see an overall calibration results summary.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.4 Calibration Results Single Control.. ... ... Calibration Results Single... This menu gives a channel specific summary of results. Note! This menu can be opened from several higher-level menus, dependent on which it now may be necessary to select a component: Multi-channel unit: In SELECT COMPONENT select the channel to be viewed. Calibration Results Single Zero Result Zero Date Span Result Span Date Calibr.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu Calibration Deviations Example: A measuring channel shows zero drift of 10 ppm per week. It is calibrated once a week. After the 3rd zero calibration, DEVIATIONS shows: ZeroDev: 10 ppm (=last calibr.) ZeroDev. total: 30 ppm (=summary of 3 calibrations carried out within 3 weeks) This menu gives a channel specific summary of deviations of calibration results.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.5 Apply Gas Menu Control.. Apply gas.. If the instrument is equipped with internal, or connected to external valves, this menu enables to apply a specific gas to the analyzer, e.g. for maintenance purposes. If the system is setup accordingly, the access code for level 1 must be entered to gain access to this menu. Multi-channel unit: In SELECT COMPONENT, select the channel for the gas to be applied.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.6 Acknowledgements Menu Control.. Acknowledgements.. Acknowledgements This submenu provides function lines to acknowledge status messages and alarms, separately (lines 1 to 5) or simultaneously (last line). To do so, highlight the relevant line and press enter.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.7 Ranges Menu Control.. Ranges.. If the system is setup accordingly, the access code for level 1 must be entered to gain access to this menu. Multi-channel unit: In SELECT COMPONENT, select the channel to be configured. Ranges Current Range Range Start Range End Concentration Range1 0.000 ppm 100000.000 ppm 25000.000 ppm Select the measuring range to be used.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.8 Data Logger Control.. Data logger.. Enter this menu to export data logger data to an external device for further processing. Data Logger Data Logger Data to USB! Cached Entries Total Entries 10 120 Number of entries currently in RAM, not yet saved to the internal data logger (total entries) file.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.2 Control Menu 6.2.2.9 Event/Calibration Logger Control.. Event/Calibration Logger.. Enter this menu to export event or calibration logger data to an external device for further processing. Note! This menu is available only, if an optional software features license code has been purchased and installed; 6-89 If the system is setup accordingly, the access code for level 1 must be entered to gain access to this menu.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3 Setup Menu Setup.. Enter this menu to change basic analyzer settings.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1 Display Setup Menu Setup.. Display.. From within this menu you can setup how measuring results show up, control menu access, and more. If the system is setup accordingly, the access code for level 1 must be entered to gain access to this menu. Adjust the display´s contrast 6-23 Selects the preferred language for the analyzer software. Available options may vary according to the software version.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.1 Display Contrast Setup Menu Setup.. Display.. Contrast Press (148). Contrast 148 Increase / decrease contrast by a value of 3, each time enter is pressed in either of these lines. Limits: 82 ... 208 Shows the currently used contrast value and is updated each time, enter is pressed in one of the above menu lines. Notes! If by mistake characters are not visible any longer, reset the contrast to the default value.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.2 Display Measurements Setup Menu Setup.. Display.. Measurements.. Enter this menu to setup measurement tags, units, precision and more for primary and secondary measurements. This specifies how the measured values are displayed. If the system is setup accordingly, the access code for level 2 must be entered to gain access to this menu. Configure the parameters for Measurements Component.. Temperature.. Pressure.. Flow..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.2.1 Display Component Setup Menu Setup.. Display.. Measurements.. Component.. Specify how the measured gas concentration values are displayed. Multi-channel unit: In SELECT COMPONENT, select the channel to be configured. Component Unit.. Precision: Gas Name: Range Tag Concentration d.ddd CO Ch1-R1 10.000 ppm Enter this line to specify the gas name for this component. This name will show up on the measurement display.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.2.1.1 Component Unit Setup Menu Setup.. Display.. Measurements.. Component.. Unit.. Within this menu, configure the component´s unit to be used for measurement results. Select the measuring unit for the component. Available options: ppm, ppb, Vol%, Custom Ch1 Component Unit Unit: ppm Unit Text: ppm Span Gas Unit: ppm Span Unit Text: ppm Custom Factor: 1.000000 Custom Offset: 0.000000 Span Gas 50000.000 ppm Concentration 0.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.2.2 Display Temperature Setup Menu Setup.. Display.. Measurements.. Temperature.. Select the temperature unit to be used for all measurements. Available options: °C, °F, K Temperature Unit Unit: Precision: °C 1 DSP-T1 30.0 °C Configure the precision of temperature displays Accepted values: 0 ... 2 Note! Conversion factors for the different units are pre-defined. 6.2.3.1.2.3 Display Pressure Setup Menu 6 Setup..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.2.4 Display Flow Setup Menu Setup.. Display.. Measurements.. Flow.. Note! This menu is available only if at least one flow sensor is installed ( 6-88 ). Flow Unit Unit: Precision: DSP-F1 l/min 2 1.00 l/min Select the flow unit to be used for all measurements. Available options: l/min, l/h, ml/min, gal/min Note! 1 gal = 1 US.liq.gal. = 3.7853 l Configure the precision of flow displays Accepted values: 0 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.3 Measurement Display Setup Menu Setup.. Display.. Measurement Display.. Enter this menu to configure the measurement display. Line 1: Line 2: Line 3: Line 4: Line 5: Comp1 Comp2 Display 2 Assignments Comp3 Comp4 Line 1: Comp1 Blank Line 2: Comp2 Line 3: Comp3 Line 4: Comp4 Line 5: Blank Labels.. Labels.. Note! Primary measurements are gas measurements.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.3.1 Setup Measurement Display Labels Setup.. Display.. Measurement Display.. Labels.. Display 1 Labels Line 1: Line 2: Line 3: Line 4: Line 5: Input Output Flow1 For each MEASUREMENT DISPLAY line you may enter an individual text, called "label". Specifications: • free alphanumeric text • maximum length: 8 characters.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Examples: MEASUREMENT DISPLAY, if besides gas names and units, only labels are setup: 0.000 ppm CO 0.000 ppm CO 0.000 ppm CO 0.000 ppm CO Input Labels Output Flow 1 MEASUREMENT DISPLAY, if also tags are setup: Gas name1) Tags Unit text1) Input Ch1 Output Ch2 Ch3 Ch4 Flow 1 0.00 l/min 0.000 ppm CO 0.000 ppm CO 0.000 ppm CO 0.000 ppm CO 0.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.1.4 Menu Access Setup Setup.. Display.. Menu access.. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Menu access Restrict Level1: Off Code Level1: 00000001 Restrict Level2: Off Code Level2: 00000002 Restrict Level3: Off Code Level3: 00000003 Activate: On Home When using access codes, we recommend NOT using the factory-set codes.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2 Calibration Setup Menu Setup.. Calibration.. Note! Multi-channel unit: Some of the following submenus require to select the component to be configured: In SELECT COMPONENT, select which channel´s calibration to be configured. Setup calibration tolerances; Ch1 6 Submenu to specify the calibration gases for the selected channel; 6-34. 6-35. Calibration Gases.. Tolerances.. Procedure.. Valve Assignment.. Program Sequence..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2.1 Setup Calibration Gases Setup.. Calibration.. Gases.. Enter this line to setup the zero gas concentration. Ch1 Gases Zero Gas: Span Gas: Range Gases.. 0.000 ppm 50000.000 ppm Current Range: Range 1 Enter this line to setup the span gas concentration. Submenu to specify different calibration gases for several ranges (if instrument is setup to below. use different ranges); Information about currently used range.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2.2 Setup Calibration Tolerances Setup.. Calibration.. Tolerances.. Multi-channel unit: In SELECT COMPONENT, select the channel to be configured. Ch2 Tolerances Off 20.0 % 20.0 % If tolerance check is enabled (On), setup the limits for zero and span gas individually. Accepted values: 0 ... 100 % (of the channel´s full range) 6 DeviatToler.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2.3 Setup Calibration Procedure Setup.. Calibration.. Procedure.. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Multi-channel unit: In SELECT COMPONENT, select the channel to be configured. The time required to completely fill the gas line with the related gas, when switching to zero or span gas 1). Adjust according your system. Accepted values: 0 ... [max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2.4 Setup Calibration Valves Note! This line is available only if “Valves” in INSTALLED OPTIONS is set to a value other 6-88 ). than none. ( Setup.. Calibration.. Valve Assignment.. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Ch1 Valve Assignment 2of3 Span1 Valve: V1 Purge Time: 1s Span2 Valve: None Purge Time: 0s Span3 Valve: None Purge Time: 0s Span4 Valve: None 0s Purge Time: On the 2nd menu page, assign the span valves, and specify their purge times for the different ranges of the selected channel.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2.5 Setup Calibration Program Sequence Setup.. Calibration.. Program Sequence.. Note! This line is available only if “Valves” in INSTALLED OPTIONS is set to a value other than none ( 6-88 ). This menu with 8 pages allows to setup a sequences of up to 30 actions (steps), to carry out individual calibration procedures.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2.6 Setup Calibration Interval Time Setup.. Calibration.. Interval Times.. Note! This line is available only if “Valves” in INSTALLED OPTIONS is set to a value other than none ( 6-88 ). Interval Times Select the procedure(s) you want to configure to be carried out on a regular (interval time) basis. Zero All.. Zero&Span All.. Programmed Sequence.. Blowback All.. 6.2.3.2.6.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu '...' in the title is replaced by 'Zero All', 'Zero&Span All', 'Programmed Sequence' or 'Blowback All', depending on the selected procedure. Month: Day: Hour: Minute: Set! Next 1 1 10 0 6/23/09 10:00 Set start date and time: The next calibration time is calculated, considering the entries in above lines and the interval time given on the previous page.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.2.7 Setup Calibration Deviations Setup.. Calibration.. Deviations.. Multi-channel unit: In SELECT COMPONENT select the channel to be setup. In the context of this menu, deviation means the value, the current zero or span calibration value is corrected by subsequent calibrations. Ch2 Deviations ZeroDev 5.000 ppm ZeroDev Total 7.000 ppm Zero Deviation Reset! SpanDev 100.000 ppm SpanDev Total 230.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3 Setup Measurement Setup.. Measurement.. If the system is setup accordingly, the access code for level 2 must be entered to gain access to this menu. Multi-channel unit: In SELECT COMPONENT, select the channel to be configured. Submenu to configure up to 4 ranges per channel; 6-44. Measurement 1of2 Ranges.. Damping.. Linearization.. Cross Interference.. Average.. Delay.. Setup t90 times; 6-47.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.1 Setup Measurement Ranges Setup.. Measurement.. Ranges.. Enter this menu to configure up to 4 ranges per channel. Select the range to be configured. Available options: Range1 ... Range4 For the current range, specify start and end concentration limits. Ch1 Ranges Current Range: Range Start: Range End: Starts & Ends.. Range Control: Auto Ranging.. Range1 0.000 ppm 50000.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.1.1 Measurement Ranges Starts & Ends Setup.. Measurement.. Ranges.. Starts & Ends.. Ch1 Measurement Autoranging 6 6.2.3.3.1.2 Select the range to be configured, and set start and end concentrations. Note! For automatic or remote range control, all ranges end values need to be different! Software Menus Starts & Ends Range1 Start: 0.000 ppm Range1 End: 500.000 ppm Range2 Start: 0.000 ppm Range2 End: 1000.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.1.2.1 Autoranging Switchover Levels Setup.. Measurement.. Ranges.. Autoranging.. Switchover Levels.. Ch1 Switchover Levels MinLevel1: 200.000 ppm MaxLevel1: 500.000 ppm MinLevel2: 400.000 ppm MaxLevel2: 750.000 ppm MinLevel3: 600.000 ppm MaxLevel3: 2000.000 ppm MinLevel4: 1800.000 ppm MaxLevel4: 5000.000 ppm Specify individual switchover levels, instead of using one single hysteresis value for all ranges. Max.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Setup Damping Setup.. Measurement.. Damping.. This software menu enables to setup an additional electronic damping (t90 time), that is added to the system damping. The reason to do so, is to e.g. have a smoother output signal. Ch1 Damping t90 Range1: t90 Range2: t90 Range3: t90 Range4: t90 Current Current Range t90 Time Max 2.0 s 2.0 s 2.0 s 2.0 s 2.0 s Ch1-R1 76.9 s Specify t90 times for each range of the selected channel.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.3 Setup Linearization Setup.. Measurement.. Linearization.. If the system is set up accordingly, the access code for level 3 must be entered to gain access to this menu. Switch On or Off the linearizer. Specify the linearization methode to be used. Available options: Splines, Polynom. Ch1 Linearization Linearizer: Off LinearMethod: Splines Calculate! Setpoints (X).. Actuals (Y).. Linearizer Min 20.000 ppm Linearizer Max 110.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Specify the linearization methode to be used. Available options: Splines, Polynom. Linearization LinearMethod: Splines LinearizerRange1: Off LinearizerRange2: Off LinearizerRange3: Off LinearizerRange4: Off Coefficients.. Current range: Range1 0.000 ppm Concentration Separately switch On or Off the Linearization for each range of the selected channel. Submenu to enter coefficients ( ). Select the currently to be used range.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.3.1 Setup Linearization Setpoints (X) Setup.. Measurement.. Linearization.. Setpoints (X).. Ch1 Setpoints (X) 1of4 Lin X1: 0.000 ppm Lin X2: 0.000 ppm Lin X3: 0.000 ppm Lin X4: 0.000 ppm Lin X5: 0.000 ppm Lin X6: 0.000 ppm Lin X7: 0.000 ppm 0.000 ppm Lin X8: To modify a linearization curve, within this menu enter up to 32 setpoint values (x values) for the new parameter sets.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.3.2 Setup Linearization Actuals (Y) Setup.. Measurement.. Linearization.. Actuals (Y).. Ch1 Actuals (Y) 1of4 0.000 ppm 0.000 ppm 0.000 ppm 0.000 ppm 0.000 ppm 0.000 ppm 0.000 ppm 0.000 ppm To modify a linearization curve, after having entered the setpoint values (x values) for the new parameter sets, within this menu enter the new actuals (y) values. 4 menu pages enable to enter up to 32 values.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.3.3 Polynomials Coefficients Setup.. Measurement.. Linearization.. Coefficients.. Assign polynomials to each range of the current channel. Multiple assignments are supported. Range1 Uses: Range2 Uses: Range3 Uses: Range4 Uses: Coeffs Poly1.. Coeffs Poly2.. Coeffs Poly3.. Coeffs Poly4.. Calculation.. 6-52 Poly1 Poly3 Poly3 Poly4 Submenus to enter polynomials coefficients; . Submenu to calculate the polynomials; .
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.3.3.1 Enter Polynomials Coefficients 10.0 % 5.0 % 100000 ppm No coeffs Enter the coefficients here for a 4th order polynomial: a4 x4 + a3 x3 + a2 x2 + a1 x + a0 6 Software Menus Overflow: Underflow: a0: a1: a2: a3: a4: RefValue State Poly1 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.3.3.2 Calculate Polynomials Ch1 LinearSetpoints.. CalcPolySet: Calculate! Cancel! State Poly1 State Poly2 State Poly3 State Poly4 Merit Submenu for linearization setpoints ( 6-55 ) All poly No coeffs No coeffs No coeffs No coeffs 0 Ch1 Approximation setup MaxItera: Cut-off: 6-54 100 0.001 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.3.3.2.1 Linearization Setpoints Ch1 Linearization setpoints 6-50 ). 6-51 ). 10000 ppm 10000 ppm 10000 ppm 10000 ppm 6 Software Menus Setpoints (X).. Actuals (Y).. RefPoly1: RefPoly2: RefPoly3: RefPoly4: Clear all points! Submenu to enter setpoints ( Submenu to enter actuals ( Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.4 Setup Cross Interference Setup.. Measurement.. Cross Interference.. This menu enables to configure up to 4 sources (internal or external) for cross interference compensation. Ch1 Cross Interference Activated: IF Source1.. IF Source2.. IF Source3.. IF Source4.. 6-56 Off Enable (On) or disable (Off) cross interference compensation for the selected channel.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Setup Cross Interference Sourcen Setup.. Measurement.. Cross Interference.. IF Source1 ... 4 Select the source of measuring values to be used for cross compensating the selected channel. Available options: None: source is disabled for cross compensation Conc1 ... Conc5: Measurement values of internal channels 1 ... 5 AIN1, AIN2: Analog input 1 or 2 Calc A ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.4.1.1 Setup Cross Interference Linearization Curve Setup.. Measurement.. Cross Interference.. IF Source1 ... 4 Linearization Curve.. Note! The menu figure to the left shows the default setup for the polynomials, which relates to a straight line. So, if your IF source signal is linear, no further actions or changes in this menu are required. Reference value to normalize the linearization curve. Accepted range: -1E+9 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.5 Setup Measurement Average Setup.. Measurement.. Average.. Some applications, like e.g. CEMS, require to calculate and monitor concentration averages. Enter this menu to setup averaging. Average 2 min 32 s 0.000 ppm Elapsed time for the next average. Last average result. 6 Averaging Time: Next Average in Last Average Software Menus In this line specify the averaging time. Accepted values: 0 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.6 Setup Measurement Delay Setup.. Measurement.. Delay.. This menu option enables to delay a measurement output (on all display, analog outputs, network, etc.). Use this option to compensate signal delays within multichannel instruments, if you need very synchronous results. Reasons for unsynchronous behaviour may be e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.3.7 Setup Pressure Compensation Setup.. Measurement.. Pressure Compensation.. Manually enter the current ambient pressure for pressure compensation. Note! If a pressure sensor is installed, this line is hidden! Ch1 Manual Pressure: 1013 hPa Pressure Pressure Status 1013 hPa Good 6 These lines show the pressure, currently used for pressure compensation and the status.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4 Setup In-/Outputs Setup.. In-/Outputs.. If the system is setup accordingly, the access code for level 2 must be entered to gain access to this menu. Enter the submenu for the in- or outputs you want to configure: In-/Outputs Analog Outputs.. Digital Outputs.. Digital Inputs.. Internal SHS.. Analog Inputs..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.1 Setup Analog Outputs Setup.. In-/Outputs.. Analog Outputs.. Configure your analyzer´s analog outputs. Analog Outputs Output1.. Output2.. Output3.. Output4.. Output5.. Emerson Process Management GmbH & Co. OHG Note! All submenus for the analog outputs settings are of an identical design. 6-63 6 Enter the submenu for the output you want to configure.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.1.1 Setup Analog Outputn Setup.. In-/Outputs.. Analog outputs.. Output1...5.. Signal: OutRange: Low Scale: Max Scale: AutoScale: FailMode: 0/4 mA: 20 mA: Hold: Comp1 0-20 mA 0.00 100.00 Yes Live 0.00 100.00 No "Signal” specifies the value associated with the selected output.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Enter a concentration, to equal the low output limit (0 or 4 mA) Enter a concentration, to equal the high output limit (20 mA) Enable (Yes) or disable (No) output autoscaling. Note! Factory settings are OutRange: 4-20 mA and FailMode: LOW - 10%, if not changed at time of order. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.1.1.1 Operation Modes Acc. NAMUR NE 43 If "OutRange" is set to 0-20 mA, a 20 mA signal is generated, if the measured concentration is equal to "Max Scale". A 0 mA signal is generated if the sample gas concentration is 0 (dead zero). However, a severed cable also results in a signal value of 0. An external data acquisition system thus cannot detect such an failure and accepts a gas concentration of 0.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu "0/4 mA" enables to finetune the analog output: Set "Signal" to 0 mA and, while measuring the output current, in this line adjust it to the expected value. Accepted range: -10,000 ... +10,000 Emerson Process Management GmbH & Co. OHG "20 mA" enables to finetune the analog output: Set "Signal" to 20 mA and while measuring the output current, in this line adjust it to the expected value. Accepted range: -10,000 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.2 Setup Digital Outputs Setup.. In-/Outputs.. Digital Outputs.. This first menu page enables to configure digital outputs 1 ... 4, which are the basic outputs, available with every X-STREAM XE analyzer ('X1' in the menu title refers to the instrument´s I/O connector X1).
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Off On Heartbeat Any Failure Any Off-Spec Any MaintRequ Any FctCheck Any calibrating Any zeroing Any spanning Any zero failed Any span failed Any range low Any range high Any conc alm Any average alm Any temper alm Any press alm Any flow alm V1 ... V20 Pump1 ... 2 Ext Alarm1 ... 8 PLC result1 ... 10 CalcA rslt LoLo CalcA rslt Lo CalcA rslt Hi CalcA rslt HiHi CalcB rslt LoLo ....
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu If your instrument features optional digital I/O boards, similiar menu pages for the additional digital outputs are unlocked. The options for each output are as described before.: Digital Outputs (X4.1) 1of3 Output5 Node: System Output5 Signal: Off Output6 Node: Digital OutputsSystem (X4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.3 Setup Digital Inputs Setup.. In-/Outputs.. Digital Inputs.. If your instrument features optional digital I/O boards, this menu appears, enabling to configure the digital inputs.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.4 Setup Internal SHS Setup.. In-/Outputs.. Internal SHS.. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.5 Setup Analog Inputs Setup.. In-/Outputs.. Analog Inputs.. Enter this menu to configure the optional analog inputs. Note! If your instrument does not feature analog inputs, this menu is not available. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Analog Inputs Analog Input1.. Analog Input2.. 6-74 Select the analog input you want to configure 6-75).
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.5.1 Setup Analog Input n Setup.. In-/Outputs.. Analog Inputs.. Analog Input1 ... 2 Max. Good Value: Min. Good Value: 800 1100 Adjust analog-digital conversion 1013 Good Shows the calculated input value, based on the conversion coefficients a0..a2, as specified in the submenu. 6-76 Adjust Conversion.. Calc. Input Value Status Input value status. Possible entries: Absent, Failure, Good, Simulated (e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.4.5.1.1 Adjust Conversion Setup.. In-/Outputs.. Analog Inputs.. Analog Input1 ... 2 Adjust Conversion.. Select the type of signal, provided by the external source. Available options: 0..10V, 0..1V, 4..20mA: voltage or current signal is applied Off: Input is switched off Test: the instrument internally generates a step-like signal for testing purposes. Adjust Conversion Input1 Mode: 4..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.5 Setup Communication Setup.. Communication.. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Setup the interface parameters to meet the configuration of your host system. Configure the serial interface ( Serial.. Web Server: Ethernet1.. Ethernet2.. On 6-78) Specify, if (On) or not (No) a web server connection is used.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.5.1 Setup Communication Serial Setup.. Communication.. Serial.. Select the Modbus protocol. Available options: MODB-RTU, None Serial Protocol: 32Bit Mode: Mbus ID: Baud Rate: Parity: Modbus-RTU 16BitLow 1 19200 Even Modbus mode of operation. Available options: 32Bit (=Daniel mode), 16BitLow (=Modicon mode, LOW word first) 16BitHigh (=Modicon mode, HIGH word first) Enter instrument ID for network. Accepted values: 1 ..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Setup Communication Ethernetn Setup.. Communication.. Ethernet1...2 Ethernet1 MAC 123456789 IP: 123.456.78.9 Subnet: 255.255.255.1 Gateway: 123.456.78.0 Use DHCP: Yes IP Status Ready Apply Configuration! Within these menus, configure the Ethernet communication for connector 1 or 2. Most entries are standard, to be setup to meet your local network configuration. Enter the second menu page to configure the Modbus parameters.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.6 Setup Alarms This menu and its submenus enable to configure a couple of alarm conditions. In case an alarm goes off, a status is set and the related pictogram shows in the display. Also digital outputs may be configured to be activated ( 6-68 ). Setup.. Alarms.. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.6.1 Setup Alarms Concentrations Setup.. Alarms.. Concentration.. Note! To configure concentration alarm levels for averaging measurements, 6-83! Ch1 Concentration Alarm Monitor: LoLo Level: Lo Level: Hi Level: HiHi Level: Hysteresis: States.. On 50.000 ppm 100.000 ppm 400.000 ppm 600.000 ppm 10.000 ppm Specify up to 4 concentration (threshold) levels: "LoLo" and "Lo" levels are to be setup below the expected concentration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.6.1.1 View Concentration Alarms States Setup.. Alarms.. Concentration.. States.. Ch1 Concentration Alarms LoLo Alarm: Lo Alarm Hi Alarm HiHi Alarm Concentration 6-82 Off On Off Off This menu gives an overview of activated alarms, based on the currently measured "Concentration". 75.000 ppm Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.6.2 Setup Concentration Average Alarms Setup.. Alarms.. Conc Average.. Note! To configure concentration alarm levels for non-averaging measurements, 6-81! If concentration averaging ( 6-59) is active, specify concentration alarms within this menu. In this case, an alarm goes off only, if an average value exceeds one of the given levels.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.6.2.1 Setup Alarms Conc Average States Setup.. Alarms.. Average.. States.. ConcAverage Alarms LoLo Alarm Lo Alarm Hi Alarm HiHi Alarm Average 6-84 Off Off On Off This menu gives an overview of activated alarms, based on the currently measured "Average" of concentration. 468.000 ppm Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Setup Temperature Alarms Setup.. Alarms.. Temperature.. 6.2.3.6.4 Configuring temperature alarms is similiar to the procedure for concentration alarms; 6-81 . Setup Pressure Alarms Configuring pressure alarms is similiar to the 6-81 procedure for concentration alarms; . 6 Setup.. Alarms.. Pressure.. Software Menus 6.2.3.6.3 6.2.3.6.5 Setup Flow Alarms Setup.. Alarms.. Flow.. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.6.6 Setup Alarms Calculatorn Setup.. Alarms.. Calculator A..D Any X-STREAM XE analyzer can be upgraded with optional software packages to provide 4 calculators (A ... D). The results can be monitored to activate alarms. Note! This menu is available only, if a valid software upgrade code has been purchased and entered ( 6-89 ). For more information on calculators, see the associated separate software features manual.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.6.6.1 Setup Alarms Calculator n States Setup.. Alarms.. Calculator n.. States.. LoLo Alarm Lo Alarm Hi Alarm HiHi Alarm 468.000 Unit A Note! The calculator result unit is shown as configured in SETUP - CALCULATOR. 6 Result A Off Off Off Off This menu gives an overview of activated alarms, based on the currently calculated "Result n" (here of Calculator A).
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.7 Setup Installed Options Setup.. Installed Options.. Note! Don´t change settings in this menu without experienced knowledge! Wrong settings may result in a defective instrument. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Some software features are optional and can be unlocked with a license code. To do so, 6-89 Installed Options 1of2 Licenses..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Setup.. Installed Options.. Licenses.. Licenses Key 1: Key 2: Key 3: Package Trial Days 88888 88888 88888 Trial 21 This menu is used to unlock software features, to be purchased separately. By default, X-STREAM XE analyzers provide a web browser interface and a basic data logger. 3 optional software packages are available, to upgrade the software: Enhanced: add PLC and calculator.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.7.2 Setup Installed Options Flow Setup.. Installed Options.. Flow.. Select the flow measurement data source for the currently selected channel. Available options: XSP F1 ...XSP F4: internal sensors, connected to the board XSP AIN1, AIN2: analog inputs Ch1 Flow Flow Source: Flow SensorMin.: SensorMax.: Flow Status XSP F1 1.00 l/min 0.00 l/min 1.50 l/min Good Currently measured flow. The sensor´s minimum and maximum limits.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.7.3 Setup Installed Options Pressure Setup.. Installed Options.. Pressure.. Pressure Pressure Source: Pressure SensorMin: SensorMax Pressure Status Reference: Manual Pressure: Compensation: AIN1 1013 hPa 800 hPa 1100 hPa Good 1013 hPa 1014 hPa Off Currently measured pressure, or currently entered manual pressure value. The sensor´s minimum and maximum limits, or the limits for manual pressure entries. The sensor´s status.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.8 Setup Save-Load Setup.. Save-Load.. Enter this menu, if you want to save or restore configuration files from/to your analyzer. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Enables to save/restore analyzer configuration data to/from a special internal memory 6-94 area; Save-Load Local Backup.. Factory Defaults.. USB Backup.. USB Firmware Update..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu Notes on Save-Load Procedures Save new local backup and overwrite old one! Are you sure? After selecting any 'backup' procedure, a safety prompt appears: select "Yes!" to start the backup; "No!" cancels.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.8.2 Save-Load Local Backup Setup.. Save-Load.. Local Backup.. This menu enables to save or restore the current analyzer configuration to/from a special internal memory area. Consider the notes on 6-93 ! Save the configuration data Date of last backup Local Backup Save.. UsrBack Date Restore.. Undo Restore! Busy Progress (0..1000) 7/29/09 14:26 0 0 Restore saved configuration data Undo a restore Wait until "Progress (0..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.8.4 Save-Load USB Backup Setup.. Save-Load.. USB Backup.. This menu enables to save or restore the current analyzer configuration to/from an external USB device. Consider the notes on 6-93 ! Save the configuration data USB Backup Restore saved configuration data 0 1000 Undo a restore 6 Save.. Restore.. Undo Restore! Busy Progress (0..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.8.5 Save-Load USB Firmware Update Setup.. Save-Load.. USB Firmware Update.. Replace current firmware! This will restart device! Are you sure? No! Yes! 6-96 Firmware is the analyzer´s basic operation software. This menu enables to update your current analyzer firmware from an external USB device, e.g. to add new features, etc.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.9 Setup Operation Hours Meter Setup.. Operation Hours Meter.. If the system is set up accordingly, the access code for level 2 must be entered to gain access to this menu. Operation Hours Meter MaintRequInterval: Hrs of Operation Reset Hours! 26000 h 145 h Operating hours since last reset. Note! If this function is enabled, a maintenance request is triggered the moment, "Hrs of Operation" matches "MaintRequInterval".
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.10 Setup Identification Setup.. Identification.. Enter analyzer identification data here. If the system is set up accordingly, the access code for level 3 must be entered to gain access to this menu. Identification System Tag: Plant Name: Customer: Device tag Plant name Company name Enter a device tag here to identify the analyzer in e.g. a network Enter a plant and customer name here. Component Tags..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.10.1 Component Tag Setup Menu Setup.. Identification.. Component Tags.. Multi-channel unit: In SELECT COMPONENT select the channel to be setup. Tags Tag: Range1 Tag: Range2 Tag: Range3 Tag: Range4 Tag: Ch1 Ch1-R1 Ch1-R2 Ch1-R3 Ch1-R4 Note! If set, the "Tag" always appears in the very top menu line, if the current menu refers to a specific component.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.11 Setup Time Setup.. Time.. Within this menu configure time settings. Note! Correct time settings are important for e.g. time interval based calibrations and log files entries. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Select, how the analyzer time is set Available options: Manual, NTP Time Time Source: NTP Server.. Manual Time..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.11.1 Setup NTP Server To sync the time with a NTP server, • select NTP as "Time Source" in the previous menu • ensure the analyzer has internet access ( SETUP - COMMUNICATION - ETHERNET1 or ETHERNET2; page 6-79 ). Next enter • a valid NTP server IP address • the difference (hrs) of your local time to the standard NTP time (UTC) Accepted values: -12 ...
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.11.2 Setup Manual Time Setup.. Time.. Manual Time.. Manual time Year: Month: Day: Hour: Minutes: Set! Time 2009 7 7 10 29 7/7/2009 13:29 Use this menu page to manually set analyzer date and time. Note! Time format is 24h (13 = 1 pm) The last line shows the current analyzer time. 6-102 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.12 Setup USB Interface Setup.. USB Interface.. If the system is set up accordingly, the access code for level 3 must be entered to gain access to this menu. Interface: USB Operation USB Total USB Free Usage Warning: enabled Mass Memory 4033.6 MB 3985.7 MB 95 % Format USB Stick.. If "Interface" is set to Disabled, the line shows Disabled, too.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.12.1 Format USB Stick Setup.. USB interface.. Format USB stick... Format USB stick and erase all existing data! Are you sure? No! Yes! Select "Yes!" to start formatting. A progress screen appears. Formating has ended when "Busy" in this screen shows 0. Note! While formatting is in progress, "USB operation" in the previous menu shows formatting.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.13 Setup Data Logger Setup.. Data Logger.. If the system is setup accordingly, the access code for level 3 must be entered to gain access to this menu. Enable (On) or disable (Off) data logging. Data logger Logging: Sample Time: Data Logger Data Delete! Cached Entries Total Entries Off 1s 14 0 Export.. Export data to USB! Number of records currently in RAM, not yet saved to the internal data logger file.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.13.1 Setup Data Logger Export Setup.. Data Logger.. Export.. Export 6-106 Concentration: Temperature: Flow: Pressure: Yes No No Yes FieldSep: TAB Menu lines 1 ... 4: Specify which data to be exported. Available options: Yes, No Select the field separator, separating the different data columns in the exported text file. Available options: TAB, Semikol, Comma Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.3 Setup Menu 6.2.3.14 Setup Event Logger Setup.. Event Logger.. Note! This menu is available only, if a valid software upgrade code has been purchased 6-89 ). See the separate and entered ( software options manual for more information on this menu. Note! This menu is available only, if a valid software upgrade code has been purchased 6-89 ). See the separate and entered ( software options manual for more information on this menu. 6 Setup.. PLC..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4 Status Menu Status.. Within this menu, select any submenu to see detailed information about related status. Note! Lines 1 to 4 conform with NAMUR status. If no status of a specific type is active (no message to show), the related menu line is hidden. Failure status information ( 6-109 ) Off-specs status information ( Failures.. OffSpecs.. Maintenance Requests.. Function Checks.. Measurement.. Calibration.. Alarms..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.1 Status Menu Failures Failures Ch1 SENSOR Communication Ti.. Ch2 SENSOR Communication Ti.. Ch3 SENSOR Communication Ti.. Ch4 SENSOR Communication Failures 2 Ti.. Failures 3 6.2.4.2 Up to 3 menu pages are prepared to show status messages of type 'Failure'. Note! If no failure is active (no message to show), this menu is hidden.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.3 Status Menu Maintenance Requests Status.. Maintenance Requests.. Maintenance Requests Maintenance Requests 2 Maintenance Requests 3 6.2.4.4 Up to 3 menu pages are prepared to show status messages of type 'Maintenance requests'. Note! If no maintenance request is active (no message to show), this menu is hidden.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.5 Status Menu Measurement Status.. Measurement.. View status information on a specific component ( 6-112 ) Component.. Temperature Sensors.. Pressure Sensors.. Flow Sensors.. View status information on pressure sensors ( 6-115 ) 6 View status information on flow sensors ( 6-115 ) Software Menus Measurement View status information on temperature sensors ( 6-114 ) Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.5.1 Status Menu Component Status.. Measurement.. Component.. Multi-channel unit: In SELECT COMPONENT, select the channel to be viewed. Ch1 Component Raw Signal1 109518.000 Raw Signal2 54321.000 Raw Signal 2.016 Concentration 1,984 % Current Range Range 1 Statistics.. Secondary Variables..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu Status Menu Component Statistics Status.. Measurement.. Component.. Statistics.. Ch1 Statistics MinConc MinDate MaxConc MaxDate 0.000 ppm 7/7/2009 07:42:49 5000.000 ppm 8/7/2009 12:11:10 StdDev Start Date Reset Statistics.. 6.2.4.5.1.1.1 100 ppm 1.1.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.5.1.2 Status Menu Secondary Variables Status.. Measurement.. Component.. Secondary Variables.. Ch1 Temperature Temp Status TempZComp TempSComp Pressure Pressure Status Flow Flow Status Source Current 6.2.4.5.2 40.0 °C Good 0.0 °C 0.0 °C 1013 hPa Good 1.1 l/min Good If the analyzer features an IR measurement, this line shows the IR source current. Status Menu Temperature Sensors Status.. Measurement.. Temperature Sensors..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu Status Menu Pressure Sensors DSP-P1 Status DSP-P2 Status AIN1 Status AIN2 Status 6.2.4.5.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.6 Status Menu Calibration Status.. Calibration.. Calibration Component.. Next Automatic Calibrations.. 6.2.4.6.1 Enter this menu to view calibration status information. View calibration data for a specific component 6-116. Enter this menu to view scheduled dates for the next automatically performed calibrations; 6-118. Status Menu Single Calibration Status.. Calibration.. Component..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.6.1.1 Calibration Status Single Menu Status.. Calibration.. Single Calibration.. Status.. This page provides calibration status information for the last calibration for the selected component. Ch1 Calibration Status Single Calibration Results Single Menu Status.. Calibration.. Single Calibration.. Results... 6 6.2.4.6.1.2 Ready 0s 0.000 ppm 0.000 ppm 5000.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.6.1.2.1 Calibration Results Single Deviations Status.. Single Calibration.. Results.. Deviations.. This menu page provides calibration results for the selected channel. Ch1 Deviations ZeroDev ZeroDev Total 0.000 % 0.000 % SpanDev SpanDev Total 0.000 % 0.000 % 6.2.4.6.2 Next Automatic Calibrations Menu ... ... Next Automatic Calibrations..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu Status Menu Alarms Alarms Concentration.. Conc Average.. Temperature.. Pressure.. Flow.. Alarms 2 Calculator A.. Calculator B.. Calculator C.. Calculator D.. Emerson Process Management GmbH & Co. OHG This menu lets you select from several alarm functions, to view detailed status information.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.7.1 Alarms Status Details Status.. Alarms.. ... All alarm status menus, accessible from STATUS - ALARMS ( previous page), are designed in a similiar way to provide the information. For submenus of the first STATUS ALARMS menu page only: Multi-channel unit: In SELECT COMPONENT select the channel to be viewed. '...' in the title is replaced by the text of the submenu line, selected in the previous menu, e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.4 Status Menu 6.2.4.7.2 Operation Hours Status Status.. Operation Hours Meter.. Multi-channel unit: In SELECT COMPONENT select the channel to be viewed. Ch1 30000 h 145 h 6 MaintRequInterval Hours of Operation Software Menus Operation Hours Meter Multi-channel unit: Press left to enter SELECT COMPONENT, to view the status for another channel. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.5 Info Menu 6.2.5 Info Menu Info.. Open this menu to view the most important information about your instrument at a glance. Info Firmware 1.0 XSP Version 1.0 Serial no 123456789 Components.. Installed Options.. Ethernet1 IP 192.168.1.1 Ethernet2 IP 192.168.1.2 10/01/10 14:00 Time Data Logger Event Logger PLC Enabled Calculator Enabled Identification..
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.5 Info Menu 6.2.5.1 Info Menu Components Info.. Components.. Components Channel1 Channel2 Channel3 Channel4 Channel5 Measurement.. Enabled Enabled Enabled Disabled Disabled Only measuring channels indicating Enabled are installed in the current analyzer. See below for the measurement info menu Info.. Components.. Measurement.. Software Menus Info Menu Measurements This menu shows the full scale limits. 6 6.2.5.1.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.5 Info Menu 6.2.5.2 Info Menu Installed Options Info.. Installed Options.. Installed Options Package Valves Pump DIO#1 Installed DIO#2 Installed Anal. Outputs AIN Installed 6-124 None None None Yes No 4 No This menu indicates, if your analyzer features any of the listed options. ("Package" refers to the software upgrade options; 6-89). Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.5 Info Menu 6.2.5.3 Info Menu Identification Info.. Identification.. Identification System Tag Plant Name Customer My Tag My Plant Me See how the instrument is identified. Info Menu Component Tags Info.. Identification.. Component Tags.. View channel and range tags for a selected component. 6 6.2.5.3.1 Multi-channel unit: In SELECT COMPONENT select the channel to be viewed.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 6.2.6 Service Menu 6.2.6 Service Menu Service.. Service Europe.. North America.. Latin America.. Asia-Pacific.. Factory Setup.. This menu provides contact information about Emerson Process Management offices in several world regions (see below).
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 Chapter 7 Maintenance and Other Procedures 7.1 Introduction This chapter gives instructions not only for maintenance procedures, but also covers several procedures useful for properly operating the instruments. Maintenance carried out on a regular basis ensures long-term efficiency of your EMERSON Process Management gas analyzer! page 7-2 for general information about maintenance procedures and intervals.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.2 General Maintenance Information 7.2 General Maintenance Information Intervals given in the following tables are based on standard operating conditions (ambient temperatures +10 ... +40 °C / +50 ... +104 °F; temperature changes < 10 K /hr). Try cleaning contaminated components.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.3 Performing a Leak Test 7.3 Performing a Leak Test To achieve best and proper measuring results you must ensure the gas path system does not have leaks. The following procedure describes how to perform a leak test with focus on the instrument. The gas path system should be leak tested at least on a bimonthly basis and after maintenance, replacement or repair of gas path parts. Note! It is recommended to include external equipment (e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4 Calibration Procedures Note! To achieve best and proper measuring results, it is recommended to perform zero and span calibrations on a regular weekly basis. Also a zero calibration must always precede a span calibration! Zero calibration To perform a zero calibration supply either nitrogen (N2) or another suitable zero gas [conditioned ambient air or industrial air (NOT for oxygen measurement!)] to the gas path.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4 Calibration Procedures Before starting any calibration take care of the table below, and the following sections, describing general preparations for calibration procedures and how to perform such calibrations! Type of Procedure Manual calibration Menu Page (CONTROL - ) ZERO CALIBRATION... SPAN CALIBRATION... Furthermore you´ll find information about calibration gases setup: 7-7.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing Calibrations 7.4.1 Preparing Calibrations Before starting calibrations it is required to tell the instrument the calibration gas concentrations. Setup.. Calibration.. Gases.. Starting from the MEASUREMENT SCREEN press down to open the MAIN MENU, enter SETUP-CALIBRATION and directly enter GASES. Gases Zero gas: Span gas: Range gases.. 0.000 ppm 50000.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing Calibrations Next enter TOLERANCES: By default the option 'Deviat.toler.' (deviation tolerance check) is disabled (Off). Tolerances Deviat. toler.: Zero limit: Span limit: Off 20.0 % 20.0 % Example: Measuring range: 0 ... 50 % Zero gas: 0 % Span gas: 50 % Tolerance limits: 20.0 % (see figure above) Situation: Due to a fault zero gas is supplied to carry out a span calibration, instead of span gas. Deviat. toler.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing Calibrations When done, press left several times to return to SETUP: If you use analog output signals, you may want to check or setup, how the analog signals proceeds during calibrations. Signal: Out range: LowScale: MaxScale: Auto scale: Fail mode: 0/4 mA: 20 mA: Hold: 0 0-20 mA 0.00 100.00 Yes Live 0.00 100.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing Calibrations 7.4.1.1 Valve Assignment for Valve Supported Calibrations As described earlier, several calibration procedures require installed internal and/or external valves. In addition this requires all requested calibration gases to be connected to the valves and the valves to be software assigned to the gases.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing a Calibration Installed Options 1of2 Licenses.. None None No No 4 No 7 Valves: Pumps: XDIO1 Installed: XDIO2 Installed: Anal. Outputs: AIN Installed: Open SETUP - INSTALLED OPTIONS and check the "Valves:" line. Available options: None: Valves are not supported Internal: Open INTSHS ( 7-12) to assign internal valves. External: Open DIGITAL OUTPUTS ( 7-13) to assign external valves.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing a Calibration 7.4.1.1.1 Internal Valve Assignment If your analyzer provides internal valves, at first open SETUP - IN-/OUTPUTS - INTERNAL SHS to assign the valves to the gas inlets: Internal SHS (1of2) Gas1 Signal: Gas2 Signal: Gas3 Signal: Gas4 Signal: Gas5 Signal: Gas6 Signal: Gas7 Signal: Gas8 Signal: This menu allows to configure the optional internal valves for routing gas.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing a Calibration External Valve to Digital Output Assignment Setup.. In-/Outputs.. Digital Outputs.. Digital Outputs (X1) Output1 Node: System Output1 Signal: Off Output2 Node: System Output2 Signal: Off Output3 Node: System Output3 Signal: Off Output4 Node: System Output4 Signal: Off Further pages are indicated by a down arrow ( ), only when at least one extension card (outputs 5 - 13) is installed: Digital Outputs (X4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing a Calibration Verify which digital outputs are connected to control your external valves, and how the valves are labelled. Next enter the menu page, showing these outputs, and for each output select System in the line "Outputn Node" (where "n" is replaced by the output number). Finally for each output setup the valve's label. Digital Outputs (X4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing a Calibration 7.4.1.1.3 Calibration Valve Assignment For each channel a valve has to be assigned zero gas valve or span gas valve, whereat the valves can be freely assigned to any channel. This includes: • selecting the same combination for all channels • selecting combinations where one valve has the same function for several channels • selecting combinations where one valve has different functions for several channels, e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.1 Preparing a Calibration Ch1 Valve assignment 2of3 Span1 valve: V1 Purge time: 1s Span2 valve: None Purge time: 0s Span3 valve: None Purge time: 0s Span4 valve: None 0s Purge time: Now open the next menu to assign up to 4 span gas valves to the selected channel: one for each range. Note! Depending on the gas analyzer and SHS configuration, is may be possible to assign a specific valve to multiple ranges.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.2 Manual Calibration Manual Calibration Starting from the MEASUREMENT SCREEN press down to open the MAIN MENU and enter CONTROL. Control.. 7.4.2.1 To start a zero calibration select the first line: Off Off Manual Zero Calibration Ch1 Zero calibration Cancel! Start! Zero gas 0.000 ppm Concentration 0.000 ppm Flow 0.00 l/min Status.. Results.. Restore! Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.2 Manual Calibration The next lines show • the calibration gas setup (here: required zero gas concentration is 0.000 ppm), • the currently measured gas concentration • and the current gas flow. Ch1 Calibration status single Cancel! Calibr.status Ready Remaining time 0s Concentration 0.000 ppm Zero gas 0.000 ppm Span gas 5000.000 ppm Current range Range 1 Applied gas Sample gas "Status..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.2 Manual Calibration 7.4.2.2 Manual Span Calibration Zero calibration.. Span calibration.. Advanced calibration.. Apply gas.. Lock menus! Acknowledgements Pump 1: Pump 2: Ranges.. Select "Span calibration.." Off Off Span calibration offers the same options as Zero calibration, so for a detailled description 7-17. 7 Span calibration Cancel! Start! Span gas 0.000 ppm Concentration 0.000 ppm Flow 0.00 l/min Status.. Results..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration 7.4.3 Advanced Calibration Standard manual calibration procedures offer limited funtionality: To zero and span calibrate a multi channel instrument the operator has to manually start 2 procedures per channel in proper sequence. In addition he has to stay at the instrument to see when the one sequence has finished and to start the following.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration Zero All Calibration Before selecting any further line make sure the required calibration gas is applied! Supply all calibration gases with the same flow as the sample gas (recommeded approx. 1 l/min), pressureless and utilizing the correct gas fitting ( sect. 3.4).
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration The analyzer immediately begins zero calibration(s), showing the CALIBRATION STATUS SUMMARY screen. Calibration Status Summary Calibration Status Single.. Current Action Ready Action Detail Off Current Duration 0s Prev.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration start zero all procedure select channel 1 start to zero all channels with the current zero gas valve assigned, considering the individual purge times is current channel the last channel? Yes END No 7 wait until all selected channels are zeroed increase channel no. by 1 Yes Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration 7.4.3.2 Span All Calibrations Before selecting any further line make sure the required calibration gas is applied! Supply all calibration gases with the same flow as the sample gas (recommeded approx. 1 l/min), pressureless and utilizing the correct gas fitting ( sect. 3.4).
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration The analyzer immediately begins span calibration(s), showing the CALIBRATION STATUS SUMMARY screen. Multi-channel unit: Select the channel in SELECT COMPONENT. Ch1 Calibration status single Cancel! Calibr.status Ready Remaining time 0s Concentration 0.000 ppm Zero gas 0.000 ppm Span gas 5000.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration start span all procedure select channel 1 select all channels with the same span gas valve assigned start to span all channels with the current span gas valve assigned, considering the individual purge times is current channel the last channel? increase channel no. by 1 Yes 7-26 END No wait until all selected channels are spanned Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration Before selecting any further line make sure the required calibration gas is applied! Supply all calibration gases with the same flow as the sample gas (recommeded approx. 1 l/min), pressureless and utilizing the correct gas fitting ( sect. 3-4).
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration start zero all procedure select channel 1 select all channels with the same span gas valve assigned select all channels with the same zero gas valve assigned increase channel no.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration The analyzer immediately begins to calibrate, showing the CALIBRATION STATUS SUMMARY screen. Ready Ch2 0s 0s 0 To see a detailled calibration status for a single channel, enter CALIBRATION STATUS SINGLE, Multi-channel unit: Select the channel in SELECT COMPONENT. Ch1 Calibration status single Cancel! Calibr.status Ready Remaining time 0s Concentration 0.000 ppm Zero gas 0.000 ppm Span gas 5000.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration 7.4.3.4 Notes on Span Calibrating Channels With Multiple Ranges X-STREAM XE series gas analyzers support up to 4 ranges per measuring channel 6-44). ( For valve supported calibrations, each range can be assigned an individual span gas valve ( 6-37 and figure to the left).
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.3 Advanced Calibration 7.4.3.5 Blowback From ADVANCED CALIBRATION you may also start a blowback procedure: Press enter in this line to . While the procedure is active, a "Function executing" message appears. 7 Maintenance & Procedures Advanced calibration 1of2 Cancel! Zero&span single.. Zero all! Span all! Zero&span all! Programmed sequence! Blowback! Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.4 Remote Calibration 7.4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.4 Remote Calibration 7.4.4.1 Calibrations Initialized by Digital Inputs As already mentioned, the analyzer must either provide internal valves or external valves (connected to its digital outputs), to make use of this feature. time slot starts after the zero calibration has ended (see fig. 7-7, signal D), otherwise it is rejected (signal B).. Chapter 4 for information about electrical data and installation of digital inputs and outputs.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.4 Remote Calibration Examples The sequences shown in Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.4 Remote Calibration Modbus Activated Calibrations Without Valves 7.4.4.3 This means, the Modbus command immediately starts the calculation. The operator has to ensure in this moment, the proper gas is applied and the measuring system is filled with calibration gas. If applicable, he also has to take care to not activate a span calibration without a preceding zero calibration. For detailled instructions about manual calibration 7-17.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.5 Unattended Automatic Calibration 7.4.5 Unattended Automatic Calibration The unattended automatic calibration feature allows to program the analyzer to automatically perform valve supported calibration procedures without the need of digital inputs or Modbus interface connections.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.5 Unattended Automatic Calibration Setup.. Calibration.. Interval times.. Zero all.. Zero & span all.. Programmed sequence.. Blowback all.. Settings Diagram 7 Several time intervals may be specified: "Zero all..": This entry specifies intervals for zero calibrations only! If there is an entry for "Zero & span all..", too, the instrument will carry out additional zero calibrations based on the "Zero & span all.." interval. "Zero & span all..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.5 Unattended Automatic Calibration All submenus opened by above menu lines have the same content, exemplarily described below: Set to Enabled, to use interval times for the selected calibration (here: Zero all) Enter the interval time Accepted range: 1 ... 10,000 h Zero all Enable: Interval: Start time.. Enabled 15 h In this submenu enter the start time for the first calibration (s.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4.6 Verifying a Calibration 7.4.6 Verifying a Calibration For instruments without internal and/or external valves simply apply either span or zero calibration gas to the sample gas inlet. If the calibration still is proper, the reading on the MEASUREMENT SCREEN should show the related value.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.4 Calibration Procedures 7.4.8 Restoring a Calibration Ch1 Zero calibration Cancel! Start! Zero gas 0.000 ppm Concentration 0.000 ppm Flow 0.00 l/min Status.. Results.. Restore! 7.4.7 In case a wrong configuration was detected after a calibration was carried out (e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5 Replacing the Electrochemical Sensor designed life (% hours) O2 concentration (%) The sensor‘s designed life under constant conditions of 20 °C is approx. 900,000 hrs. The lifetime at 21 % oxygen is therefore calculated to approx. 42,857 hrs, corresponding to approx. 5 years. Irrespective of all calculations above: A sensor is worn-out when, connected to ambient air, the output voltage is less than 2.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor HAZARD FROM WEAK ACID AQUEOUS SOLUTION If the electrolyte leaks due to sensor damage, put the sensor in a plastic bag so that the solution will not be smeared on other places and return the sensor to Emerson Process Management or an industrial waste management contractor.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5.1 General Hints on Handling the Sensor GENERAL HINTS ON HANDLING THE SENSOR Do not expose the sensor to a temperature other than the temperature range of -20 to +60‘C (-4 to +140 F). Exposing to a temperature outside the temperature range may cause abnormal output or leak of the electrolyte due to parts degradation or damage. Make sure to prevent condensation of the oxygen concentration detecting part.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5.2 Opening X-STREAM Analyzers ELECTRICAL SHOCK HAZARD Live parts are accessible when working at open instruments! Take care to observe all applicable safety instructions! 7.5.2.1 How to Open X-STREAM XEGP Remove the top cover after loosening the 12 screws. If your instrument features an internal heated fig. 7-11 on next for information on box, how to open. 7.5.2.2 Maintenance & Procedures Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5.2.3 How to Open X-STREAM XEF / XDF Depending on the individual analyzer configuration, either open the upper or lower front door to the left, utilizing the two sash fasteners. sash fasteners 7.5.2.4 How to Open X-STREAM XEFD To open a X-STREAM XEFD loosen the 20 screws located at the instrument´s flange.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5.4 Disassembling the Sensor Unit The sensor unit consists of a holder, an electronics board and the sensor itself, all together installed on a base plate (Fig. 7-13). After loosening the nut (5), push the holder (3) with sensor (1) until the nut is above the hole (see details), then lift the holder from the base plate (4). The sensor is still fixed in the holder by means of a clip (8).
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5.5 Adjusting the Output Signal Having replaced the worn sensor, the board´s output signal requires some adjustment. ELECTRICAL SHOCK HAZARD Tp 1 Strip P3 for sensor connection Fig. 7-14: OXS Board, Top View Emerson Process Management GmbH & Co. OHG Tp 2 Procedure: • power on the open instrument. • Supply ambient air (approx.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.5 Replacing the Electrochemical Sensor 7.5.6 Finalizing the Sensor Replacement • • • Disconnect the analyzer from power Re-install the sensor unit into the analyzer Close the housing. Take care to use all screws, especially if the instrument is to be used in hazardous areas! In a next step for proper measuring results, perform a zero and a span calibration at least for the channel with the replaced sensor.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.6 Cleaning the Instrument´s Outside 7.6 Cleaning the Instrument´s Outside Use a liquid general purpose detergent and a lint-free cloth for cleaning the analyzer´s outside.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7.7 Save / Restore Configuration Data Sets After some time of operating the instrument, one can assume all the parameters (calibration gases setup, measuring ranges, inputs and outputs, etc) are setup to meet the application´s and operator´s needs. To save these settings for means of restoring them in case of failures, data loss or even overwriting, use the options of the SETUP - SAVE-LOAD menus.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7.7.1 Local Backup - Save Starting at the MEASUREMENT SCREEN press down to open the MAIN MENU, enter SETUP and next SAVE-LOAD.. . Setup.. Save-Load.. If system is setup accordingly, access level 3 code must be entered to gain access to this menu. Save-Load Local backup.. Factory defaults.. USB backup.. Highlight "Local backup.." and press enter. Local backup Save.. UsrBack date Restore..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7.7.2 Local Backup - Restore Starting at the MEASUREMENT SCREEN press down to open the MAIN MENU, enter SETUP and next SAVE-LOAD.. . Setup.. Save-Load.. If system is setup accordingly, access level 3 code must be entered to gain access to this menu. Save-Load Local backup.. Factory defaults.. USB backup.. Local backup Save.. UsrBack date Restore..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7.7.3 Factory Defaults - Restore Setup.. Save-Load.. Starting at the MEASUREMENT SCREEN press down to open the MAIN MENU, enter SETUP and next SAVE-LOAD.. . If system is setup accordingly, access level 3 code must be entered to gain access to this menu. Save-Load Local backup.. Factory defaults.. USB backup.. Highlight "Factory defaults.." and press enter. Factory defaults FacBack date Restore..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7.7.4 USB Backup - Save IMPORTANT INFORMATION! Read carefully before activating USB procedures! Installation Mass storage devices can be hot-plugged. After attaching a device, the analyzer will automatically recognize it, if the USB interface is enabled; 6-103. However, do not remove a memory device, while data transmission is ongoing, this can cause loss of data! Fig.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets Starting at the MEASUREMENT SCREEN press down to open the MAIN MENU, enter SETUP and next SAVE-LOAD. Setup.. Save-Load.. If system is setup accordingly, access level 3 code must be entered to gain access to this menu. Save-Load Local backup.. Factory defaults.. USB backup.. Highlight "USB backup.." and press enter. USB backup Save.. Restore.. Undo restore! Busy Progress Highlight "Save..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7 Maintenance & Procedures The backup files are stored within the USB device' file structure, in a subdirectory 'config'. For more information on the USB device file system structure, 6-104 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7.7.5 USB Backup - Restore Starting from the MEASUREMENT SCREEN, press down to open the MAIN MENU, enter SETUP and next SAVE-LOAD.. . Setup.. Save-Load.. If system is setup accordingly, access level 3 code must be entered to gain access to this menu. Save-Load Local backup.. Factory defaults.. USB backup.. In this line press enter. USB backup Save.. Restore..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.7 Save / Restore Configuration Data Sets 7.7.6 Undo Restore Local backup 3/16/10 14:26 0 0 7 Maintenance & Procedures Save.. UsrBack date Restore.. Undo restore! Busy Progress Each backup menu has a function line called "Undo restore!" to undo the last restore backup operation, as shown exemplarily shown by the figures to the left (local backup menu).
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.8 Logfiles 7.8 Handling Log Files Log files are created by the internal data logger, event logger and calibration logger, whereat the latter are part of optional software upgrade packages. 7.8.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.8 Logfiles Press left to return to the previous menu, and • enter a sample time to specify the time interval between entries • turn "Logging" On, to start logging. 14 0 Data Selection.. Export data to USB! 7.8.2 Exporting Log Files There are two options to export log files to an USB device: Data logger Logging: Sample time: Data logger data delete! Cached entries Total entries On 1s Data Selection..
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.8 Logfiles 2nd option: From within CONTROL - DATA LOGGER (this may require to enter the access code for level 1) Data logger Export data to USB! Cached entries Total entries 0 0 The line "Export data to USB!" enables to export the total entries to a connected USB device. Before starting to export, the analyzer automatically adds the cached entries to the total entries file, so all available data is exported to the log file.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.8 Logfiles Log Files Content Example Imported into a text file, for a 3 channel instrument the above settings would give the following log file layout: The exported log file does not only show the discussed entries data, but also separate lines with • the type of log file • the analyzer tag, if such has been setup ( 6-98 ) • the analyzer serial number • column headings for the entries fields For further processing, import that file e.g.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.9 Files on USB Stick 7.9 Files on USB Memory Device After connecting or formatting an USB device, or after a first log file export, a special file structure is present on the stick, figure below. 7.9.1 Furthermore, two files are created within this structure: • autorun.inf • xe_win_tools.zip autorun.inf 'Autorun.inf' can be used to automatically start actions, when the USB device, it is saved on, is connected to the analyzer.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.9 Files on USB Stick 7.9.2 xe_win_tools.zip This ZIP contains some files to be used with a Microsoft Windows based computer only. Chiptool for Ethernet connections, enables to remotely identify an analyzer by its IP address, without requiring front panel access. 7 Maintenance & Procedures USB driver for MS Windows PC Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.10 Web Browser 7.10 Web Browser Field housings Table top analyzers To gain access to the instrument´s web browser interface, first ensure the instrument is powered and connected to your network via Ethernet1 connector (see figure on the left) Ethernet1 connector Fig. 7-21: Ethernet Connectors Next enter INFO, to check if the instrument has been assigned a valid network IP address: Info Firmware 1.0 DSP version 1.0 Serial no 123456789 Components.
Instruction Manual X-STREAM XE HASXEE-IM-HS 10/2010 7.10 Web Browser Connect your computer to the network, open a web browser and enter the instrument´s network IP address, and if all is configured properly, a screen like the following appears: IP address in web browser´s address line Analyzer information IP address on logon screen MAC address of interface Analyzer serial number After logon, the measurements screen appears.
X-STREAM XE 7-70 Instruction Manual HASXEE-IM-HS 10/2010 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 Chapter 8 Troubleshooting 8.1 Abstract 8.2 Problems indicated by status messages Analyzer related messages Channel related messages As the analyzer software is not capable to detect all problems and faults, section 8.3 describes such faults, their consequences, gives hints on potential causes and on how to solve the problem(s). Section 8.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2 Solving Problems Indicated by Status Messages 8.2 Solving Problems Indicated by NAMUR Status Messages As mentioned before, status messages show up in the measuring display´s last line. Multiple status messages, active at a time, show up sequentially in this status line. To see all status messages at a glance, enter STATUS: If any status is set, the corresponding menu line appears, whereat only the first 4 lines are of interest here (NAMUR status).
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.1 Analyzer Related Messages Message Status level Factory file read error Maintenance Factory file write error Maintenance FATAL!! Configuration data destroyed Failure FATAL: FRAM read/write error Maintenance Flash write count over limit Maintenance Explanation Recommended Actions Reading the factory configuration file is not possible Writing the factory configuration file is not possible.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.1 Analyzer Related Messages Message Status level USB free space warning Maintenance USB stick full The attached USB storage devices´s free memory exceeded the setup limit ( 6-103 ) The attached USB storage device has not sufficient free memory to store data Recommended Actions • Replace USB device by another one with sufficient free memory • On the USB device, delete files not required. To do so, attach it to a computer.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.2 Channel Related Messages 8.2.2 Channel Related Messages (preceded by Channel Tag, e.g. CO2.1) Message Status level Concentration Is Higher Than Limit Off-spec Concentration Is Lower Than Limit Off-spec Device Not in Service Function check External Failure Description Recommended Actions Currently the actual concentration is outside the analyzer's range limits. The shown measuring value does not comply to the actual concenctration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.2 Channel Related Messages Message Description Status level Invalid Interference Value Off-spec Linearizer Overflow Off-spec Linearizer Underflow Off-spec Operation Hours Exceeded Maintenance No Sample Gas Function check Range Overflow Off-spec A measuring value used for Check status of interfering cross interference compensacomponents tion is found to be erroneous.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.2.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.3 Solving Problems Not Indicated by Status Messages 8.3 Solving Problems Not Indicated by Status Messages The following table lists possible faults not detectable by the instrument‘s software, gives hints on the potential causes, and tips on how to solve the problems.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.3 Solving Problems Not Indicated by Status Messages Situation Description Recommended Actions • XDIO: If LED "TIMEOUT" glows - check jumpers on XDIO. XDIO #1: jumper on ADR2 XDIO #2: jumpers on ADR2 Digital Outputs Not Working Outputs on extension board(s) & ADR0 Properly (cont.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components 8.4 Troubleshooting on Components This section gives information on how to check and replace internal components.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components 8.4.1 Opening X-STREAM Analyzers ELECTRICAL SHOCK HAZARD Live parts are accessible when working at open instruments! Take care to observe all applicable safety instructions! 8.4.1.1 How to Open X-STREAM XEGP Remove the top cover after loosening the 12 screws. If your instrument features an internal heated fig. 8-4 on next page for information box, on how to open. Fig. 8-2: X-STREAM XEGP 8.4.1.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components 8.4.1.3 How to Open X-STREAM XEF / XDF Depending on the individual analyzer configuration, either open the upper or lower front door to the left, utilizing the two sash fasteners. sash fasteners 8.4.1.4 How to Open X-STREAM XEFD To open a X-STREAM XEFD loosen the 20 screws located at the instrument´s flange.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components 8.4.2 Fig. 8-5: 8-22 Signal Connectors on XSP Board XSP - Allocation of Signal Connectors Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components 8.4.3 Sample Pump: Replacement of Diaphragm 8 To do so you need to dismantle the pump from your analyzer. Troubleshooting This instruction explains the procedure to replace the diaphragms of sample gas pumps (PN 42716569) used in the X-STREAM series gas analyzers. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Required parts for the spare parts kit for the pump (PN 0375946). Step 1: S1 8-24 If applicable: Remove the screws S1 on both sides of the pump. Take off the cover. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components S3 S2 Step 2: 8 Troubleshooting Remove the screws S2 and screw S3. Step 3: Take out the pump assembly. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 4: Mark the pump assy. before disassembly. Step 5: Remove the white block. 8-26 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 6: Troubleshooting Remove the teflon gasket. 8 Step 7: Remove the remaining two pump parts. Clean the white plate for the gas inand outlet. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Clamp S4 Step 8: Disassemble the lower block and the clamp. Loosen the screw S4 and the nut N1. N1 8-28 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 9: 8 Troubleshooting Remove the two washers on the diaphragm. Step 10: Replace the old with the new diaphragm and assemble the washers and the clamp in reverse order (step 9 and 8). Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 11: Remove the locking springs on both sides of the white block and take out the old diaphragms on both sides. 8-30 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 12: Clean the white block. Afterwards put in the new dia-phragms and fix them with the new locking springs. 1 Assemble the pump assy. Take care of your marker ( step 4) 8 1. Put the two upper plates under the clamp ( steps 6 & 7 for reference). Troubleshooting Step 13: 2. Put the white block and the new teflon gasket between the lower block and the in-outlet plate. 2 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 14: Assemble the pump assy in reverse order. S3 Put it in the pump housing and fix it with the screws S2. Fix the clamp with screw S3 and the black buffer. S2 8-32 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 15: 8 Finally re-install the pump into your analyzer, to complete the replacement of pump diaphragm. Troubleshooting S1 If applicable: Install the cover and fix it with screws S1 at both sides. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components 8.4.4 Paramagnetic Oxygen Cell for Standard Applications: Adjustment of Physical Zero GND +6 V -6 V Signal connectors Signal pins To adjust the physical zero you need to measure some voltages on the XSP board: Depending on which channel the cell is assigned to, the measuring signal (+) can be measured at pin 3 of the related connector. GND (-) is available at a separate pin (see figure).
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 3: Apply N2 to the analyzer. Step 4: A Step 5: Tighten the screw S2 with care, close the cover and check the zero point again. Note! If the cell itself does not provide a cover, close the instrument while checking the cell! You might have to re-adjust the zero point several times until it remains at the expected value. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 6: Fix the closed cell´s cover with screw S1. S1 8-36 This completes the zero point adjustment procedure. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components 8.4.5 Thermal Conductivity Cell: Adjustment of Output Signal A digital voltmeter (DVM) is required to measure and adjust several voltages! 8 Troubleshooting To adjust the zero signal of this measuring cell you need to have access to both sides of the related electronics board WAP 100. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 1: Check the solder bridges, located at the solder side of the board, for proper configuration: LB20 LB10 LB10 LB4 2-5 LB21 1-4 LB20 LB21 LB4 open closed closed open Step 2: Switch on the analyzer. The onboard LED will light up red and green. LED LED When the warmup time has elapsed, the LED flashes green. LED 8-38 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 3: Locate test connector P4 to measure the bridge voltage: P4.16 P4.15 R60 Bridge voltage (+) Bridge voltage (-); GND CAUTION! Do not short-circuit pins! P4.16 + Test connector P4 _ 8 Troubleshooting P4.15 Alternatively the GND signal (-) is accessible on the main board BKS, too: fig. 8-3, page 8-16) . Locate X11 ( The bridge voltage depends on range and sample gas and should be between 3V and 5V.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 4: To adjust the physical zero point: Apply zero gas to the analyzer. P11/ P17 Connect the DVM to the following pins: P4.5 Raw signal (+) P4.15 Bridge voltage (-); GND CAUTION! Do not short-circuit pins! P4 P4.15 P4.5 Raw To adjust the physical zero point, it is necessary to install a resistor between P11/ P17 at position 1, 2, 3 or 4 (the following figure shows it at position 4).
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 5: To adjust the physical span: LB3 R119 P4 Apply span gas to the analyzer. Do not disconnect the DVM: P4.5 Raw signal (+) P4.15 Bridge voltage (-); GND CAUTION! Do not short-circuit pins! P4.15 P4.
Instruction Manual X-STREAM XE HASXEE-IM-HS 06/2010 8.4 Troubleshooting on Components Step 7: LB10 To finetune the physical zero point: Close solder bridge LB10. Apply zero gas to the analyzer. Do not disconnect the DVM: P4.5 Raw signal (+) P4.15 Bridge voltage (-); GND CAUTION! Do not short-circuit pins! Now you can finetune the zero point to a minimum value, using R103. R103 P4.15 P4.5 Raw P4 Check the zero point with zero gas again and perform a zero calibration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Chapter 9 Modbus Functions 9.1 Abstract This chapter lists all Modbus functions and registers supported by X-STREAM gas analyzers. Refer to the www.Modbus-IDA.org website for detailled documentation about programming the interface. At date of creation of this instruc-tion manual the following documents were used: • MODBUS Serial Line Implementation Guide: Modbus_over_serial_line_V1.pdf.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.2 Modbus - Supported Functions 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Modicon Daniel Data Type Client Access Control.Acknowledge.AllStates 2091 2091 Boolean R/W Control.Acknowledge.Failure 2092 2092 Boolean R/W Control.Acknowledge.FctChecks 2095 2095 Boolean R/W Control.Acknowledge.LevelAlarms 2096 2096 Boolean R/W Control.Acknowledge.MaintRequests 2094 2094 Boolean R/W Control.Acknowledge.OffSpecs 2093 2093 Boolean R/W Control.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Address Modicon Daniel Data Type Client Access Control.ApplyGas.Span3Valve2 2067 2067 Boolean R/W Control.ApplyGas.Span3Valve3 2071 2071 Boolean R/W Control.ApplyGas.Span3Valve4 2075 2075 Boolean R/W Control.ApplyGas.Span3Valve5 2079 2079 Boolean R/W Control.ApplyGas.Span4Valve1 2064 2064 Boolean R/W Control.ApplyGas.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Info.ProgramVersion Info.SensorVersion Info.SerialNumber PV1 PV2 PV3 PV4 PV5 Service.General.ChannelActive1 Service.General.ChannelActive2 Service.General.ChannelActive3 Service.General.ChannelActive4 Service.General.ChannelActive5 Service.General.Identification.CPLDVersion Service.General.Identification.ManufacturingInfo Service.General.Identification.ProgramVersionDate Service.General.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Service.Measurement.AbsMinRange2 Service.Measurement.AbsMinRange3 Service.Measurement.AbsMinRange4 Service.Measurement.AbsMinRange5 Service.Measurement.Compensation.PfactCoeffs1 Service.Measurement.Compensation.PfactCoeffs2 Service.Measurement.Compensation.PfactCoeffs3 Service.Measurement.Compensation.PfactCoeffs4 Service.Measurement.Compensation.PfactCoeffs5 Service.Measurement.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Service.Measurement.Compensation.ToffCoeffs3 Service.Measurement.Compensation.ToffCoeffs4 Service.Measurement.Compensation.ToffCoeffs5 Service.Measurement.Compensation.ToffEnable1 Service.Measurement.Compensation.ToffEnable2 Service.Measurement.Compensation.ToffEnable3 Service.Measurement.Compensation.ToffEnable4 Service.Measurement.Compensation.ToffEnable5 Service.Measurement.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Service.Measurement.Lin.MinValue4 Service.Measurement.Lin.MinValue5 Service.Measurement.Lin.OverflowPerc1 Service.Measurement.Lin.OverflowPerc2 Service.Measurement.Lin.OverflowPerc3 Service.Measurement.Lin.OverflowPerc4 Service.Measurement.Lin.OverflowPerc5 Service.Measurement.Lin.RangePolySet1 Service.Measurement.Lin.RangePolySet2 Service.Measurement.Lin.RangePolySet3 Service.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Data Type Client Access Float R/W Float R/W Float R/W Float R/W Float R/W Float R/W Float R/W Float R/W Float R/W Description Modicon 7341… 7350 7541… 7550 7741… 7750 7941… 7950 7151… 7160 7351… 7360 7551… 7560 7751… 7760 7951… 7960 Daniel 9671… 9675 9771… 9775 9871… 9875 9971… 9975 9576… 9580 9676… 9680 9776… 9780 9876… 9880 9976… 9980 Service.Measurement.Lin.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Service.Measurement.Lin.TableYValues5 Service.Measurement.Lin.UnderflowPerc1 Service.Measurement.Lin.UnderflowPerc2 Service.Measurement.Lin.UnderflowPerc3 Service.Measurement.Lin.UnderflowPerc4 Service.Measurement.Lin.UnderflowPerc5 Service.Measurement.Simulation.PVARawValue1 Service.Measurement.Simulation.PVARawValue2 Service.Measurement.Simulation.PVARawValue3 Service.Measurement.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Service.Status.DeviceStates.ChStateForce1 Service.Status.DeviceStates.ChStateForce2 Service.Status.DeviceStates.ChStateForce3 Service.Status.DeviceStates.ChStateForce4 Service.Status.DeviceStates.ChStateForce5 Service.Status.DeviceStates.ChStateInhibit1 Service.Status.DeviceStates.ChStateInhibit2 Service.Status.DeviceStates.ChStateInhibit3 Service.Status.DeviceStates.ChStateInhibit4 Service.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Setup.Calibration.CurrentSpangas3 Setup.Calibration.CurrentSpangas4 Setup.Calibration.CurrentSpangas5 Setup.Calibration.CurrentZerogas1 Setup.Calibration.CurrentZerogas2 Setup.Calibration.CurrentZerogas3 Setup.Calibration.CurrentZerogas4 Setup.Calibration.CurrentZerogas5 Setup.Calibration.Range1Spangas1 Setup.Calibration.Range1Spangas2 Setup.Calibration.Range1Spangas3 Setup.Calibration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Setup.Calibration.Range2Zerogas2 Setup.Calibration.Range2Zerogas3 Setup.Calibration.Range2Zerogas4 Setup.Calibration.Range2Zerogas5 Setup.Calibration.Range3Spangas1 Setup.Calibration.Range3Spangas2 Setup.Calibration.Range3Spangas3 Setup.Calibration.Range3Spangas4 Setup.Calibration.Range3Spangas5 Setup.Calibration.Range3Zerogas1 Setup.Calibration.Range3Zerogas2 Setup.Calibration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Setup.Calibration.Range4Zerogas4 Setup.Calibration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Address Data Type Client Access String R/W String R/W String R/W String R/W String R/W Modicon 3101… 3104 3105… 3108 3109… 3112 3113… 3116 3117… 3120 Daniel 3101… 3104 3105… 3108 3109… 3112 3113… 3116 3117… 3120 Setup.Display.MeasDisplay.Dis1Line1 3091 3091 Word R/W Setup.Display.MeasDisplay.Dis1Line2 3092 3092 Word R/W Setup.Display.MeasDisplay.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Setup.In/Outputs.AO.AdjustEnd5 Setup.In/Outputs.AO.AdjustStart1 Setup.In/Outputs.AO.AdjustStart2 Setup.In/Outputs.AO.AdjustStart3 Setup.In/Outputs.AO.AdjustStart4 Setup.In/Outputs.AO.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Address Modicon Daniel Data Type Client Access Setup.In/Outputs.AO.Hold2 3280 3280 Word R/W Setup.In/Outputs.AO.Hold3 3285 3285 Word R/W Setup.In/Outputs.AO.Hold4 3290 3290 Word R/W Setup.In/Outputs.AO.Hold5 3295 3295 Word R/W Setup.In/Outputs.AO.NumberOuts 3270 3270 Word R/W Setup.In/Outputs.AO.OutRange1 3272 3272 Word R/W Setup.In/Outputs.AO.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Setup.In/Outputs.DO.Signal_DO14 Setup.In/Outputs.DO.Signal_DO15 Setup.In/Outputs.DO.Signal_DO16 Setup.In/Outputs.DO.Signal_DO17 Setup.In/Outputs.DO.Signal_DO18 Setup.In/Outputs.DO.Signal_DO19 Setup.In/Outputs.DO.Signal_DO2 Setup.In/Outputs.DO.Signal_DO20 Setup.In/Outputs.DO.Signal_DO21 Setup.In/Outputs.DO.Signal_DO22 Setup.In/Outputs.DO.Signal_DO3 Setup.In/Outputs.DO.Signal_DO4 Setup.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Modicon Daniel 6055… 9028 Setup.Measurement.EndOfCurrentRange_4 6056 6059… 9030 Setup.Measurement.EndOfCurrentRange_5 6060 6143… 9072 Setup.Measurement.EndOfRange1_1 6144 6159… 9080 Setup.Measurement.EndOfRange1_2 6160 6175… 9088 Setup.Measurement.EndOfRange1_3 6176 6191… 9096 Setup.Measurement.EndOfRange1_4 6192 6207… 9104 Setup.Measurement.EndOfRange1_5 6208 6147… 9074 Setup.Measurement.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Setup.Measurement.StartOfCurrentRange_2 Setup.Measurement.StartOfCurrentRange_3 Setup.Measurement.StartOfCurrentRange_4 Setup.Measurement.StartOfCurrentRange_5 Setup.Measurement.StartOfRange1_1 Setup.Measurement.StartOfRange1_2 Setup.Measurement.StartOfRange1_3 Setup.Measurement.StartOfRange1_4 Setup.Measurement.StartOfRange1_5 Setup.Measurement.StartOfRange2_1 Setup.Measurement.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Setup.Measurement.StartOfRange4_5 Address Modicon Daniel 6217… 9109 6218 Data Type Client Access Float R/W Description start of range4 of comp5 Compute cross interference factor comp1 (0=Src1, 1=Src2 etc.) Compute cross interference factor comp2 (0=Src1, 1=Src2 etc.) Compute cross interference factor comp3 (0=Src1, 1=Src2 etc.) Compute cross interference factor comp4 (0=Src1, 1=Src2 etc.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Address Modicon 6737… Setup.Measurement.XIntf.LinearReferences3 6744 6797… Setup.Measurement.XIntf.LinearReferences4 6804 6857… Setup.Measurement.XIntf.LinearReferences5 6864 4051… Setup.Measurement.XIntf.SignalStates1 4054 4055… Setup.Measurement.XIntf.SignalStates2 4058 4059… Setup.Measurement.XIntf.SignalStates3 4062 4063… Setup.Measurement.XIntf.SignalStates4 4066 4067… Setup.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Status.Diagnostics.DSPMuxValue1 Status.Diagnostics.DSPMuxValue2 Status.Diagnostics.DSPMuxValue3 Status.Diagnostics.DSPMuxValue4 Status.Diagnostics.DSPMuxValue5 Status.Diagnostics.DSPMuxValue6 Status.Diagnostics.DSPMuxValue7 Status.Diagnostics.DSPMuxValue8 Status.Diagnostics.RawMeasConce1 Status.Diagnostics.RawMeasConce2 Status.Diagnostics.RawMeasConce3 Status.Diagnostics.RawMeasConce4 Status.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Status.Diagnostics.RawRefConce5 Status.Flow1 Status.Flow2 Status.Flow3 Status.Flow4 Status.Flow5 Status.NamurStates.FailAlarmComp1 Status.NamurStates.FailAlarmComp2 Status.NamurStates.FailAlarmComp3 Status.NamurStates.FailAlarmComp4 Status.NamurStates.FailAlarmComp5 Status.NamurStates.FailAlarmDev Status.NamurStates.FailureActive Status.NamurStates.FailureAlarm Status.NamurStates.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Status.NamurStates.MaintRequAlarmComp1 Status.NamurStates.MaintRequAlarmComp2 Status.NamurStates.MaintRequAlarmComp3 Status.NamurStates.MaintRequAlarmComp4 Status.NamurStates.MaintRequAlarmComp5 Status.NamurStates.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Status.Pressure2 Status.Pressure3 Status.Pressure4 9-30 Address Modicon Daniel 6063… 9032 6064 6065… 9033 6066 6067… 9034 6068 Data Type Client Access Float RO pressure of comp2 Float RO pressure of comp3 Float RO pressure of comp4 Description Emerson Process Management GmbH & Co.
Instruction Manual HASXEE-IM-HS 04/2010 X-STREAM XE 9.4 List of Parameters and Registers - Sorted by Daniel Registers Emerson Process Management GmbH & Co. OHG 9-31 9 Address Data Client Description Tag Name Type Access Daniel Modicon 2001 2001 Control.Calibration.Zero_1 Boolean R/W Zero calibration comp1 (1=start) 2002 2002 Control.Calibration.Zero_2 Boolean R/W Zero calibration comp2 (1=start) 2003 2003 Control.Calibration.Zero_3 Boolean R/W Zero calibration comp3 (1=start) 2004 2004 Control.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers 3039 3040 3040 3091 3091 3092 3092 3093 3093 3094 3094 3095 3095 3096 3096 3097 3097 3098 3098 3099 3099 3100 3100 3151 3152 3153 3154 3156 3157 3158 3159 3161 3162 3163 3164 3165 3166 3167 3168 3169 3171 3172 3173 3174 3175 3151 3152 3153 3154 3156 3157 3158 3159 3161 3162 3163 3164 3165 3166 3167 3168 3169 3171 3172 3173 3174 3175 Setup.Display.Component.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Data Type Client Access Description Word R/W signal of SHS Pump1 3221 3222 3223 3224 3225 3226 3227 3228 3229 3230 3231 3232 3233 3234 3236 3237 3238 3239 3240 3241 3242 3243 3244 3245 3246 3247 3248 3249 3270 3271 3272 Setup.In/Outputs.SHS.Signal_ Pump2 Setup.In/Outputs.DI.Node_DI1 Setup.In/Outputs.DI.Node_DI2 Setup.In/Outputs.DI.Node_DI3 Setup.In/Outputs.DI.Node_DI4 Setup.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Address Tag Name Daniel Modicon 3281 3281 Setup.In/Outputs.AO.SignalAsgn3 3282 3282 Setup.In/Outputs.AO.OutRange3 Data Type Word Word 3283 3283 Setup.In/Outputs.AO.AutoScale3 Word 3284 3284 Setup.In/Outputs.AO.FailMode3 Word 3285 3286 3287 3285 3286 3287 Setup.In/Outputs.AO.Hold3 Setup.In/Outputs.AO.SignalAsgn4 Setup.In/Outputs.AO.OutRange4 Word Word Word 3288 3288 Setup.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers 3404 3405 3405 3406 3406 3407 3407 3408 3408 3409 3409 3410 3410 3411 3411 3412 3412 3413 3413 3414 3414 3415 3415 3416 3416 3417 3417 3418 3418 3419 3419 3420 3420 3421 3421 3422 3422 3423 3423 3424 3424 3425 3425 3426 3426 3427 3427 Service.Measurement.Compensation.ToffEnable4 Service.Measurement.Compensation.ToffEnable5 Service.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Address Daniel Modicon Tag Name Service.Measurement.Compensation.PfactSensorAssign3 Service.Measurement.Compensation.PfactSensorAssign4 Service.Measurement.Compensation.PfactSensorAssign5 Service.Measurement.Simulation. SimPVARawEnable1 Service.Measurement.Simulation. SimPVARawEnable2 Service.Measurement.Simulation. SimPVARawEnable3 Service.Measurement.Simulation.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Data Type Client Access Float R/W end of current range of comp3 Float R/W start of current range of comp4 Float R/W end of current range of comp4 Float R/W start of current range of comp5 Float R/W end of current range of comp5 Status.Pressure1 Float RO pressure of comp1 Status.Pressure2 Float RO pressure of comp2 Status.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Tag Name Setup.Calibration.Range3Spangas2 Setup.Calibration.Range4Spangas2 Setup.Calibration.Range1Spangas3 Setup.Calibration.Range2Spangas3 Setup.Calibration.Range3Spangas3 Setup.Calibration.Range4Spangas3 Setup.Calibration.Range1Spangas4 Setup.Calibration.Range2Spangas4 Setup.Calibration.Range3Spangas4 Setup.Calibration.Range4Spangas4 Setup.Calibration.Range1Spangas5 Setup.Calibration.Range2Spangas5 Setup.Calibration.Range3Spangas5 Setup.Calibration.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Tag Name Data Type Client Access Setup.In/Outputs.AO.AdjustStart2 Float R/W fine adjustment for start range of output2 Setup.In/Outputs.AO.AdjustEnd2 Float R/W fine adjustment for end range of output2 Setup.In/Outputs.AO.StartRange3 Float R/W Setup.In/Outputs.AO.EndRange3 Float R/W Setup.In/Outputs.AO.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Tag Name Data Type Setup.Display.Component.PV_UnitString String5 Client Access Description R/W unit displayed for comp5 String R/W displayed tag for component1 Setup.Display.Component.Tag2 String R/W displayed tag for component2 Setup.Display.Component.Tag3 String R/W displayed tag for component3 Setup.Display.Component.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Tag Name Service.Status.DeviceStates.ChStateInhibit1 Service.Status.DeviceStates.ChStateInhibit2 Service.Status.DeviceStates.ChStateInhibit3 Service.Status.DeviceStates.ChStateInhibit4 Service.Status.DeviceStates.ChStateInhibit5 Service.Status.DeviceStates.DvcStateForce Service.Status.DeviceStates.ChStateForce1 Service.Status.DeviceStates.ChStateForce2 Service.Status.DeviceStates.ChStateForce3 Service.Status.DeviceStates.ChStateForce4 Service.Status.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Setup.Measurement.XIntf.SignalValues5 Setup.Measurement.XIntf.InterfereFactors5 Setup.Measurement.XIntf.LinearReferences5 Setup.Measurement.XIntf.LinearPolyCoeffs5 Service.Measurement.Simulation. SimXSPMuxValue1 Service.Measurement.Lin.TableXValues1 Service.Measurement.Lin.TableYValues1 Service.Measurement.Lin.Set1Coeffs1 Service.Measurement.Lin.Set2Coeffs1 Service.Measurement.Lin.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 9.4 List of Parameters and Registers - Sorted by Daniel Registers Address Daniel Modicon 9766… 7531… 9770 7540 9771… 7541… 9775 7550 9776… 7551… 9780 7560 9781… 7561… 9784 7568 9785… 7569… 9788 7576 9801… 7601… 9803 7606 9831… 7661… 9833 7666 9861… 7721… 9865 7730 9866… 7731… 9870 7740 9871… 7741… 9875 7750 9876… 7751… 9880 7760 9881… 7761… 9884 7768 9885… 7769… 9888 7776 9901… 7801… 9903 7806 9931… 7861… 9933 7866 9-56 Tag Name Service.Measurement.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Chapter 10 Service Information 10.1 Return of Material Secure a return authorization from a Rosemount Analytical Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted. 2. In no event will Rosemount be responsible for equipment without proper authorization and identification. 3.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 10 Service Information 10.2 Customer Service 10.3 Training For order administration, replacement parts, applicaton assistance, on-site or factory repair, service or maintenance contract information, contact: In Europe: Emerson Process Management GmbH & Co. OHG Service Department Germany T +49 6055 884-470/-472 A comprehensive Factory Training Program of operator and service classes is available.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Chapter 11 Dismounting and Disposal 11.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 11 Dismounting & Disposal When the instrument has reached the end of its useful life, do not throw it in a trash can! This instrument has been made of materials to be recycled by recyclers specialised in this field. Let the instrument and the packing material duly and environmently friendly be disposed of. Ensure the equipment is free of dangerous and harmful substances (decontaminated). Take care of all local regulations for waste treatment.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Appendix This chapter contains page A-2 EC declaration of conformity page A-12 CSA Certificate of Compliance page A-14 Block diagram page A-21 Water Vapor: Conversion of Dewpoint, Vol.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.1 Modbus Specification A.1 Modbus Specification A-2 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Modbus Specification A Appendix A.1 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.1 A-4 Modbus Specification Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Modbus Specification A Appendix A.1 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.1 A-6 Modbus Specification Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Modbus Specification A Appendix A.1 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.1 A-8 Modbus Specification Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Modbus Specification A Appendix A.1 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.1 A-10 Modbus Specification Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Modbus Specification A Appendix A.1 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.2 EC Declaration of Conformity A.2 EC Declaration of Conformity A-12 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Appendix A.2 EC Declaration of Conformity Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.3 CSA Certificate of Compliance A.3 CSA Certificate of Compliance Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2247530 Date Issued: 2010/01/21 Issued to: Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.3 CSA Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2247530 Date Issued: 2010/01/21 • X-STREAM Gas Analyzer Core (X2CA) Table Top or Rack Mount with appliance inlet for indoor use and no display (optional with field wiring terminals for indoor use); Gas analyzer, Model: X-STREAM, rated 100-240Vac, 50/60 Hz, 5.5 - 3A, Class I, Pollution Degree II.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.3 CSA Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2247530 Date Issued: 2010/01/21 Class I, Zone 1, AEx d IIB+H2, T3 Gas analyzer , Model: X-STREAM, rated 100-240Vac, 50/60 Hz, 3 - 1.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.3 CSA Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2247530 Date Issued: 2010/01/21 CAN/CSA-C22.2 No. 94-M91 (R2001) - Special Purpose Enclosures CAN/CSA C22.2 No. 60529:05 - Degrees of protection provided by enclosure (IP Code) ANSI/ISA-12.00.01-2002 (IEC 60079-0 Mod) - Electric Apparatus for Use in Class I, Zones 0, 1 & 2 Hazardous (Classified) Locations: General Requirements ANSI/ISA-12.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.4 Block Diagram A Appendix X-STREAM Version A.4 Block Diagram Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.4 Block Diagram A-22 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Appendix A.4 Block Diagram Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.4 Block Diagram A-24 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Appendix A.4 Block Diagram Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.4 Block Diagram A-26 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Appendix A.4 Block Diagram Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.4 Block Diagram A-28 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Appendix A.4 Block Diagram Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.4 Block Diagram A-30 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Appendix A.4 Block Diagram Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.4 Block Diagram A-32 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Appendix A.4 Block Diagram Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.5 Calculation of Water Vapor A.5 Water Vapor: Relationship of Dewpoint, Vol.-% and g/Nm³ Dewpoint Content of Water Water Concentration °C °F Vol. -% g/Nm³ Dewpoint Content of Water Water Concentration °C °F Vol.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.6 Declaration of Decontamination A.6 Declaration of Decontamination Because of legal regulations and for the safety of Emerson Process Management employees and operating equipment, we need this "Declaration of Decontamination", signed by an authorized person, prior to processing your order. Ensure to include it with the shipping documents, or (recommended) attach it to the outside of the packaging.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.7 PLC Quick Reference A.7 PLC Quick Reference PLC Quick Reference Card Rev. 2010‐04 PLC Quick Reference Card If timers need to be used, they have to be setup at the beginning of the program (see Example Program) For USB transfer save PLC.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.7 PLC Quick Reference PLC Quick Reference Card Rev.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.7 PLC Quick Reference PLC Quick Reference Card Rev.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.7 PLC Quick Reference PLC Quick Reference Card Rev.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.7 PLC Quick Reference PLC Quick Reference Card Rev.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.7 PLC Quick Reference PLC Quick Reference Card Rev.
X-STREAM XE A-42 Instruction Manual HASXEE-IM-HS 04/2010 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A.8 Assignment of Terminals and Sockets A.8 Assignment of Terminals and Sockets A.8.
P3.1 P3.2 P3.3 P3.4 P3.5 P3.6 P3.7 P3.8 P3.9 P3.10 P3.11 P3.
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 A Access levels see Menu access Acknowledgements 6-17, 7-17, 7-19, 7-40 Advanced Calibration 6-8 Alarms 6-80 Setup 6-80 Alarms Status 6-119 Ambient temperature 1-13, 2-2, 3-9, 3-15, 4-3 Analog Input 6-74 Analog inputs 4-28, 6-57, 6-75 Conversion 6-76 Analog outputs see Outputs: Analog outputs Analog Outputs 6-63 Apply gas 6-16, 7-17, 7-19, 7-40 ATEX 1-1, 1-2, 1-13, 1-23...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Index Cross Interference Linearization Curve 6-58 CSA 2-2, A-1, A-14...A-20 Certificate of Compliance A-1, A-14...A-20 D Damping see Setup Menu Data logger 1-4, 6-19, 6-21, 6-105, 6-106, 6-122, 7-66...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Setup 6-88 Intended use statement S-5 Internal SHS 6-62, 6-73, 7-12 see also Setup Menu: Internal SHS Interval Time 6-40 see Calibration: Interval Time IP address see Ethernet: Setup; see also Info menu J K L Labels 6-30 Language 1-3, 5-12, 5-13, 6-21, 6-22 Leak test S-8, 7-1, 7-4, 8-15 Leak Test 7-4 Licenses 6-89 Linearization Setup 6-48 Local backup 6-94 Lock menus 6-2, 6-32, 7-17, 7-19, 7-40 Log files see Logger: Log files Handling see Logger: Log f
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Index Flow monitor switch 1-6 Flow sensor 1-6, 6-88 Heated area 1-8 Installed options 5-14, 6-36...6-40, 6-88...
Instruction Manual X-STREAM XE HASXEE-IM-HS 04/2010 Index T Tags 6-30 Technical data 2-1 Model-specific 2-6 Temperature Alarms 6-85 Temperature Alarms 6-85 Temperature Setup 6-27 Terminal adapters see Screw terminals; see Terminals Terminals 1-27, 4-21, 4-24, 4-27, 4-28, A-1, A-43, A-44 Time 6-102 Manual time setup 6-102 NTP Server 6-101 Realtime clock 1-4 Setup 6-100 Time server 1-4 Tool see Status Line Troubleshooting 8-1 U USB Files 7-71 Preparation 7-61 Supported Mass Storage Device Types 7-61 USB
X-STREAM XE I-6 Instruction Manual HASXEE-IM-HS 04/2010 Emerson Process Management GmbH & Co.
Rosemount Analytical Process Gas Analyzer Center of Excellence Emerson Process Management GmbH & Co. OHG Industriestrasse 1 63594 Hasselroth Germany T +49 6055 884 0 F +49 6055 884 209 Rosemount Analytical Inc. Gas Analyzer Service Center 6565P Davis Industrial Parkway Solon, OH 44139 USA T +1 440.914.1261 Toll Free in US and Canada 800.433.6076 F +1 440.914.1262 e-mail: gas.csc@emerson.com US Response Center 800.654.