Troubleshooting guide

78
Reference Manual
00809-0100-4308, Rev BA
Section 6: Service and Troubleshooting
December 2014
Service and Troubleshooting
6.3 Troubleshooting guide
If there is a malfunction despite the absence of alerts, see Tab le 6 .3 .1 for information on
possible causes and recommended actions.
The troubleshooting guide contains the following symptoms:
Incorrect level readings, see page 78.
Incorrect or missing interface level reading, see page 80.
Power module troubleshooting, see page 81.
Device display troubleshooting, see page 81.
Wireless Network troubleshooting, see page 82.
6.3.1 Incorrect level readings
Symptom Possible cause and recommended actions
The level readings do not correspond to a
reference measurement, for example a
handgauged value.
Check the Tank Height parameter, refer to
“Tank Height” on page 161.
Check Thresholds, refer to “Adjusting
thresholds” on page 84.
Run Verify Level, see “Verify Level” on
page 59.
Check transmitter configuration. Run Basic
Setup, refer to “Basic Setup” on page 58.
There is no level reading.
The tank is empty. No action is needed.
Check Thresholds, refer to “Adjusting
thresholds” on page 84.
Level spikes or level is suddenly reported as full
or empty.
Check the Upper Product Dielectric
Constant, see “Upper Product Dielectric
Constant” on page 163.
The transmitter is configured with wrong
Probe Type, refer to “Probe Type” on
page 161.
Check Thresholds, see “Adjusting
thresholds” on page 84.
The transmitter has locked on disturbing
obstacles at top of the tank. See “Handling
disturbances at the top of the tank” on
page 92 for recommended actions.
The surface is turbulent. Set the
Performance Mode to High to get a stable
measurement signal, refer to “Noise or weak
surface echoes” on page 97.