Troubleshooting guide

36
Reference Manual
00809-0100-4308, Rev BA
Section 3: Installation
December 2014
Installation
3.6.4 Tank connection with Tri-Clamp
Figure 3-16. Tank connection with Tri-Clamp
1. Place a gasket on top of the tank flange.
2. Lower the transmitter and probe into the tank.
3. Fasten the Tri-Clamp to the tank with a clamp.
4. Loosen the nut that connects the transmitter head to the probe slightly.
5. Rotate the transmitter head so the device display faces the desired direction.
6. Tighten the nut. Max torque is 30 Lbft (40 Nm).
7. Continue with the grounding step.
A. Transmitter head
B. Nut
C. Tri-Clamp
D. Probe
E. Clamp
F. G as k et
G.Tank