Troubleshooting guide
177
Reference Manual
00809-0100-4308, Rev BA
Appendix D: Configuration Parameters
December 2014
Configuration Parameters
Near Zone Threshold
Near Zone Trimming
Threshold Control The Near Zone Threshold can be automatically calculated by
device or manually set by user.
Threshold Threshold to filter out noise in a zone near the device. Noise
below the threshold is suppressed. This threshold replaces
the Surface threshold in the zone where it is applicable.
Distance Distance from Upper Reference Point (normally the lower side
of device flange) to point where the near zone threshold ends.
Trim Near Zone Select to activate trimming.
The Trim Near Zone method is used to fine tune performance
in the area close to the tank top. This function is normally not
used. Trim Near Zone is typically used if there are problems
related to the nozzle. For more information see section
“Using the Trim Near Zone function” on page 94.
Near Zone Has Been
Trimmed
Indicates if the Trim Near Zone method has been used for this
device.