Appendix D: Configuration Parameters
December 2014
Configuration Parameters
Geometry
Probe Angle
(Only applicable to rigid
probes)
Defines the angle compared to the plumb line at which the
device with probe is mounted (0 means that probe is
mounted vertically).
Enter the angle between the probe and the vertical line. Do
not change this value if the transmitter is mounted with the
probe along the vertical line (which is normally the case).
Remote Housing If the transmitter head is mounted apart from the probe, the
length of cable between probe and remote housing must be
configured.
User Defined Probe Settings Parameters for user defined probe.
Note
These settings should only be modified for customized
probes. The settings are typically provided by factory.
Calibration Offset Difference between surface distance measured by device and
the same distance measured by e.g. handgauging with a
measurement tape. A positive Calibration Offset value will
increase the presented Level value.
Show Level Below Probe End
as Zero
When this setting is selected and the product surface is at or
below the probe end, the level measurement output will be
zero.
Note
Only applicable for negative probe end peak.
0
o
Probe Angle