Troubleshooting guide

166
Reference Manual
00809-0100-4308, Rev BA
Appendix D: Configuration Parameters
December 2014
Configuration Parameters
Units
The units for length, volume, and temperature are selectable. After appropriate units have been
selected, all configuration parameters and transmitter variables will be expressed in these units.
HART - Variable Mapping
HART - Percent of Range
HART - Data Collection
HART - Variable History
Length Unit Used unit for Level and Interface Level values.
Volume Unit Used unit for Volume values.
Temperature Unit Used unit for Electronics Temperature value.
Primary Variable Primary dynamic variable in the HART protocol which will be
assigned as a variable from the device.
Secondary Variable Second dynamic variable in the HART protocol which will be
assigned as a variable from the device.
Third Variable Third (Tertiary) dynamic variable in the HART protocol which
will be assigned as a variable from the device.
Fourth Variable Fourth (Quaternary) dynamic variable in the HART protocol
which will be assigned as a variable from the device.
Upper Range Value Value for Primary Variable (PV) corresponding to 100% range.
Lower Range Value Value for Primary Variable (PV) corresponding to 0% range.
Upper Sensor Limit The upper boundary for the range over which the sensor
works properly.
Lower Sensor Limit The lower boundary for the range over which the sensor
works properly.
Measurement and Status
Log
Alternatives for data collection in the device.
Configure Data History Data History is a series of 12 data points stored in the
transmitter. To enable Data History trending select either to
enable single data point trending (recommended) or enable
filtered trending.
If Data History is enabled, select which Device Variable to
store, and then type the time between each sample into the
Sample Interval box (4 to 7200 seconds).