Troubleshooting guide

94
Reference Manual
00809-0100-4308, Rev BA
Section 6: Service and Troubleshooting
December 2014
Service and Troubleshooting
Using the Trim Near Zone function
The Trim Near Zone function is used to fine tune performance in the area close to the tank top.
Normally it is not necessary to use the function, but if you experience problems related to the
nozzle, pipe, or chamber installation, you may need to use this function.
Figure 6-14 describes the Trim Near Zone function and its effect on the echo curve. This effect
is only visible if measurement conditions so require.
Figure 6-14. Echo Curve before and after Trim Near Zone
Note
Make sure the product level is below the Near Zone region (0-3.3 ft. (0-1 m) below the Upper
Reference Point) before performing the Trim Near Zone.
Note
The Trim Near Zone function should only be used for reducing impact from stationary
disturbances. For occasional disturbances, use the Near Zone Threshold.
To use the Trim Near Zone function, do the following:
1. From the Home Screen, go to Service Tools > Echo Tuning and click Near Zone.
2. Under Near Zone Compensation, click Trim Near Zone and follow the on-screen
instructions.
Changing the Upper Null Zone
Measurements are not performed within the Upper Null Zone (UNZ). The Upper Null Zone can
be extended to block out disturbing echoes close to the tank top, caused by for example a
narrow nozzle with rough walls, obstacles close to the probe, or a nozzle that protrudes into the
tank. See Figure 6-16.
Note
Desired measuring range must be below the Upper Null Zone.
Trim Near Zone
Reference Peak Reference Peak